| Literature DB >> 34349370 |
Emilie Acx1, Els Van Caelenberg2, Luc De Baerdemaeker3, Marc Coppens3.
Abstract
BACKGROUND AND AIMS: The Laryngeal Mask Airway (LMA) Protector™ is one of the latest introduced supraglottic airway devices. It provides access and functional separation of the respiratory and digestive tracts. Compared to the LMA Supreme™, it has two digestive ports, one to provide suction in the pharyngeal region and one for gastric tube insertion. High oropharyngeal leak pressure is a marker for safe ventilation when using LMA devices. We hypothesized that oropharyngeal leak pressure of the LMA Protector™ is 5 cm H2O higher than the oropharyngeal leak pressure of the LMA Supreme™ at various cuff volumes. Secondary outcome measures were ease of insertion of both masks, fiberoptic confirmation of correct positioning, failures of insertion, presence of blood staining, sore throat, presence of air leak and insertion time.Entities:
Keywords: Laryngeal mask; laryngeal mask airway protector; supraglottic airway device
Year: 2021 PMID: 34349370 PMCID: PMC8289652 DOI: 10.4103/joacp.JOACP_416_19
Source DB: PubMed Journal: J Anaesthesiol Clin Pharmacol ISSN: 0970-9185
Demographics
| LMA Protector | LMA Supreme | ||
|---|---|---|---|
| Age (years) | 42 (15.3) | 41 (14.6) | 0.8 |
| Height (cm) | 171 (10.8) | 170 (9.9) | 0.9 |
| Weight (kg) | 73 (13.6) | 73 (14.3) | 0.7 |
Data are presented as mean (standard deviation)
Figure 1Oropharyngeal airway leak pressures at different cuff volumes and cuff pressure of 65 cm H2O
Oropharyngeal leak pressures at different cuff volumes and at cuff pressure of 65 cm H2O
| Cuff volumes | Cuff pressure | |||||
|---|---|---|---|---|---|---|
| 0 ml | 10 ml | 20 ml | 30 ml | 40 ml | 65 cm H2O | |
| LMA Supreme | 10 | 13 | 18 | 20 | 22 | 21 |
| (5) CI (9-12) | IQR (8) | (6) CI (17-20) | (6) CI (18-22) | (7) CI (20-24) | (6) CI (19-23) | |
| LMA Protector | 14 | 21 | 28.5 | 32 | 33.5 | 30 |
| IQR (12) | IQR (13) | IQR (19) | IQR (18) | IQR (17) | IQR (18) | |
| <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | |
Pressures are presented in cm H
Secondary outcome measures
| LMA Supreme ( | LMA Protector ( | ||
|---|---|---|---|
| Insertion success, | 38 (95%) | 36 (90%) | 0.675 |
| Second attempt needed, | 2 (5%) | 4 (10%) | 0.675 |
| Third attempt needed, | 1 (2.5%) | 0 | 1.000 |
| Etiology of failure, | |||
| Failed passage into pharynx | 1 (2.5%) | 2 (5%) | 1.000 |
| Malposition (air leak) | 1 (2.5%) | 2 (5%) | 1.000 |
| Failed ventilation | 1 (2.5%) | 0 | 1.000 |
| Visible blood staining, | 4 (10%) | 8 (20%) | 0.348 |
| Sore throat (VAS) | 0.13 (0.56) | 0.18 (0.55) | 0.689 |
| Presence of air leak, | |||
| Listening at the mouth | 4 (10%) | 1 (2.5%) | 0.359 |
| Gastric auscultation | 0 | 0 | 0.494 |
| Laryngeal auscultation | 2 (5%) | 0 | |
| At drain tube | 0 | 0 | |
| End-tidal >45 mmHg | 0 | 0 | |
| Insertion time (s) | 19 IQR (16,22) | 29IQR (23,35) | 0.0001 |