Balazs Feher1,2, Florian Frommlet3, Reinhard Gruber1,4,5, Lena Hirtler6, Christian Ulm2, Ulrike Kuchler2.
Abstract
OBJECTIVES: Resonance frequency analysis (RFA) is used to monitor implant stability. Its output, the Implant Stability Quotient (ISQ), supposedly correlates with insertion torque, a common measurement of primary stability. However, the reliability of RFA in condensed bone remains unclear.Entities:
Keywords: bone condensation; edentulous jaw; immediate implant placement; implant stability; resonance frequency analysis
Mesh:
Substances:
Year: 2021 PMID: 34358360 PMCID: PMC9292279 DOI: 10.1111/clr.13817
Source DB: PubMed Journal: Clin Oral Implants Res ISSN: 0905-7161 Impact factor: 5.021
FIGURE 1Region of interest definition. A Top‐down view. The inner yellow circle with radius r is the region of interest is an expansion of the implant site by 𝛿 = 0.5 mm. The outer yellow circle with radius R is the outer bounds of the region of interest. The white area between the two yellow circles represents a slice of the volumetric region of interest. B Lateral reconstruction view. The implant length is h = 10 mm. The inner yellow line is the of the sites with radius r in the coronal plane. The outer yellow line represents the reconstruction of the outer bounds of the region of interest after the expansion by 𝛿 = 0.5 mm. The white area between the two yellow lines is the reconstruction of a slice of the volumetric region of interest
Correlation analysis
| Edentulous jaws | Fresh extraction sockets | |||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| Peak insertion torque—Implant Stability Quotient | ||||||
| Uncondensed bone | 13 | .63 | .02 | 13 | .78 | < .01 |
| Condensed bone | 13 | .82 | < .01 | 16 | .15 | .58 |
| Combined | 26 | .78 | < .0001 | 29 | .61 | < .001 |
| Preoperative BV/TV—peak insertion torque | ||||||
| Uncondensed bone | 13 | .82 | < .01 | 7 | .61 | .17 |
| Condensed bone | 13 | .93 | < .0001 | 13 | –.33 | .27 |
| Combined | 26 | .90 | < .0001 | 20 | .33 | .15 |
| Preoperative BV/TV—Implant Stability Quotient | ||||||
| Uncondensed bone | 13 | .51 | .08 | 7 | .71 | .09 |
| Condensed bone | 13 | .65 | .02 | 13 | –.26 | .38 |
| Combined | 26 | .64 | < .001 | 20 | .38 | .09 |
All r and p values using Spearman's rank correlation coefficient.
Abbreviations: BV, bone volume and TV, total volume of the region of interest.
FIGURE 2Correlations in edentulous jaws. Blank circles show sites without, filled circles show sites with bone condensing. All r and p values using Spearman's rank correlation coefficient
Split‐mouth comparisons in edentulous jaws
| Uncondensed bone | Condensed bone |
| |
|---|---|---|---|
| Clinical parameters | |||
| Peak insertion torque, Ncm | 28 (22–37) | 26 (18–46) | .67 |
| Implant Stability Quotient | 74 (71–77) | 74 (63–79) | .67 |
| Radiographic parameters | |||
| BV/TV, preoperative CT | 0.49 (0.38–0.66) | 0.67 (0.33–0.74) | .74 |
| BV/TV, postoperative CT | 0.41 (0.35–0.50) | 0.39 (0.27–0.49) | .11 |
| BV/TV, postoperative µCT | 0.88 (0.77–0.93) | 0.87 (0.76–0.95) | .95 |
Medians and interquartile ranges. P‐values using Wilcoxon matched‐pairs signed‐rank test.
Abbreviations: BV, bone volume; CT, computed tomography; TV, total volume of the region of interest; and µCT, micro‐computed tomography.
FIGURE 3Preoperative relative bone volume and its absolute change in edentulous jaws. Blank circles show sites without, filled circles show sites with bone condensing. All r and p values using Spearman's rank correlation coefficient
FIGURE 4Correlations in fresh extraction sockets. Blank circles show sites without, filled circles show sites with bone condensing. All r and p values using Spearman's rank correlation coefficient
FIGURE 5Preoperative relative bone volume and its absolute change in fresh extraction sockets. Blank circles show sites without, filled circles show sites with bone condensing. All r and p values using Spearman's rank correlation coefficient
FIGURE 6Joint analysis of edentulous jaws and fresh extraction sockets. Blank squares show fresh extraction sockets, filled squares show sites in empty jaws