| Literature DB >> 34405488 |
Linda K Bawua1, Christine Miaskowski1, Xiao Hu2, George W Rodway3, Michele M Pelter1.
Abstract
BACKGROUND: Respiratory rate (RR) is one of the most important indicators of a patient's health. In critically ill patients, unrecognized changes in RR are associated with poorer outcomes. Visual assessment (VA), impedance pneumography (IP), and electrocardiographic-derived respiration (EDR) are the three most commonly used methods to assess RR. While VA and IP are widely used in hospitals, the EDR method has not been validated for use in hospitalized patients. Additionally, little is known about their accuracy compared with one another. The purpose of this systematic review was to compare the accuracy, strengths, and limitations of VA of RR to two methods that use physiologic data, namely IP and EDR.Entities:
Keywords: electrocardiography; hospitalized patients; impedance pneumography; respiratory rate; sensitivity/specificity; visual assessment
Mesh:
Year: 2021 PMID: 34405488 PMCID: PMC8411767 DOI: 10.1111/anec.12885
Source DB: PubMed Journal: Ann Noninvasive Electrocardiol ISSN: 1082-720X Impact factor: 1.468
FIGURE 1An Illustration of how electrocardiographic (ECG) limb leads I, II, and III are obtained using skin electrodes placed on the right arm (RA), left arm (LA) and left leg (LL). Impedance respiration is typically generated using one or two of these ECG leads using the bedside monitor. A single chest (C) electrode is shown that is routinely placed in the V1 position for in‐hospital arrhythmia monitoring and the right leg (RL) electrode, that is required to record lead V1. Lead V1 is not used for deriving respirations. Figure from Drew et al., PLoS One https://doi.org/10.1371/journal.pone.0110274.g003 (Drew et al., 2014)
FIGURE 2(a–c) Accurate (a), inaccurate (b), and motion artifact (c) respiratory waveforms using the impedance pneumography (IP) respiration method. (a) Normal respirations are generated from a 10 s IP waveform. Note the upward flag on the inspiratory waveform and the downward flag on the expiratory waveform, which are added by this particular manufacturer (GE Healthcare). (b) Inaccurate respiratory rate from a 10 s IP waveform recording. Note that occurrence of indistinguishable waveforms that are indicative of inspiration and expiration and the random flags throughout the tracing. (c) An illustration of a 20 s IP waveform during motion artifact, which resulted in an alarm for a respiratory rate of 55 breaths/min. Note that flags are present on the tracing that coincide with the oscillations of the IP waveform
FIGURE 3An illustration of a respiratory sinus arrhythmia (RSA) derived respiratory rate (a,b), which uses varying RR intervals (horizontal arrows) from QRS complexes on the electrocardiogram (ECG). Note that the circles and arrowheads of the horizontal arrows de‐note the QRS complexes. The inverse of the RR intervals is shown as vertical arrows (d), which are exaggerated for illustration. A heart rate is computed, which is used as amplitude knots for cubic spline interpolation to create the RSA‐derived respiration waveform (c). Reprinted with permission from the journal (Helfenbein et al., 2014)
FIGURE 4A diagrammatic representation of the literature search strategy using the PRISMA format (Moher et al., 2009)
Summary of the findings from studies that compared respiratory rates (RR) identified using visual assessment (VA), impedance pneumography (IP), and/or electrocardiographic‐derived (EDR) methods
| Author, year, country, purpose, setting, and study design | Sample characteristics | Study procedures and methods of data analysis | Main findings | Strengths and limitations |
|---|---|---|---|---|
| VA compared to IP | ||||
|
Author: Lovett et al. ( Country: USA Purpose: Measure the variability and accuracy of triage nurses' measurements of RR relative to criterion standard measurements and Evaluate the variability and accuracy of electronic measurements of RR recorded using a cardiac monitor equipped with transthoracic impedance (IP) Setting: urban teaching ED Design: Cross‐sectional |
Sample size: 159 consecutive patients who presented to the ED Age (years) 18–29 = 34.0% 30–39 = 22.6% 40–49 = 14.5% 50–59 = 8.8% 60–69 = 7.5% 70–79 = 5.7% 80–89 = 1.3% NR = 5.7% Mean Age = 39.41 Female = 50.9% Hispanic = 41.5% White = 46.5% |
Description of study procedures: Triage nurses' measurements of RR were recorded from the medical record Research Assistants (RAs) were trained in standardized methods to collect criterion standard measurements of RR. RAs observed respirations and auscultated RR at a single location for one minute. When auscultation could not be performed, observed RR was used in the analyses. RR using the IP method was captured at 60‐s intervals. Data analysis: Variability—was estimated by calculating the SD of each of the measures. Differences among the nurse, RA, and IP measures were evaluated using ANOVA. Sensitivity and specificity of triage nurses versus IP were cross‐tabulated measures against criterion standard measurements of respiratory values: Low = <12 breaths per minute Normal = 12–20 breaths per minute High = >20 breaths per minute Bland–Altman analyses were done that compared for—(a) triage nurses RR to criterion standard RR and (b) criterion standard RR to IP rates Agreement Bias 95% limits of agreement |
Variability for triage nurses' measurements of RR (3.3) was significantly lower than for IP (4.1) and criterion standard (4.8, Variability for IP measure was significantly lower than for criterion standard measure ( Accuracy of detecting bradypnea and tachypnea—neither triage nurses nor IP measures of RR were accurate in detecting bradypnea or tachypnea Bradypnea (<12 breaths/min) Nurse versus criterion measure
Sensitivity = 0.00 (0.00–0.35) Specificity = 1.00 (0.97–1.00) IP versus criterion measure
Sensitivity = 0.25 (0.07–0.59) Specificity = 0.98 (0.94–0.99) Tachypnea (>20 breaths/min) Nurse versus criterion measure
Sensitivity = 0.38 (0.25–0.53) Specificity = 0.84 (0.75–0.90) IP versus criterion measure
Sensitivity = 0.40 (0.28–0.55) Specificity = 0.86 (0.78–0.92) Agreement between triage nurses and criterion measure of RR was poor (95% limits of agreement −8.6 to 9.5) Agreement between IP and criterion measure of RR was poor (95% limits of agreement −9.9 to 7.5) Systematic bias was small for triage nurses' measurements of RR (+0.0) and electronic measurements of RR (−1.2) |
Strengths
Data collected in an ED during triage The criterion reference standard used for comparison Use of Bland–Altman analyses Limitations
The majority of the patients were less than 39 years of age
Triage nurses were aware that their assessments of RR were being collected
Criterion measure of RR was obtained after the triage visit, not simultaneously with triage nurses' assessment of RR
No inter‐rater reliability estimates were done with the RAs |
|
Author: Chand et al. ( Country: India Purpose: Examine differences between VA and electronic (IP) measurements of vital signs in cardiac patients Setting: Advanced Cardiac Centre ICU Design: Comparative study |
Sample size: 50 patients admitted in CTVS‐ICU and CCU CTVS‐ICU = 21 (42%) CCU = 29 (58%) Mean age (Years) =55.9 Females = 49.25 (range 25–58) Females = 16% Ethnicity = NR |
Description of study procedures: VA—By floor RNs IP—By the cardiac monitor Four measurements of temperature, pulse, respiration, and blood pressure were recorded at 30‐min intervals, consecutively. The measurement of each vital sign was done simultaneously. Data analysis: Paired The coefficient of variation was calculated to quantify the variation between the VA and IP measures |
A total of 200 measurements were done using each method The mean difference in RR between the VA and IP methods was not significant (i.e., 0.015 (±1.16), The coefficient of variation between the VA (26.25%) and IP (25.48%) was similar |
Strengths
Measurements made simultaneously Limitations
Purposive sampling
Type of physiologic monitor not reported
Unclear if nurses were blinded to values obtained with the IP methods
Small sample size
Bland–Altman analyses were not performed |
|
Author: Granholm et al. ( Country: Denmark Purpose: Evaluate the agreement between RR rates done using three methods (i.e., standardized approach, VA by ward staff, IP) Setting: Medical unit Design: Prospective, observational study |
Sample size: 50 patients admitted to an acute medical unit Median age (years) =71.5 Female = 54% Ethnicity = NR |
Description of study procedures: VA—Ward staff performed all assessments as usual. Data obtained from medical record IP—Sensium Vitals wireless patch measures RR, heart rate, and axillary temperature every 2 min Standardized approach—Trained researchers counted the patient's RR over 60 s. Patients were instructed to lie still and refrain from talking Data analysis: Bland–Altman analysis used to evaluate the agreement between the methods with 95% LOA and 95% CI |
Agreement between standardized VA by researcher versus IP
Mean difference was 0.3 b/m (95% CI −1.4 to 2.0 b/m) Lower and upper 95% LOAs were −11.5 b/m (95% CI −14.5 to −8.6 b/m) and 12.1 b/m (95% CI 9.2 to 15.1 b/m) respectively Large RR differences (>10 b/m) were found in three outliers (i.e., one obese patient with respiratory disease; one elderly patient with respiratory disease, atrial fibrillation, and prior cardiac surgery; one slim young patient with a non‐respiratory‐related infection)
The mean difference after removing three outliers was −0.1 b/m (95% CI −0.7 to 0.5 b/m). Without outliers' differences were normally distributed Agreement between VA by ward staff versus IP
Mean difference was 1.7 b/m (95% CI −0.5 to 3.9 b/m) Lower and upper 95% LOAs were −13.3 b/m 95% CI −17.2 to −9.5 b/m and 16.8 b/m (95% CI 13.0 to 20.6 b/m), respectively
RR by ward staff was not normally distributed, with digit preferences of 16, 18, and 20 b/m |
Strengths
One trained researcher recorded the standardized approach
The single paired measurement used for each patient minimized bias caused by within‐subject correlations Limitations
No repeated measurements
RR done by ward staff were obtained from the electronic health record, which could affect comparison with IP (i.e., inaccurate times recorded)
Small sample size |
| VA compared to EDR | ||||
|
Author: Kellett et al. ( Country: Ireland Purpose: Evaluate for the association between VA and EDR measured RR and their relationships to in‐hospital mortality Setting: Acute medical unit in a small rural hospital Design: Descriptive, correlational |
Sample size: 377 acutely ill medical patients Mean age (years) – 68.3 ± 16.8 Alive = 67.9 (±17.0) Dead = 77.1 (±9.2) Sex = NR Ethnicity = NR |
Description of study procedures: VA of RR was obtained by one of eight nurses on the patient's admission to the unit. Nurses were not given any instructions on how to measure or record RR. EDR: RR was obtained using a BT16/Piezoelectric belt for 5 min after admission. Data were transmitted to a separate computer system for subsequent analyses. Data analysis: Paired Correlation coefficients were calculated for VA versus EDR measures of RR. Bland–Altman plots were done to evaluate the limits of agreement between the VA and IP measures of RR |
The mean RR measured by VA (20.9 (±4.8) breaths/min) was significantly different from that obtained by EDR (19.9 (±4.5) breaths/min), The correlation coefficient between VA and EDR was 0.50. Visual inspection of the scatter plots illustrated that RR obtained using VA clustered around rates of 18, 20, and 22 breaths/min. The RR rates obtained using EDR were more variable. Bland–Altman plots revealed that the 95% LOA between VA and EDR for RR were −8.2 and 10.3 breaths/min |
Strengths
Relatively large sample size Limitation
Demographic and clinical characteristics of the sample (e.g., acuity level, use of medications) were not reported Only eight nurses participated in this study, and their characteristics were not reported Lack of standardization in the VA or RR Bland–Altman plots not included in the paper |
Abbreviations: b/m, breaths per minute; CTVS‐ICU, cardiothoracic and vascular surgery‐intensive care unit; CCU, critical care unit; CI, confidence interval; CSM, criterion standard measurement; EDR, electrocardiographic‐derived respiration; IP, impedance pneumography; ED, emergency department; LOA, limits of agreement; NR, not reported; PACU, post anesthesia care unit; VA, visual assessment; RN, registered nurse; RR, respiratory rate; SD, standard deviation.
Comparison of the strengths and limitations of the visual assessment (VA), impedance pneumography (IP), and ECG‐derived respiration (EDR) methods for assessment of respiratory rate
| Methods | Strengths | Limitations |
|---|---|---|
| VISUAL |
Traditional method to assess RR Easy and safe to perform Breathing characteristics (e.g., depth, accessory muscles, skin color) can be assessed |
Time‐consuming for clinicians Numerous omissions and guessed measurements (Cooper et al., Low precision and/or variability because respiratory rate is often counted for 30 s, and then, the value is multiplied by two to get the number of breaths/minute VA is a snapshot of a patient's RR at prescribed intervals (e.g., every 30 min). Acute changes and early identification of patient deterioration can be missed |
| IP |
Simpler, less time‐consuming than VA Safe to use Continuous measurement of RR Coherence analysis concluded that IP is more reliable than EDR (Ernst et al., |
Studies found that the IP method was prone to erratic artifacts, false‐positive readings and was sensitive to motion and cardiac artifacts (Drew et al., A device's internal impedance, such as cables and wires, can be a source of measurement error (Landon, IP can generate false positives from movement and interruptions by the examinee and affect the readings and values (Krapohl & Shaw, IP method is influenced by behaviors that occur naturally (e.g., talking, coughing) (Krapohl & Shaw, IP is predisposed to signal degeneration with body position changes because the thoracic signal depends on posture, making it difficult to evaluate tidal volume (Landon, |
| EDR |
The EDR algorithm can be added to existing ECG to extract respiratory signals from the ECG signal without new transducers, devices, or accessories required for monitoring (Charlton et al., Continuous monitoring and non‐invasive (Charlton et al., The sensitivity and specificity of the EDR algorithm to identify RR were high (99%/97%) in cardiac patients compared with other methods (Babaeizadeh et al., Alterations in the RR are easily detected |
RSA aspect weakens with aging, which may lead to inaccurate measurements in older individuals. Patient movement and noise can cause artifacts and lead to inaccurate values. Method lacks validation in the hospital setting. EDR measurement can be affected by the natural decline in RSA, as well as arrhythmias (e.g., atrial fibrillation) and the effects of medications that affect heart rate and rhythm |
FIGURE 5False apnea alarm in an intensive care unit patient measured using the impedance method. The respiratory waveform (bottom waveform labeled “RESP”) is essentially a flat line. Therefore, respiratory rated calculated using the impedance method alarmed for apnea. The monitor default setting for apnea is cessation of breathing for >20 s. However, this patient was not in acute respiratory distress at the time of this alarm. Note at the top of the alarm tracing is an erroneous respiratory rate (RR) of 6 breaths/min, yet the oxygen saturation measure from the Sp02 probe is 95%