Standardization of PD-L1 immunohistochemistry.

Sandra Martinez-Morilla1, Myrto Moutafi1, David L Rimm2.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 34508229      PMCID: PMC8860739          DOI: 10.1038/s41379-021-00917-4

Source DB:  PubMed          Journal:  Mod Pathol        ISSN: 0893-3952            Impact factor:   7.842


× No keyword cloud information.
The immunohistochemistry (IHC) assays for PD-L1 have been a source of great uncertainty in both oncologists and pathologists. The wide range of FDA-approved assays with differential sensitivity and scoring system[1], and the lack of success in harmonization of these assays[2,3] has led to confusion in pathology labs. Currently, in breast cancer, the oncologist’s treatment plan must either be known in advance of the assay performance, or two non-concordant assays (Ventana SP142 and Agilent 22c3) must be performed since each qualifies patients for different PD-1 axis therapy (Atezolizumab and Pembrolizumab, respectively). Although the problem of different scoring systems is not easily resolved, in this issue of Modern Pathology, Sompuram et al. provide potential tools to standardize the sensitivity of the assays and to maintain lab-to-lab or week-to-week reproducibility of assessment of PD-L1 expression. In this study, a new tool is described that allows evaluation of IHC assays using analytic parameters including limit of detection (LOD) and dynamic range of PD-L1. This National Institute of Standards (NIST)-based tool uses calibrators consisting of microbeads coated with a range of peptides representing the antigenic epitope portion of PD-L1 for each antibody/assay. In each spot of matrix containing beads on the slide, the number of molecules of the protein per bead has been previously determined spectroscopically. This system has been explored by the authors for a number of applications, including a recently published study with the calibrators for estrogen receptor earlier this year[4], piloting methods that are then used for the PD-L1 calibrators toward standardization of many of the current PD-L1 assays. In the current study, the PD-L1 calibrators were used to compare four FDA-approved PD-L1 IHC assays and a series of LDT assays, among 41 laboratories from four different countries. After image analysis, they observed that the four FDA-approved PD-L1 assays could be grouped in three different levels of analytic sensitivity: SP263 - 22C3/28-8 - SP142 in order of decreasing analytic sensitivity (increasing LOD). In fact, this approach shows that in some labs the Ventana SP142 assay is analytically ten times less sensitive than the Ventana SP263 assay. The authors also show that the performance of the assessed LDT PD-L1 assays was similar to FDA-approved assays in many labs, with respect to LOD and analytic response curves. Can this tool be used to harmonize the PD-L1 assay and potentially use a single assay to provide the correct information for each drug? Unfortunately, no. The reason is that the FDA approved PD-L1 assays have two components, the biochemical part of the assay which happens in the autostainer or on the slide, and the interpretive part, which happens when the pathologist converts what they see on the slide to a “score”. The score part cannot be addressed by this tool, but the biochemical part can! In spite of the common use of autostaining platforms, all of the PD-L1 IHC assays have shown to vary in intensity from week to week by as much as two SD[5]. This calibrator tool can be used to show week-to-week or month-to-month reproducibility of any assay and alert a lab director to a light or dark batch. Although Sompuram et al. do not show any images of the bead spots and only measure the beads using a stain intensity percentage (not explained in this paper but defined in Vani et al.[6] as a complex MatLab algorithm), the inventor envisioned “by eye” assessment in clinical labs, where the beads appear increasingly brown by conventional light microscopic evaluation (see Fig. 1 produced in our lab). Using this tool, the IHC lab director could assess, by eye, the lightest positive spot, compared to the last negative spot and be sure that every stainer run in their lab results in the same threshold cut-point. This strategy should work regardless of which assay is used. Even though SP263 would show a different threshold than 22c3 and both would be more sensitive than SP142, the lab could be sure that they have reproducibility beyond that achievable with a single cell line control supplied by Agilent or the suggestion of a tonsil control in the label of the SP142 assay. Furthermore, using calibrators at the initial stages of assay development could lead to more accurate diagnostic tests that would stratify patients precisely during the clinical trial with standardization to NIST and actual numbers of molecules detected by the assay. The use of this tool could impact patient care in the near term as the patients are more likely to be correctly scored and receive the appropriate treatment based on an accurate and standardized measurement of the specific biomarker.
Fig. 1

Representative image showing a panoramic view of the PD-L1 calibrator slide and three sequential levels of PD-L1 expression after DAB E1L3N LDT assay in two different magnification scales (100 µm top row and 20 µm bottom row).

Level 4 corresponds to 150,157 PD-L1 molecules/bead, Level 5 287,131 molecules/bead and Level 6 642,523 molecules/bead. In this case Level 4 is below the LOD for this assay where Levels 5 and 6 are above. The small brown beads are used as a standard for automated reading. The larger beads show some refractility, but on the microscope, the last negative spot (Level 4) can be distinguished from the first positive spot (Level 5, the LOD). PD-L1 Programmed Death cell Ligand 1, DAB 3,3′-Diaminobenzidine, LDT Laboratory Development Test.

Representative image showing a panoramic view of the PD-L1 calibrator slide and three sequential levels of PD-L1 expression after DAB E1L3N LDT assay in two different magnification scales (100 µm top row and 20 µm bottom row).

Level 4 corresponds to 150,157 PD-L1 molecules/bead, Level 5 287,131 molecules/bead and Level 6 642,523 molecules/bead. In this case Level 4 is below the LOD for this assay where Levels 5 and 6 are above. The small brown beads are used as a standard for automated reading. The larger beads show some refractility, but on the microscope, the last negative spot (Level 4) can be distinguished from the first positive spot (Level 5, the LOD). PD-L1 Programmed Death cell Ligand 1, DAB 3,3′-Diaminobenzidine, LDT Laboratory Development Test. While the work of Sompuram et al. has the potential to change practice in pathology labs across the world, there are limitations. The most significant limitation is that this standardization approach does not address the problems associated with interpretation or scoring of PD-L1 IHC[7,8]. Until there are automated objective systems that can assess PD-L1 expression levels and localization, that issue is likely to generate a lot of publications but will not be easily solved. Another limitation is the substrate itself. While the peptides are designed to carefully mimic the antibody recognition sequence, the antigen is synthetic and may not exactly mimic the antigen seen in real human tumor tissue. However, this limitation should not affect the key utility of this tool as a standard. Finally, the ability to easily define the PD-L1 expression threshold bead number can be affected by the counterstain. Good laboratory practice, with minimal hematoxylin counterstain can address this limitation. In summary, Sompuram et al. have produced a NIST-standardized calibrator tool that can solve the biochemical aspects of the variability of the PD-L1 assays, and its incorporation into IHC labs is likely to increase the chances that the right patients get the right drugs.
  8 in total

1.  The PD-L1 Immunohistochemistry Biomarker: Two Steps Forward, One Step Back?

Authors:  Keith M Kerr
Journal:  J Thorac Oncol       Date:  2018-03       Impact factor: 15.609

2.  PD-L1 Immunohistochemistry Comparability Study in Real-Life Clinical Samples: Results of Blueprint Phase 2 Project.

Authors:  Ming Sound Tsao; Keith M Kerr; Mark Kockx; Mary-Beth Beasley; Alain C Borczuk; Johan Botling; Lukas Bubendorf; Lucian Chirieac; Gang Chen; Teh-Ying Chou; Jin-Haeng Chung; Sanja Dacic; Sylvie Lantuejoul; Mari Mino-Kenudson; Andre L Moreira; Andrew G Nicholson; Masayuki Noguchi; Giuseppe Pelosi; Claudia Poleri; Prudence A Russell; Jennifer Sauter; Erik Thunnissen; Ignacio Wistuba; Hui Yu; Murry W Wynes; Melania Pintilie; Yasushi Yatabe; Fred R Hirsch
Journal:  J Thorac Oncol       Date:  2018-05-22       Impact factor: 15.609

3.  Prospective multi-institutional evaluation of pathologist assessment of PD-L1 assays for patient selection in triple negative breast cancer.

Authors:  Emily S Reisenbichler; Gang Han; Andrew Bellizzi; Veerle Bossuyt; Jane Brock; Kimberly Cole; Oluwole Fadare; Omar Hameed; Krisztina Hanley; Beth T Harrison; M Gabriela Kuba; Amy Ly; Dylan Miller; Mirna Podoll; Anja C Roden; Kamaljeet Singh; Mary Ann Sanders; Shi Wei; Hannah Wen; Vasiliki Pelekanou; Vesal Yaghoobi; Fahad Ahmed; Lajos Pusztai; David L Rimm
Journal:  Mod Pathol       Date:  2020-04-16       Impact factor: 7.842

4.  Analytic Response Curves of Clinical Breast Cancer IHC Tests.

Authors:  Kodela Vani; Seshi R Sompuram; Anika K Schaedle; Anuradha Balasubramanian; Steven A Bogen
Journal:  J Histochem Cytochem       Date:  2017-01-01       Impact factor: 2.479

5.  Development and Validation of Measurement Traceability for In Situ Immunoassays.

Authors:  Emina E Torlakovic; Seshi R Sompuram; Kodela Vani; Lili Wang; Anika K Schaedle; Paul C DeRose; Steven A Bogen
Journal:  Clin Chem       Date:  2021-04-29       Impact factor: 8.327

Review 6.  PD-L1 as a biomarker of response to immune-checkpoint inhibitors.

Authors:  Deborah Blythe Doroshow; Sheena Bhalla; Mary Beth Beasley; Lynette M Sholl; Keith M Kerr; Sacha Gnjatic; Ignacio I Wistuba; David L Rimm; Ming Sound Tsao; Fred R Hirsch
Journal:  Nat Rev Clin Oncol       Date:  2021-02-12       Impact factor: 66.675

7.  Quantitative assessment of PD-L1 as an analyte in immunohistochemistry diagnostic assays using a standardized cell line tissue microarray.

Authors:  Sandra Martinez-Morilla; John McGuire; Patricia Gaule; Lauren Moore; Balazs Acs; Delphine Cougot; Allen M Gown; Hadi Yaziji; Wei-Lien Wang; Richard W Cartun; Jason L Hornick; Lynette M Sholl; Jingxin Qiu; Mari Mino-Kenudson; Eunhee S Yi; Mary Beth Beasley; Daniel T Merrick; Abiy B Ambaye; Zhong J Zhang; Jill Walker; David L Rimm
Journal:  Lab Invest       Date:  2019-08-13       Impact factor: 5.662

8.  PD-L1 Immunohistochemistry Assay Comparison in Atezolizumab plus nab-Paclitaxel-Treated Advanced Triple-Negative Breast Cancer.

Authors:  Hope S Rugo; Sherene Loi; Sylvia Adams; Peter Schmid; Andreas Schneeweiss; Carlos H Barrios; Hiroji Iwata; Véronique Diéras; Eric P Winer; Mark M Kockx; Dieter Peeters; Stephen Y Chui; Jennifer C Lin; Anh Nguyen Duc; Giuseppe Viale; Luciana Molinero; Leisha A Emens
Journal:  J Natl Cancer Inst       Date:  2021-06-07       Impact factor: 13.506

  8 in total
  1 in total

1.  Expression of CD274 mRNA Measured by qRT-PCR Correlates With PD-L1 Immunohistochemistry in Gastric and Urothelial Carcinoma.

Authors:  So Young Kang; You Jeong Heo; Ghee Young Kwon; Kyoung-Mee Kim
Journal:  Front Oncol       Date:  2022-04-27       Impact factor: 5.738

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.