A radiation quality correction factor k for well-type ionization chambers for the measurement of the reference air kerma rate of (60)Co HDR brachytherapy sources.

Andreas Schüller1, Markus Meier1, Hans-Joachim Selbach1, Ulrike Ankerhold1.   

Abstract

PURPOSE: The aim of this study was to investigate whether a chamber-type-specific radiation quality correction factor kQ can be determined in order to measure the reference air kerma rate of (60)Co high-dose-rate (HDR) brachytherapy sources with acceptable uncertainty by means of a well-type ionization chamber calibrated for (192)Ir HDR sources.
METHODS: The calibration coefficients of 35 well-type ionization chambers of two different chamber types for radiation fields of (60)Co and (192)Ir HDR brachytherapy sources were determined experimentally. A radiation quality correction factor kQ was determined as the ratio of the calibration coefficients for (60)Co and (192)Ir. The dependence on chamber-to-chamber variations, source-to-source variations, and source strength was investigated.
RESULTS: For the PTW Tx33004 (Nucletron source dosimetry system (SDS)) well-type chamber, the type-specific radiation quality correction factor kQ is 1.19. Note that this value is valid for chambers with the serial number, SN ≥ 315 (Nucletron SDS SN ≥ 548) onward only. For the Standard Imaging HDR 1000 Plus well-type chambers, the type-specific correction factor kQ is 1.05. Both kQ values are independent of the source strengths in the complete clinically relevant range. The relative expanded uncertainty (k = 2) of kQ is UkQ = 2.1% for both chamber types.
CONCLUSIONS: The calibration coefficient of a well-type chamber for radiation fields of (60)Co HDR brachytherapy sources can be calculated from a given calibration coefficient for (192)Ir radiation by using a chamber-type-specific radiation quality correction factor kQ. However, the uncertainty of a (60)Co calibration coefficient calculated via kQ is at least twice as large as that for a direct calibration with a (60)Co source.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26133626     DOI: 10.1118/1.4922684

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Impacts on air dose rates after the Fukushima accident over the North Pacific from 19 March 2011 to 2 September 2015.

Authors:  Kuo-Ying Wang; Philippe Nedelec; Hannah Clark; Neil Harris; Mizuo Kajino; Yasuhito Igarashi
Journal:  PLoS One       Date:  2022-08-24       Impact factor: 3.752

2.  Source strength determination in iridium-192 and cobalt-60 brachytherapy: A European survey on the level of agreement between clinical measurements and manufacturer certificates.

Authors:  Javier Vijande; Åsa Carlsson Tedgren; Facundo Ballester; Dimos Baltas; Panagiotis Papagiannis; Mark J Rivard; Frank-André Siebert; Larry De Werd; Jose Perez-Calatayud
Journal:  Phys Imaging Radiat Oncol       Date:  2021-07-30

3.  Dosimetric evaluation of newly developed well-type ionization chamber for use in the calibration of brachytherapy sources.

Authors:  Sathiyan Saminathan; Henry Finlay Godson; Retna Ponmalar; Ravikumar Manickam; James Mazarello
Journal:  J Med Phys       Date:  2016 Oct-Dec
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.