Literature DB >> 29057775

Influence of the Pixel Sizes of Reference Computed Tomography on Single-photon Emission Computed Tomography Image Reconstruction Using Conjugate-gradient Algorithm.

Kyohei Okuda1, Shota Sakimoto1, Susumu Fujii1, Tomonobu Ida1, Shigeru Moriyama1.   

Abstract

PURPOSE: The frame-of-reference using computed-tomography (CT) coordinate system on single-photon emission computed tomography (SPECT) reconstruction is one of the advanced characteristics of the xSPECT reconstruction system. The aim of this study was to reveal the influence of the high-resolution frame-of-reference on the xSPECT reconstruction.
METHODS: 99mTc line-source phantom and National Electrical Manufacturers Association (NEMA) image quality phantom were scanned using the SPECT/CT system. xSPECT reconstructions were performed with the reference CT images in different sizes of the display field-of-view (DFOV) and pixel.
RESULTS: The pixel sizes of the reconstructed xSPECT images were close to 2.4 mm, which is acquired as originally projection data, even if the reference CT resolution was varied. The full width at half maximum (FWHM) of the line-source, absolute recovery coefficient, and background variability of image quality phantom were independent on the sizes of DFOV in the reference CT images.
CONCLUSION: The results of this study revealed that the image quality of the reconstructed xSPECT images is not influenced by the resolution of frame-of-reference on SPECT reconstruction.

Entities:  

Keywords:  computed-tomography (CT) coordinate system; display field-of-view (DFOV); frame-of-reference; single-photon emission computed tomography (SPECT)

Mesh:

Year:  2017        PMID: 29057775     DOI: 10.6009/jjrt.2017_JSRT_73.10.1039

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Validation of the influence of CT slice thickness on the quantitative accuracy and image quality of single photon emission computed tomography.

Authors:  Tomohiro Sato; Takashi Takagi
Journal:  Asia Ocean J Nucl Med Biol       Date:  2021
  1 in total