| Literature DB >> 34505101 |
Leonardo Cavinatto1, Omar Khatib1, Alexander Martusiewicz1, Denise M Koueiter1, Brett P Wiater1, J Michael Wiater1.
Abstract
BACKGROUND: The purpose of this study was to compare the accuracy of anatomic reconstruction of three different humeral head designs after anatomic total shoulder arthroplasty.Entities:
Keywords: Anatomic total Shoulder arthroplasty; Humeral head reconstruction; Radiographic study; Stemless Elliptical humeral head; Stemless humeral head; Stemmed humeral head
Year: 2021 PMID: 34505101 PMCID: PMC8411049 DOI: 10.1016/j.jseint.2021.04.015
Source DB: PubMed Journal: JSES Int ISSN: 2666-6383
Figure 1Postoperative true AP radiographs for each implant desingn revealing the anatomic circle with its CoR (blue circle) using three preserved bone landmarks (blue dots) according to Youderian et al, and the postoperative implant circle and its CoR (yellow circle).
Figure 2Postoperative axillary radiographs for each implant design used to calculate the percentage of overhang of the each implant by dividing the anterior to posterior dimension of the implant “d” (red line) divided by the anterior to posterior extend of the prepared osteotomy “d” (yellow line).
Reconstruction index scoring system.
| Parameter | Score | Definition |
|---|---|---|
| Humeral head height | 0 | >6 mm difference between anatomic and prosthetic |
| 1 | 3-6 mm difference between anatomic and prosthetic | |
| 2 | 0-3 mm difference between anatomic and prosthetic | |
| Center of rotation | 0 | > 6 mm difference between anatomic and prosthetic |
| 1 | 3-6 mm difference between anatomic and prosthetic | |
| 2 | 0-3 mm difference between anatomic and prosthetic | |
| Overhang | 0 | >10% anterior-posterior overhang of the implant |
| 1 | 5-10% anterior-posterior overhang of the implant | |
| 2 | 0-5% anterior-posterior overhang of the implant |
Reconstruction scores, percent per implant type.
| Reconstruction score | Stemless elliptical | Stemless spherical | Stemmed |
|---|---|---|---|
| Poor score, 0-2 | 0.0% | 4.2% | 8.7% |
| Satisfactory score, 3-4 | 31.9% | 62.5% | 63.0% |
| Excellent score, 5-6 | 68.1% | 33.3% | 28.3% |
Mean measurements.
| Paramenter | Stemless elliptical | Stemless spherical | Stemmed spherical | |
|---|---|---|---|---|
| Change in humeral head height, mm | 1.28 ± 0.72 | 2.85 ± 1.68 | 1.76 ± 1.26 | <.001 |
| Shift in center of rotation, mm | 3.88 ± 1.75 | 3.34 ± 2.20 | 3.00 ± 1.56 | .060 |
| Percent overhang | 1.04 ± 0.02 | 1.08 ± 0.04 | 1.10 ± 0.04 | <.001 |
Score per parameter, percent based on implant type.
| Parameter | Score | Stemless elliptical | Stemless spherical | Stemmed spherical |
|---|---|---|---|---|
| Humeral head height | 0 | 0.0% | 4.2% | 2.2% |
| 1 | 2.1% | 37.5% | 10.9% | |
| 2 | 97.9% | 58.3% | 87.0% | |
| Center of rotation | 0 | 6.4% | 8.3% | 8.7% |
| 1 | 63.8% | 37.5% | 30.4% | |
| 2 | 29.8% | 54.2% | 60.9% | |
| Overhang | 0 | 0.0% | 29.2% | 52.2% |
| 1 | 34.0% | 41.7% | 30.4% | |
| 2 | 66.0% | 29.2% | 17.4% |