Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol.

Alexandra C Cuckson1, Annemarie Reinders, Husain Shabeeh, Andrew H Shennan.   

Abstract

BACKGROUND: The market for devices for the self-measurement of blood pressure is growing, and as accuracy is of prime importance, there is increasing pressure for manufacturers to provide evidence of independent testing. Recent reviews have shown that only five automated upper arm devices for self-measurement of blood pressure have been recommended for use. We tested the Microlife BP 3BTO-A, a lightweight, upper arm, automated oscillometric device, according to a modified version of the British Hypertension Society protocol and also analysed the computer-generated oscillograms for possible causes of inaccuracy.
METHODS: One hundred and twenty-six subjects were recruited from general medical and specialist clinics and from amongst the staff at Guy's and St Thomas' Hospital, London, UK. Only 85 of these were included in the final analysis. Nine sequential readings were taken by two trained observers alternating between the mercury sphygmomanometer and the device. The last seven readings were analysed according to the British Hypertension Society protocol. Modifications to the protocol were: (1) the exclusion of patients whose blood pressure varied by more than 15 mmHg between sequential observer readings and (2) limited testing in the low systolic pressure range.
RESULTS: The Microlife achieved a grade A for both systolic and diastolic pressure according to the British Hypertension Society protocol. The mean differences (standard deviation) between the observers and the device were -1.6 (7.7) mmHg and -2.1 (6.3) mmHg for systolic and diastolic blood pressure, respectively, therefore also fulfilling the criteria set by the Association for the Advancement of Medical Instrumentation. Sub-analysis for different pressure ranges showed that the device was less accurate in the high-pressure range (>160/100 mmHg).
CONCLUSION: The Microlife can be recommended for clinical use in an adult population. Copyright 2002 Lippincott Williams & Wilkins

Entities:  

Mesh:

Year:  2002        PMID: 12488652     DOI: 10.1097/00126097-200212000-00005

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  22 in total

1.  Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines.

Authors:  Rajab Ali Khawaja; Riaz Qureshi; Abdul Hay Mansure; Mohammad Emran Yahya
Journal:  J Saudi Heart Assoc       Date:  2010-03-10

2.  Hypertension awareness and control among young adults in the national longitudinal study of adolescent health.

Authors:  Holly C Gooding; Shannon McGinty; Tracy K Richmond; Matthew W Gillman; Alison E Field
Journal:  J Gen Intern Med       Date:  2014-02-28       Impact factor: 5.128

3.  Hypertension prevalence and control in Ulaanbaatar, Mongolia.

Authors:  Harry Potts; Uurtsaikh Baatarsuren; Maral Myanganbayar; Baigal Purevdorj; Burtu-Ujin Lkhagvadorj; Namuun Ganbat; Alimaa Dorjpalam; Delgerbat Boldbaatar; Khulan Tuvdendarjaa; Dulmaa Sampilnorov; Khatantuul Boldbaatar; Myagmartseren Dashtseren; Batbold Batsukh; Namkhaidorj Tserengombo; Tsolmon Unurjargal; Enkhtuya Palam; Roberta Bosurgi; Geoffrey So; Norm R C Campbell; Andreas Bungert; Naranbaatar Dashdorj; Naranjargal Dashdorj
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-01-08       Impact factor: 3.738

4.  Distribution of hyperglycaemia and related cardiovascular disease risk factors in low-income countries: a cross-sectional population-based survey in rural Uganda.

Authors:  Dermot Maher; Laban Waswa; Kathy Baisley; Alex Karabarinde; Nigel Unwin; Heiner Grosskurth
Journal:  Int J Epidemiol       Date:  2010-10-05       Impact factor: 7.196

5.  Acute and chronic effects of resistance exercise on blood pressure in elderly women and the possible influence of ACE I/D polymorphism.

Authors:  Márcio Rabelo Mota; Ricardo Jacó Oliveira; Denize Faria Terra; Emerson Pardono; Maurílio Tiradentes Dutra; Jeeser Alves de Almeida; Francisco Martins Silva
Journal:  Int J Gen Med       Date:  2013-07-12

6.  Effect of peripheral edema on oscillometric blood pressure measurement.

Authors:  Shamsi Ghaffari; Majid Malaki; Afshin Rezaeifar; Shahin Abdollahi Fakhim
Journal:  J Cardiovasc Thorac Res       Date:  2014-12-30

7.  Validation of Blood Pressure Measurement Using a Smartwatch in Patients With Parkinson's Disease.

Authors:  Jong Hyeon Ahn; Joomee Song; Inyoung Choi; Jinyoung Youn; Jin Whan Cho
Journal:  Front Neurol       Date:  2021-06-29       Impact factor: 4.003

8.  Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Authors:  Jalil Belghazi; Ramzi N El Feghali; Thérèse Moussalem; Maya Rejdych; Roland G Asmar
Journal:  Vasc Health Risk Manag       Date:  2007

9.  Sarcopenia Is Associated with High Pulse Pressure in Older Women.

Authors:  Hélio José Coelho Júnior; Samuel da Silva Aguiar; Ivan de Oliveira Gonçalves; Ricardo Aurélio Carvalho Sampaio; Marco Carlos Uchida; Milton Rocha Moraes; Ricardo Yukio Asano
Journal:  J Aging Res       Date:  2015-08-05

10.  Sustained effect of resistance training on blood pressure and hand grip strength following a detraining period in elderly hypertensive women: a pilot study.

Authors:  Dahan da Cunha Nascimento; Ramires Alsamir Tibana; Franklin M Benik; Keila Elizabeth Fontana; Frederico Ribeiro Neto; Frederico Santos de Santana; Leopoldo Santos-Neto; Renato André Sousa Silva; Alessandro Oliveira Silva; Darlan Lopes Farias; Sandor Balsamo; Jonato Prestes
Journal:  Clin Interv Aging       Date:  2014-01-20       Impact factor: 4.458

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.