Vaishali Narendra Naik1, R S Gamad2, P P Bansod3.
Abstract
BACKGROUND: Carotid intima-media thickness (IMT) measured in B-mode ultrasound image is an important indicator of Atherosclerosis disease. Speckle noise inherently present in ultrasounds' thereby degrades the visual evaluation and limits the automated segmentation performance. The objective of this study is to investigate the effects of three despeckle filters on the segmentation of carotid IMT in ultrasound image.Entities:
Keywords: Atherosclerosis; Carotid artery; Despeckle filter; Intima-media thickness; Segmentation
Mesh:
Year: 2021 PMID: 34273550 PMCID: PMC9486865 DOI: 10.1016/j.bj.2021.07.002
Source DB: PubMed Journal: Biomed J ISSN: 2319-4170 Impact factor: 7.892
Fig. 1Block diagram of IMT measurement scheme.
Comparison between Manual, Without_F and Median filter (with square window size (N) = 3, 5 and 7) here measurand is IMT in mm.
| n = 151 | Manual | ||||
|---|---|---|---|---|---|
| Mean | 0.9536 | 1.004427 | 0.9832 | 0.9471 | 0.9063 |
| std | 0.2854 | 0.306744 | 0.2946 | 0.2824 | 0.2805 |
| CV% | 21.1597 | 21.59448 | 21.1849 | 21.0860 | 21.8854 |
| r | 1 | 0.924617 | 0.9322 | 0.9823 | 0.9624 |
| tp (sec) | – | 14.07 | 14.43 | 13.71 | 13.72 |
Abbreviations: n: number of samples; std: standard deviation; CV%: coefficient of variation; r: Pearson's correlation; tp: average processing time.
Comparison between Manual, Without_F and Hybrid median filter (with square window size 3, 5 and 7) here measurand is IMT in mm.
| n = 151 | Manual | ||||
|---|---|---|---|---|---|
| Mean | 0.9536 | 1.0044 | 0.9903 | 0.9715 | 0.9469 |
| std | 0.2854 | 0.3067 | 0.3002 | 0.2953 | 0.2914 |
| CV% | 21.1597 | 21.5945 | 21.4358 | 21.4891 | 21.7588 |
| r | 1 | 0.9246 | 0.9350 | 0.9416 | 0.9434 |
| tp (sec) | – | 14.07 | 14.04 | 14.34 | 14.28 |
Abbreviations: n: number of samples; std: standard deviation; CV%: coefficient of variation; r: Pearson's correlation; tp: average processing time.
Comparison between Manual, Without_F and IACDF filter (for tD (sec): 0.5, 0.6, 0.65, 0.7, 0.75, 0.8, 0.85, 0.9 and 0.95) here measurand is IMT in mm.
| n = 151 | Manual | tD = 0.5 | tD = 0.6 | tD = 0.65 | tD = 0.7 | tD = 0.75 | tD = 0.8 | tD = 0.85 | tD = 0.9 | tD = 0.95 | tD = 1 | tD = 2 | tD = 3 | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Mean | 0.9536 | 1.0044 | 0.9733 | 0.9720 | 0.9722 | 0.9684 | 0.9675 | 0.9621 | 0.9601 | 0.9479 | 0.9533 | 0.9542 | 0.9339 | 0.9198 |
| std | 0.2854 | 0.3067 | 0.2947 | 0.29459 | 0.2984 | 0.2989 | 0.2914 | 0.2948 | 0.2949 | 0.2916 | 0.2904 | 0.2934 | 0.2983 | 0.2994 |
| CV% | 21.1597 | 21.5945 | 21.4108 | 21.4228 | 21.7066 | 21.8238 | 21.2971 | 21.6646 | 21.7197 | 21.7518 | 21.5421 | 21.7391 | 22.5841 | 23.0156 |
| r | 1 | 0.9246 | 0.9615 | 0.9601 | 0.9600 | 0.9576 | 0.9617 | 0.9614 | 0.9587 | 0.8863 | 0.9601 | 0.9508 | 0.9462 | 0.9348 |
| tp (sec) | – | 14.07 | 14.16 | 14.20 | 14.21 | 14.25 | 14.26 | 14.31 | 14.32 | 14.36 | 14.38 | 14.41 | 14.94 | 15.39 |
Abbreviations: n: number of samples; std: standard deviation; CV%: coefficient of variation; r: Pearson's correlation; tp: average processing time; tD: diffusion time.
Comparison between Without_F, , H_medianW7 and IACDF (measurand: IMT in mm).
| n = 151 | Manual | IACDF | |||
|---|---|---|---|---|---|
| Mean | 0.9536 | 1.004427 | 0.9471 | 0.9469 | 0.9675 |
| std | 0.2854 | 0.306744 | 0.2824 | 0.2914 | 0.2914 |
| CV% | 21.1597 | 21.59448 | 21.0860 | 21.7588 | 21.2971 |
| r | – | 0.924617 | 0.9823 | 0.9434 | 0.9617 |
| tp (sec) | – | 14.07 | 13.71 | 14.28 | 14.26 |
Abbreviations: n: number of samples; std: standard deviation; CV%: coefficient of variation; r: Pearson's correlation; tp: average processing time; tD: diffusion time.
Fig. 3Performance of Median_W5, H_medianW7 and IACDF_0.75 filters against speckle noise variance 0, 0.05, 0.1, 0.15, 0.2 and 0.25, plot between noise and (A) average IMT error; (B) average CV%; and (C) average correlation between manual and automated methods; calculated from a total of ten images.
Fig. 4Performance of Median_W5, H_medianW7 and IACDF_0.75 filters against salt and pepper noise density 0, 0.05, 0.1, 0.15, 0.2 and 0.25, plot between noise and (A) average IMT error; (B) average CV%; and (C) average correlation between manual and automated methods; calculated from a total of ten images.