Forecasting of the composite dose for organs at risk and solid targets with random movements during different image-guided scenarios of the photon radiation therapy. Solution for the Varian therapeutic line.

Adam Ryczkowski1,2, Tomasz Piotrowski1,2.   

Abstract

BACKGROUND: This study aims to develop a useful tool for robust plan analysis which includes the effects of soft tissue deformations on simulated dose distributions. The solution was benchmarked in the light of the commercial method implemented in EclipseTM treatment planning system (TPS).
MATERIALS AND METHODS: Study was carried out on data of one patient with prostate-restricted cancer. The workflow of the procedure developed focused on three executive elements: in-house script to create a set of artificial CT images and for movement simulation of the CT V; the VelocityTM software for the calculations of the deformation matrixes and, then, to generate deformed CT sets; the EclipseTM TPS for dose re-calculations and analysis. Two scenarios were examined - first when the recalculation was done for the original geometry and second, when the isocentre from the original plan geometry was moved according to the movement of the CT V. The dose distributions were analysed on dose volume histograms (DVHs) in the light of the results obtained from the method implemented in the EclipseTM TPS.
RESULTS: The DVHs from our methods are more informative than the DVH from commercially implemented tools. For the first scenario, the highest impact on dose uncertainty has boundary positions of the CT V to the CT V-PTV margin. Using the second scenario, it is the relation of the CT V position to the whole body that has the highest effect on dose uncertainty.
CONCLUSION: Our method enables a more accurate analysis of the treatment plan robustness than the method currently implemented in EclipseTM TPS.
© 2021 Greater Poland Cancer Centre.

Entities:  

Keywords:  prostate radiotherapy; robust evaluation; robust plan analysis; robustness

Year:  2021        PMID: 34277106      PMCID: PMC8281900          DOI: 10.5603/RPOR.a2021.0051

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


Introduction

The most used radiation therapy treatments based on treatment plans, in which the delineation of the target implicates the necessary definition of the margin around the clinical target volume (CTV). CTV with this specific, geometrical, buffer zone is defined as planning target volume (PTV). The size of the CTV-PTV margin depends on the internal movement of the tumour and the errors caused by the patient set-up and the dose delivery methods or specific dose distributions obtained during treatment. Correctly applied CTV-PTV margin should provide the delivery of the planned dose to the CTV during the treatment [1-4]. Taking into account the accuracy of dose delivery - the size of the CTV-PTV margin depends on the radiation therapy method selected in the treatment and on the image-guidance scenarios used during the treatment. Numerous investigators show that increasing the frequency of image guidance (IG) allows for smaller CTV-PTV margins [5-8], or that the magnitude of margins depends on the modality used for IG [9, 10]. The safety of the PTV-based treatments makes two main assumptions. First is that moderate changes in patient position or anatomy do not affect the dose distribution and the second that the effects of the interplay between intra-fractional motion and dynamic fluence patterns are negligible [11]. Taking into account 5% as the acceptable dosimetric accuracy between planned and delivered doses [12] a lot of studies confirmed that the safety PTV-based concept is suitable for photon radiation therapy [13-18]. Nevertheless, assuming the correctness of the dose delivery at the accelerator level, the distribution of the dose uncertainties could differ for different IG scenarios used during therapy. Therefore, the analysis of the plan robustness on the potential movement of the CTV and the deformations of the tissues caused from this movement is one of the crucial methods to obtain information about possible dose distributions during correctly performed radiation therapy. There are two methods of robust plan analysis in the EclipseTM treatment planning system (TPS) (Varian Medical Systems, Palo Alto, CA, USA) that are available in the “Plan Uncertainty Parameters” dialogue box (PUP). First of them estimates how a change in patient set-up may affect the dose distribution, the other one forecasts the effect on the dose distribution if the target moves to other structures. While the first method simulates the dose distributions during the radiation therapy managed and controlled by classical IG, such as no action level (NAL) protocol, the other one tries to simulate the dose distributions obtained during daily IG procedures. Unfortunately, these scenarios do not include in the final budget of the dose distributions the effect caused by deformations of the soft tissues that are accompanying the change of position of the CTV. This study aimed to present the solution, based mainly on Varian commercial products, for robust plan analysis, which includes the effects of the soft tissue deformations on the simulated dose distributions.

Materials and methods

Study was carried out on data of one patient with prostate-restricted cancer. The workflow of the developed procedure focused on three executive elements: in-house script for movement simulation of the CTV written in the Python programming language; the VelocityTM software (Varian Medical Systems, Palo Alto, CA, USA) for calculations of the deformation matrixes; the EclipseTM TPS for dose re-calculations and analysis. Figure 1 shows the general workflow of the authors’ solution.
Figure 1

General workflow of the authors’ solution. aCT — artificial set of CT images; dCT — deformed set of CT images; 1st method — re-calculation is done for the original geometry (simulation of IG based on the bony anatomy); 2nd method — the isocentre from original plan geometry is moved according to the movement of the CT V (simulation of IG based on soft tissues)

The contours, previously prepared by a radiation oncologist on original set of CT images, are used by an in-house script to create a set of artificial set of CT images — aCT(0,0,0). Our script used external (body), bones and CTV contours to create masks on CT images (Fig. 1. — In-house engine section). The pixel value assigned to the body was 0 HU, the pixel value assigned to the bone and CTV was 1000 HU. In the next step, the in-house script simulates on aCT(0,0,0) all possible positions of the CTV and, finally, generates the group of aCT sets that differ in the position of the CTV in each direction (x, y and z) — aCT(x,y,z). The positions of the CTV are quantified by the pixel value (i.e. 0.98 mm) and are ranged in the CTV-PTV margin (10 mm buffer zone added spherically to CTV). All of the aCT sets are uploaded to the Velocity software where they are used to generate the deformation matrixes and then to generate deformed CT sets (dCT) based on the original (planning) CT. The whole procedure was based on manual scripting in the VelocityTM software and included respectively: rigid registrations between aCT(0,0,0) and any other aCT(x,y,z); deformable registrations between aCT(0,0,0) and any other aCT(x,y,z) to generate deformation matrixes (DM(x,y,z)) caused from different positions of CTV on aCT(0,0,0) and aCT(x,y,z); appling DM(x,y,z) to original set of CT images and generating deformed CT sets (dCT). The same deformation matrixes were used to deform OARs contours. The deformable multipass algorithm was used for the last two steps of the procedure [19]. The dCT sets are moved to the Eclipse software where they are linked with the original plan geometry. For each dCT set, the doses were re-calculated. Two possible scenarios of IG procedures were considered [7, 20] when doses were re-calculated: the first method: re-calculation is done for the original geometry. It simulates IG based on the bony anatomy (not included bone-independent movements of the CTV) — Figure 2A;
Figure 2

Schemes of uncertainty simulations used in the study

the second method: the isocentre from the original plan geometry is moved according to the movement of the CTV (simulates IG based on soft tissues) — Figure 2B. Finally, the dose distributions for the CTV, rectum and bladder for both scenarios were analysed used dose-volume histogram (DVH) metrics, in the same way as is possible for the currently developed tool in EclipseTM TPS (Fig. 2C).

Results

There is about 1700 possible deformations of the original set of CT images (dCT). This value was calculated by a script which moves a CTV mask in the step of one voxel in all directions and checks if it is still inside PTV. Number of possible positions of the CTV mask is dependent on the CTV-PTV margin. Twenty-six of them were selected as the most significant for the results. These were the most extreme positions of the CTV inside the PTV in each direction (x, y, z) and their combinations (xy, yz, xz, xyz). The dCT sets, obtained in this way, were exported to EclipseTM TPS where the doses were re-calculated for them according to two proposed scenarios. Additionally, the original treatment plan has been simulated by the Varian’s formula implemented in the System. Figure 3 shows the uncertainty of the dose distribution in CTV, bladder and rectum for two scenarios of IG, calculated by authors’ solution and by the tool commercially implemented in EclipseTM TPS.
Figure 3

The uncertainty of the dose distribution in CT V, bladder and rectum for two scenarios of IG, calculated by authors’ solution (A, B) and the Varian’s formula implemented in the EclipseTM TPS (C)

Discussion

The term “robustness” in the light of the radiotherapeutic plans is considered in two ways. First of them is a robust plan optimisation — the set of procedures that allows minimising potential uncertainties of the dose distribution during the plan preparation. During the robust optimisation, a lot of scenarios were considered. Each scenario represents a possible treatment course and should include a specification of all the errors that are needed to calculate the final dose distribution [21]. The second way is a robust plan analysis — the analysis of the potential uncertainties of dose distribution for a single plan prepared in a conventional way (i.e., patient in a fixed nominal position and considering CTV-PTV margins). In other words, the intention of a robust optimisation is a reduction of potential uncertainties of doses on the level of plan preparation and, then, proper irradiation. In contrast, a robust plan analysis answers the question of how accurate is the treatment based on the plan created conventionally. This study focuses on a robust plan analysis. EclipseTM TPS has a tool for calculation of the dose uncertainty for conventionally prepared treatment plans. The procedure of uncertainty simulation in this tool is quite simplified and based on moving the whole patient’s body around the isocentre of the treatment (Fig. 2C). The scenarios of available simulations there do not include the changes of dose distribution caused by deformations of the soft tissues that are accompanying the shift in the position of the CTV. Moreover, they do not precisely simulate the different scenarios of IG that have an impact on the difference in doses. When daily is IG based on the bony anatomy, the isocentre of the treatment is fixed in the same geometrical place for each fraction. Potential differences in the fraction doses in the tumour (CTV) as well as in organs at risk (OARs) are caused by the movement of the CTV and different filling of OARs (i.e. bladder, rectum) during fractions (Fig. 2A). In the case of daily IG based on soft tissues, the isocentre of the treatment depends on the current position of the CTV before the fraction dose delivery. While the impact of OARs deformation on the doses is similar, the inter-fraction tracking of CTV position reduces effectively the dose differences in the CTV caused by different positions of the CTV. Nevertheless, the different distance between the skin and the isocentre may cause a difference in doses (Fig. 2B). The observations listed above motivated us to develop an in-house solution that enabled more precise simulations of dose uncertainties. In our solution, as in the Varian’s tool, the dose uncertainty analysis is based on the information gathered on the DVH. The DVHs from our method are more informative than the DVH from commercially implemented tools. For the first scenario (IG based on bony anatomy, Fig. 3A), the highest impact on dose uncertainty has boundary positions of the CTV to the CTV-PTV margin and relocation of the CTV that causes changes of OARs volumes that receive high doses. For the second scenario (IG based on soft tissues, Fig. 3B), dose uncertainty is affected to the highest degree by the CTV position in relation to the whole body and relocation of the isocentre that causes changes in the general dose distribution (visible for OARs volume receiving low and intermediate doses). The current version of our procedure is limited to the solid (nondeformed) CTVs. Except the in-house engine used to generate aCT sets with different position of the CTV, the tools (i.e. VelocityTM and EclipseTM) used in our simulations are developed by the Varian Medical Systems. They are being used commercially in a lot of centres based on the Varian’s radiotherapeutic line. Moreover, it is possible to automatise the procedures carried out in VelocityTM and EclipseTM by scripting language implemented in these programs. Due to this fact, it is easy to implement this solution in centres based on Varian’s line.

Conclusion

The proposed method of simulating the uncertainty of dose distributions enables a more accurate analysis of the treatment plan robustness than the method currently implemented in EclipseTM TPS.
  18 in total

1.  The effect of breathing and set-up errors on the cumulative dose to a lung tumor.

Authors:  M Engelsman; E M Damen; K De Jaeger; K M van Ingen; B J Mijnheer
Journal:  Radiother Oncol       Date:  2001-07       Impact factor: 6.280

2.  Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation.

Authors:  Thomas Bortfeld; Kimmo Jokivarsi; Michael Goitein; Jong Kung; Steve B Jiang
Journal:  Phys Med Biol       Date:  2002-07-07       Impact factor: 3.609

3.  Effects of motion on the total dose distribution.

Authors:  Thomas Bortfeld; Steve B Jiang; Eike Rietzel
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

Review 4.  MRI-guided prostate adaptive radiotherapy - A systematic review.

Authors:  A J McPartlin; X A Li; L E Kershaw; U Heide; L Kerkmeijer; C Lawton; U Mahmood; F Pos; N van As; M van Herk; D Vesprini; J van der Voort van Zyp; A Tree; A Choudhury
Journal:  Radiother Oncol       Date:  2016-05-06       Impact factor: 6.280

5.  Evaluation of image-guidance protocols in the treatment of head and neck cancers.

Authors:  Omar A Zeidan; Katja M Langen; Sanford L Meeks; Rafael R Manon; Thomas H Wagner; Twyla R Willoughby; D Wayne Jenkins; Patrick A Kupelian
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-12-29       Impact factor: 7.038

6.  Effect of imaging frequency on PTV margins and geographical miss during image guided radiation therapy for prostate cancer.

Authors:  Meetakshi Gupta; Poonam Gamre; Sadhana Kannan; Ganesh Rokde; Rahul Krishnatry; Vedang Murthy
Journal:  Pract Radiat Oncol       Date:  2017-10-07

7.  Use of a realistic breathing lung phantom to evaluate dose delivery errors.

Authors:  Laurence E Court; Joao Seco; Xing-Qi Lu; Kazuyu Ebe; Charles Mayo; Dan Ionascu; Brian Winey; Nikos Giakoumakis; Michalis Aristophanous; Ross Berbeco; Joerg Rottman; Madeleine Bogdanov; Deborah Schofield; Tania Lingos
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

8.  Software simulation of tumour motion dose effects during flattened and unflattened ITV-based VMAT lung SBRT.

Authors:  Marta Adamczyk; Marta Kruszyna-Mochalska; Anna Rucińska; Tomasz Piotrowski
Journal:  Rep Pract Oncol Radiother       Date:  2020-06-11

9.  Evaluation of image-guidance strategies for prostate cancer.

Authors:  T Piotrowski; K Kaczmarek; T Bajon; A Ryczkowski; A Jodda; J Kaźmierska
Journal:  Technol Cancer Res Treat       Date:  2013-08-31

10.  Benchmarking of five commercial deformable image registration algorithms for head and neck patients.

Authors:  Jason Pukala; Perry B Johnson; Amish P Shah; Katja M Langen; Frank J Bova; Robert J Staton; Rafael R Mañon; Patrick Kelly; Sanford L Meeks
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102