| Literature DB >> 34267718 |
Jong Hyeon Ahn1,2, Joomee Song1,2, Inyoung Choi1,2, Jinyoung Youn1,2, Jin Whan Cho1,2.
Abstract
Objectives: We aimed to validate the accuracy of blood pressure (BP) measurement using a smartwatch in patients with Parkinson's disease (PD). Materials andEntities:
Keywords: Parkinson's disease; blood pressure; photoplethysmography; sphygmomanometer; validation; wearable device
Year: 2021 PMID: 34267718 PMCID: PMC8275847 DOI: 10.3389/fneur.2021.650929
Source DB: PubMed Journal: Front Neurol ISSN: 1664-2295 Impact factor: 4.003
Clinical characteristics and demographics of the study population.
| Age (years) | 66.9 ± 9.1 |
| Sex (male) | 33 (58.9%) |
| Disease duration (month) | 61.8 ± 51.6 |
| BMI (kg/m2) | 23.6 ± 3.3 |
| No. of qualified participants | 56 |
| No. of BP pairs | 168 |
| UPDRS III (on) | 17.8 ± 10.7 |
| H & Y stage (on) | 2.1 ± 0.9 |
| MMSE | 28.3 ± 1.2 |
| LED (mg) | 496.7 ± 317.3 |
| Circumstance of arm (cm) | 26.5 ± 2.8 |
| SBP (proportion of measurements), mm Hg | |
| ≥160 | 10 (6.0%) |
| ≥140 | 15 (8.9%) |
| ≤ 100 | 16 (9.5%) |
| DBP (proportion of measurements), mm Hg | |
| ≥100 | 9 (5.3%) |
| ≥85 | 32 (19.0 %) |
| ≤ 60 | 10 (3.0%) |
| Heart rate (pulse/min) | 76.4 ± 11.7 |
Data are expressed as the number (%) or mean ± standard deviation. Fifty-six participants, with a total of 168 blood pressure measurements. BMI, body mass index; BP, blood pressure; UPDRS, United Parkinson's disease rating scale; H & Y, Hoehn and Yahr; MMSE, mini-mental state examination; LED, Levodopa equivalent dose; SBP, systolic blood pressure; DBP, diastolic blood pressure.
Figure 1Bland–Altman plots for the differences between the SM-R850 readings and the reference measurements for SBP (A), and DBP (B) (n = 168). SBP, systolic blood pressure; DBP, diastolic blood pressure.