Sarah Bouari1, Özgür Eryigit1, Ron W F de Bruin1, Jan N M IJzermans1, Robert C Minnee1.
Abstract
We present an updated overview of the literature comparing normothermic with hypothermic machine perfusion in porcine kidneys. We conducted a systematic literature review in Embase, Medline Epub (Ovid), Cochrane Central, Web of Science, and Google Scholar on studies comparing normothermic (NMP) to hypothermic machine perfusion (HMP) in porcine kidneys. A meta-analysis was judged inappropriate because of heterogeneity in study design and perfusion methods. The quality of evidence of each included study was assessed. We included 8 studies. One out of 5 studies reported a significant difference in peak renal blood flow in favor of NMP. Oxygen consumption was significantly higher in NMP kidneys in 2 out of 5 studies. Peak creatinine clearance in NMP was significantly higher than that in HMP in 1 out of 6 studies. Two out of 4 studies reported a higher degree of epithelial vacuolation in kidneys receiving NMP over HMP. None of the studies found a significant difference between NMP and HMP in peak serum creatinine or graft survival after autotransplantation. The results need to be interpreted with caution in view of the diversity in perfusion protocols, the low quality of evidence, and the limited sample sizes.Entities:
Keywords: graft assessment; graft viability; hypothermic machine perfusion; normothermic machine perfusion; organ preservation
Mesh:
Year: 2021 PMID: 34309868 PMCID: PMC8596691 DOI: 10.1111/aor.14039
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 3.094
Study characteristics
| Study | Year | Weight range of animals (kg) | Study groups (duration) | No. of animals | Perfusion machine brand | Perfusate | Warm ischemia time (min) | Oxygenated | Autotransplantation of graft |
|---|---|---|---|---|---|---|---|---|---|
| Bagul et al | 2008 | 60‐70 | HMP (18 hours) | 6 | LP | Kidney perfusion solution (KPS‐1) | 10 | No | No |
| HMP (16 hours) → NMP (2 hours) | 6 | IOPS | NMP with whole autologous blood | 10 | Yes | ||||
| Blum et al | 2017 | Not mentioned in the study | HMP (8 hours) | 5 | n/a | Belzer's Machine Perfusion Solution | 45 | No | No |
| NMP (8 hours) | 5 | n/a | Erythrocyte‐based solution | 45 | Yes | ||||
| Darius et al | 2019 | 40 | SCS (20 hours) → HMP (2 hours) | 7 | LP | Kidney perfusion solution (KPS‐1) | 30 | Yes | Yes |
| SCS (20 hours) → NMP (2 hours) | 6 | LP | Autologous leukocyte‐depleted blood | 30 | Yes | ||||
| Hosgood et al | 2011 | 37.0‐43.8 | HMP (22 hours) | 6 | LP | Kidney perfusion solution (KPS‐1) | 32 | No | Yes |
| HMP (20 hours) → NMP (2 hours) | 6 | LP | NMP with autologous leukocyte‐depleted blood | 33 | Yes | ||||
| Hoyer et al | 2014 | 30 | HMP (7 hours) | 5 | LP + HFox | Custodiol‐N solution including 5 g/dL of dextran 40 | 30 | Yes | No |
| SNMP (7 hours) (20℃) | 5 | LP + HFox | Custodiol‐N solution including 5 g/dL of dextran 40 | 30 | Yes | ||||
| Metcalfe et al | 2002 | 80‐100 | SCS (2 hours) → HMP (16 hours) | 6 | RM3 system (Waters, Massachusettss, USA) | Belzer's Machine Perfusion Solution | 8 | No | No |
| SCS (2 h) → NMP (16 hours) | 6 | POPS | Whole autologous blood | 8 | Yes | ||||
| Schopp et al | 2015 | 25‐30 | HMP (18 hours) | 6 | n/a | Custodiol‐N solution including 5 g/dL of dextran 40 | n/a | No | No |
| SCS (18 hours) → COR (3 hours) | 6 | n/a | Custodiol‐N solution including 5 g/dL of dextran 40 | n/a | Yes | ||||
| Urbanellis et al | 2020 | 30 | HMP (16 hours) | 5 | LP | Belzer's Machine Perfusion Solution | 30 | No | Yes |
| NMP (16 hours) | 5 | n/a | Leukocyte‐depleted blood | 30 | Yes |
Abbreviations: COR, controlled oxygenated rewarming; HMP, hypothermic machine perfusion; (s)NMP, (sub)normothermic machine perfusion; SCS, static cold storage.
LifePort hypothermic preservation machine (Organ Recovery Systems, Des Plaines, Illinois, USA).
Isolated organ perfusion system (Medtronic 550 Bio‐console system; Medtronic, Watford, UK).
LifePort hypothermic preservation machine (Organ Recovery Systems, Des Plaines, Illinois, USA) with hollow fiber oxygenator (Minimax; Medtronics; Minneapolis, Minnesota, USA).
POPS (pulsatile organ perfusion system) (Transmedics, Boston, Massachusetts, USA).
Overview of study outcomes
| Study | Year | Perfusion method | Peak renal blood flow (mL/min) | Oxygen consumption (mL/min/100 g) | Intrarenal resistance (mm Hg/mL/min) | Peak serum creatinine (μmol/L) | Peak creatinine clearance (mL/min) | Graft survival after autotransplantation (%) | Epithelial vacuolation (grade) | Tubular dilatation (grade) | Epithelial shedding (grade) |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Bagul et al | 2008 | NMP | 90 | 40 | 0.4 | 1756 | 1.8 | n/a | 1.5 | No significant difference | No significant difference |
| HMP | Not reported | 35.1 | Not reported | 2156 | 2.5 | n/a | 0.8 | ||||
| Blum et al | 2017 | NMP | 4.4 | 0.18 | 6.5 | n/a | 1.7 [0.9‐2.3] | 1.6 [0.1‐2.3] | 1.1 [0.6‐1.9] | ||
| HMP | n/a | 1 | 9.3 | n/a | 0 [0‐0.5] | 0.9 [0.1‐2.2] | 1.3 [1.2‐1.9] | ||||
| Darius et al | 2019 | NMP | 73.35 | 0.96 | Not reported | ||||||
| HMP | 49.03 | 0.38 | Not reported | ||||||||
| Hosgood et al | 2011 | NMP | 55.33 | 28.3 | 0.45 | 1553 | 2.3 | 66.67 | 1.1 | 1.9 | |
| HMP | 55.33 | n/a | 0.46 | 1736 | Not reported | 83.33 | 1.2 | 2.2 | |||
| Hoyer et al | 2014 | sNMP | 212.82 | 3.67 | 0.46 | 18.18 | n/a | [2‐3] | [2‐2.5] | Not observed | |
| HMP | 176.37 | 3.5 | 1.08 | 12.29 | n/a | [2‐3] | [2‐3] | Not observed | |||
| Metcalfe et al | 2002 | NMP | No significant difference | No significant difference | n/a | ||||||
| HMP | No significant difference | No significant difference | n/a | ||||||||
| Schopp et al | 2015 | COR | 6.28 | 1.13 | 16.95 | n/a | |||||
| HMP | 3.93 | 1.45 | 9.58 | n/a | |||||||
| Urbanellis et al | 2020 | NMP | 215 | 0.45 | 100 | [1‐1.5] | |||||
| HMP | 53.6 | 0.45 | 100 | [1‐1.5] |
Abbreviations: COR, controlled oxygenated rewarming; HMP, hypothermic machine perfusion; (s)NMP, (sub)normothermic machine perfusion; SCS, static cold storage.
Data originally published as a graph; extracted using DataThief III.
in kPa.mL/min/g.
Exact measurements not reported; no significant difference between groups.
No statistical analysis performed.
A 5‐point scale was used instead of a 3‐point scale.
Significant difference.