Abstract
OBJECTIVE: To explore the quantitative adjustment of ablation index (AI) under different baseline impedance to achieve similar lesion dimensions.Entities:
Year: 2021 PMID: 34336272 PMCID: PMC8289601 DOI: 10.1155/2021/5574125
Source DB: PubMed Journal: Cardiol Res Pract ISSN: 2090-0597 Impact factor: 1.866
Figure 1Schematic diagram of the in vitro ablation model.
Figure 2The lesion dimensions in increased baseline impedance and similar AI.
The relationship among different basic impedance, power, and the ranges of ablation damage when the current and pressure were relatively constant.
| Impedance (Ω) | Power (W) | AI | Average pressure (g) | Depths (mm) | Widths (mm) |
|---|---|---|---|---|---|
| 99.0 ± 1.0 | 25 | 392.6 ± 6.9 | 10.0 ± 1.0 | 4.13 ± 0.28 | 6.40 ± 0.20 |
| 119.8 ± 1.5 | 30 | 452.2 ± 7.8 | 11.2 ± 0.8 | 4.14 ± 0.24 | 6.57 ± 0.90 |
| 140.4 ± 0.9 | 35 | 486.2 ± 12.6 | 11.0 ± 1.0 | 4.41 ± 0.23 | 6.68 ± 0.30 |
| 159.2 ± 1.6 | 40 | 519.6 ± 10.0 | 10.6 ± 0.5 | 4.38 ± 0.32 | 6.01 ± 0.44 |
| 180.0 ± 0.0 | 45 | 559.6 ± 11.3 | 11.0 ± 1.0 | 4.37 ± 0.28 | 6.17 ± 0.23 |
AI = Ablation index.
Figure 3The relationship between the AI and the baseline impedance in the similar ablation depths.
Figure 4The lesion depths, baseline impedance, and AI values in formula-guided ablation in the common AI groups.
Figure 5The variation tendency of lesion dimensions in the common AI groups.
The target AI, actual AI, and damage ranges obtained in different impedance under the guidance of empirical formula.
| Impedance (Ω) | Target AI | Actual AI | Depths (mm) | Widths (mm) |
|---|---|---|---|---|
| 120.0 ± 1.1 | 400 | 402.4 ± 1.6 | 3.70 ± 0.35 | 6.30 ± 0.33 |
| 140.0 ± 0.5 | 433 | 433.6 ± 2.4 | 3.81 ± 0.43 | 6.07 ± 0.30 |
| 159.9 ± 0.4 | 466 | 463.4 ± 2.6 | 3.85 ± 0.35 | 6.11 ± 0.38 |
| 179.9 ± 0.8 | 499 | 497.6 ± 1.7 | 3.87 ± 0.20 | 6.18 ± 0.38 |
| 200.4 ± 0.7 | 531 | 533.9 ± 1.6 | 3.82 ± 0.22 | 6.09 ± 0.55 |
Figure 6Baseline impedance and AI values in the high AI groups and the corresponding lesions created under these settings.
The lesion depths in the high AI groups and the P values between the experimental groups and control group.
| 550/120 Ω | 608/140 Ω | 666/160 Ω | 723/180 Ω | 781/200 Ω | |
|---|---|---|---|---|---|
| Depths (mm) | 5.778 ± 0.282 | 5.766 ± 0.205 | 5.778 ± 0.452 | 5.472 ± 0.395 | 5.23 ± 0.325 |
|
| 0.941 | 1 | 0.196 | 0.022 |
The lesion widths in the high AI groups and the P values between the experimental groups and control group.
| 550/120 Ω | 608/140 Ω | 666/160 Ω | 723/180 Ω | 781/200 Ω | |
|---|---|---|---|---|---|
| Widths (mm) | 10.002 ± 0.401 | 9.308 ± 0.506 | 9.486 ± 0.581 | 8.198 ± 0.470 | 8.436 ± 0.674 |
|
| 0.053 | 0.141 | 0.000 | 0.002 |