The effect of breathing and set-up errors on the cumulative dose to a lung tumor.

M Engelsman1, E M Damen, K De Jaeger, K M van Ingen, B J Mijnheer.   

Abstract

BACKGROUND AND
PURPOSE: To assess the impact of both set-up errors and respiration-induced tumor motion on the cumulative dose delivered to a clinical target volume (CTV) in lung, for an irradiation based on current clinically applied field sizes.
MATERIALS AND METHODS: A cork phantom, having a 50 mm spherically shaped polystyrene insertion to simulate a gross tumor volume (GTV) located centrally in a lung was irradiated with two parallel opposed beams. The planned 95% isodose surface was conformed to the planning target volume (PTV) using a multi leaf collimator. The resulting margin between the CTV and the field edge was 16 mm in beam's eye view. A dose of 70 Gy was prescribed. Dose area histograms (DAHs) of the central plane of the CTV (GTV+5 mm) were determined using radiographic film for different combinations of set-up errors and respiration-induced tumor motion. The DAHs were evaluated using the population averaged tumor control probability (TCP(pop)) and the equivalent uniform dose (EUD) model.
RESULTS: Compared with dose volume histograms of the entire CTV, DAHs overestimate the impact of tumor motion on tumor control. Due to the choice of field sizes a large part of the PTV will receive a too low dose resulting in an EUD of the central plane of the CTV of 68.9 Gy for the static case. The EUD drops to 68.2, 66.1 and 51.1 Gy for systematic set-up errors of 5, 10 and 15 mm, respectively. For random set-up errors of 5, 10 and 15 mm (1 SD), the EUD decreases to 68.7, 67.4 and 64.9 Gy, respectively. For similar amplitudes of respiration-induced motion, the EUD decreases to 68.8, 68.5 and 67.7 Gy, respectively. For a clinically relevant scenario of 7.5 mm systematic set-up error, 3 mm random set-up error and 5 mm amplitude of breathing motion, the EUD is 66.7 Gy. This corresponds with a tumor control probability TCP(pop) of 41.7%, compared with 50.0% for homogeneous irradiation of the CTV to 70 Gy.
CONCLUSION: Systematic set-up errors have a dominant effect on the cumulative dose to the CTV. The effect of breathing motion and random set-up errors is smaller. Therefore the gain of controlling breathing motion during irradiation is expected to be small and efforts should rather focus on minimizing systematic errors. For the current clinically applied field sizes and a clinically relevant combination of set-up errors and breathing motion, the EUD of the central plane of the CTV is reduced by 3.3 Gy, at maximum, relative to homogeneous irradiation of the CTV to 70 Gy, for our worst case scenario.

Entities:  

Mesh:

Year:  2001        PMID: 11410310     DOI: 10.1016/s0167-8140(01)00349-8

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  24 in total

1.  A novel technique for markerless, self-sorted 4D-CBCT: feasibility study.

Authors:  Irina Vergalasova; Jing Cai; Fang-Fang Yin
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  Development of a novel post-processing treatment planning platform for 4D radiotherapy.

Authors:  Lan Lin; Chengyu Shi; Yaxi Liu; Gregory Swanson; Nikos Papanikolaou
Journal:  Technol Cancer Res Treat       Date:  2008-04

Review 3.  Robustness Analysis for External Beam Radiation Therapy Treatment Plans: Describing Uncertainty Scenarios and Reporting Their Dosimetric Consequences.

Authors:  Adam D Yock; Radhe Mohan; Stella Flampouri; Walter Bosch; Paige A Taylor; David Gladstone; Siyong Kim; Jason Sohn; Robert Wallace; Ying Xiao; Jeff Buchsbaum
Journal:  Pract Radiat Oncol       Date:  2018-12-15

Review 4.  MRI in medical practice and its future use in radiation oncology. Resume of XXV GOCO Congress (Montpellier) 2017.

Authors:  Xavier Druet; Estrella Acosta Sanchez; Ken Soleakhena; Anne Laprie; Jordi Sáez; Stéphanie Nougaret; Olivier Riou; Elodie Rigal; Laura Kibranian; Miguel Palacios; Ismael Membrive
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-05

5.  A phase I/II study on stereotactic body radiotherapy with real-time tumor tracking using CyberKnife based on the Monte Carlo algorithm for lung tumors.

Authors:  Hiromitsu Iwata; Satoshi Ishikura; Taro Murai; Michio Iwabuchi; Mitsuhiro Inoue; Koshi Tatewaki; Seiji Ohta; Naoki Yokota; Yuta Shibamoto
Journal:  Int J Clin Oncol       Date:  2017-04-20       Impact factor: 3.402

Review 6.  Motion management in gastrointestinal cancers.

Authors:  Hassan Abbas; Bryan Chang; Zhe Jay Chen
Journal:  J Gastrointest Oncol       Date:  2014-06

7.  Patient-specific quantification of respiratory motion-induced dose uncertainty for step-and-shoot IMRT of lung cancer.

Authors:  Heng Li; Peter Park; Wei Liu; Jason Matney; Zhongxing Liao; Peter Balter; Yupeng Li; Xiaodong Zhang; Xiaoqiang Li; X Ronald Zhu
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

8.  Adaptive radiation for lung cancer.

Authors:  Daniel R Gomez; Joe Y Chang
Journal:  J Oncol       Date:  2010-08-04       Impact factor: 4.375

9.  On the accuracy of a moving average algorithm for target tracking during radiation therapy treatment delivery.

Authors:  Rohini George; Yelin Suh; Martin Murphy; Jeffrey Williamson; Elizabeth Weiss; Paul Keall
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

10.  Advances in 4D radiation therapy for managing respiration: part II - 4D treatment planning.

Authors:  Mihaela Rosu; Geoffrey D Hugo
Journal:  Z Med Phys       Date:  2012-07-15       Impact factor: 4.820

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.