| Literature DB >> 34527327 |
Yao Zhang1, Haozhe Huang2, Shuhui Cao1, Jiajun Teng1, Yue Wang1, Xuxinyi Ling1, Yan Zhou1, Jianlin Xu1, Wentao Li2, Hua Zhong1.
Abstract
BACKGROUND: The purpose of this study was to retrospectively evaluate the clinical value of an electromagnetic navigation system for CT-guided percutaneous lung biopsy of peripheral lung lesions.Entities:
Keywords: CT-guided biopsy; Lung lesions; efficiency; electromagnetic navigation; lung biopsy; safety
Year: 2021 PMID: 34527327 PMCID: PMC8411150 DOI: 10.21037/jtd-21-395
Source DB: PubMed Journal: J Thorac Dis ISSN: 2072-1439 Impact factor: 2.895
Figure 1The components of Electromagnetic navigation system components. (A) Computer and monitor (yellow arrow), field emitter (red arrow); (B) data collection chip (arrow); (C) patch sensor (arrow).
Figure 2The electromagnetic navigation system guided lung biopsy. (A) The operator can visualize the needle advancement using the electromagnetic navigation system in real time; (B) the enlarged view of the screen.
Figure 3The electromagnetic navigation system’s screen capture, which shows the position of the digital tip (cross) of the biopsy needle and needle route (line) in the axial and sagittal planes.
Clinical characteristics of patients and their lesions
| Variables | Group A (n=141) | Group B (n=96) | P value |
|---|---|---|---|
| Age, year (median, range) | 65 [32–79] | 65 [34–80] | 0.1871 |
| Sex, n (%) | 0.5362 | ||
| Male | 78 (55.3) | 57 (59.4) | |
| Female | 63 (44.7) | 39 (40.6) | |
| Puncture position, n (%) | 0.2182 | ||
| Supine | 67 (47.5) | 39 (40.6) | |
| Lateral | 5 (3.5) | 8 (8.3) | |
| Prone | 69 (48.9) | 49 (51.0) | |
| Lesion maximum diameter, mm (mean ± SD) | 19.8±6.4 | 20.9±5.9 | 0.2113 |
| Lesion lobar location, n (%) | 0.8232 | ||
| Right upper | 33 (23.4) | 21 (21.9) | |
| Right middle | 13 (9.2) | 6 (6.3) | |
| Right lower | 23 (16.3) | 19 (19.8) | |
| Left upper | 46 (32.6) | 29 (30.2) | |
| Left lower | 26 (18.4) | 21 (21.9) | |
| Distance from pleura, n (%) | 0.8242 | ||
| ≤2 cm | 95 (67.4) | 66 (68.8) | |
| >2 cm | 46 (32.6) | 30 (31.3) | |
| Lesion characteristics, n (%) | 0.3912 | ||
| Mixed ground glass opacity | 20 (14.2) | 8 (8.3) | |
| Pure solid | 113 (80.1) | 82 (85.4) | |
| Others | 8 (5.7) | 6 (6.3) | |
| Lesion histopathology, n (%) | 0.0662 | ||
| Benign | 51 (36.2) | 25 (26.0) | |
| Malignant | 86 (61.0) | 63 (65.6) | |
| Non-defined | 4 (2.8) | 8 (8.3) | |
| Emphysema, n (%) | 30 (21.3) | 15 (15.6) | 0.2762 |
| Pulmonary fibrosis, n (%) | 13 (9.2) | 5 (5.2) | 0.2522 |
| Pulmonary bullae, n (%) | 26 (18.4) | 12 (12.5) | 0.2212 |
1, Mann-Whitney U test; 2, Chi-square test/Fishers’ exact test; 3, independent t-test.
Technical information
| Variables | Group A (n=141) | Group B (n=96) | P value |
|---|---|---|---|
| Operation time, min (median, range) | 20 [3–48] | 18 [6–42] | 0.8511 |
| Number of adjust times in CT-scanning, n (median, range) | 3 [0–15] | 3 [1–11] | 0.7511 |
| Distance between skin entry point and target lesion, mm (mean ± SD) | 65.7±19.1 | 67.1±19.0 | 0.5852 |
| Technical success rate, n (%) | 141 (100.0) | 96 (100.0) | – |
| Technical efficacy rate, n (%) | 131 (92.9) | 87 (90.6) | 0.5253 |
1, Mann-Whitney U test; 2, independent t-test; 3, Chi-square test/Fishers’ exact test.
Biopsy-related complications and grading by the Society of Interventional Radiology guidelines
| Complications | Group A (n=141) (%) | Group B (n=96) (%) | P value* |
|---|---|---|---|
| Incidence of complications | 52 (36.9) | 35 (36.5) | 0.947 |
| Grade of complications | |||
| Grade 1–2 | 52 (36.9) | 35 (36.5) | – |
| Type of complications | |||
| Pneumothorax | 29 (20.6) | 16 (16.7) | 0.452 |
| Hemorrhage | 20 (14.2) | 17 (17.7) | 0.463 |
| Hemoptysis | 3 (2.1) | 2 (2.1) | 1.000 |
*, the P value was calculated using the chi-square test.