| Literature DB >> 34249976 |
Shaomin Zhang1, Hong Xian1, Yi Chen1, Yue Liao1, Nan Zhang1, Xinyu Guo1, Ming Yang1, Jinhui Wu1.
Abstract
Background: AMAZFIT®, a novel wearable electrocardiogram (ECG)-recording system is used for the measurement, acquisition, and storage of single-lead cardiac waveforms for adults. The aim of the study was to evaluate the accuracy of AMAZFIT® for diagnosing arrhythmia in older patients.Entities:
Keywords: AMAZFIT; arrhythmia; atrial fibrillation; detection; diagnostic trial
Year: 2021 PMID: 34249976 PMCID: PMC8264252 DOI: 10.3389/fmed.2021.685999
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Figure 1AMAZFIT® location diagram. (A): wrist (B): chest wall.
Figure 2Comparison of the waveforms from ECG and from AMAZFIT®. (A): sinus rhythm; (B): AF; (C): PACs; (D): PVCs; (E): first-degree AVB. AF, arterial fibrillation; PACs, premature arterial contractions; PVCs, premature ventricular contractions; AVB, atrioventricular block.
Baseline characteristics of the study population.
| Age, mean (SD), years | 76 ± 10 | 84 ± 9 | 80 ± 10 | 76 ± 9 | 79 ± 9 |
| Female | 56(47.1) | 7(25.0) | 13(52.0) | 6(21.4) | 38(40.4) |
| Hypertension | 65(54.6) | 23(82.1) | 19(76.0) | 20(71.4) | 59(62.8) |
| COPD | 22(18.5) | 11(39.3) | 4(16.0) | 5(17.9) | 24(25.5) |
| Diabetes | 19(16.0) | 12(42.9) | 5(20.0) | 5(17.9) | 25(26.6) |
| Coronary artery disease | 27(22.7) | 12(42.9) | 7(28.0) | 6(21.4) | 26(27.7) |
| Stroke | 22(18.5) | 6(21.4) | 8(32.0) | 6(21.4) | 9(9.6) |
| β- blocker | 45(37.8) | 7(25.0) | 11(44.0) | 14(50.0) | 28(29.8) |
| CCB | 34(28.6) | 13(46.4) | 13(52.0) | 10(35.70) | 24(25.5) |
| Digoxin | 18(15.1) | 0(0.0) | 0(0.0) | 2(7.1) | 1(1.1) |
| Amiodarone | 21(17.6) | 1(3.6) | 6(24.0) | 3(10.7) | 7(7.5) |
AF, atrial fibrillation; AVB, atrioventricular block; PACs, premature atrial contraction; PVCs, premature ventricular contraction; ECG, electrocardiography; COPD, chronic obstructive pulmonary disease; CCB, calcium channel blockers.
Figure 3Participant selection and summary of ECG results.
Diagnostic accuracy of AMAZFIT® against ECG.
| Arrhythmia | 197 | 89.34 | 92.55 | 97.59 | 96.17 | 80.56 | 11.99 | 0.12 |
| AF | 119 | 93.28 | 95.35 | 94.50 | 93.28 | 95.35 | 20.06 | 0.07 |
| First-degree AVB | 28 | 32.14 | 97.72 | 91.41 | 60.00 | 93.12 | 14.10 | 0.69 |
| PACs | 25 | 84.00 | 96.62 | 95.53 | 70.00 | 98.47 | 24.00 | 0.16 |
| PVCs | 28 | 89.29 | 93.92 | 93.47 | 60.98 | 98.80 | 14.69 | 0.11 |
| Normal ECG | 94 | 92.55 | 89.34 | 90.38 | 80.56 | 96.17 | 8.68 | 0.08 |
PPV, positive predictive value; NPV, negative predictive value; PLR, positive likelihood ratio; NLR, negative likelihood ratio; AF, atrial fibrillation; AVB, atrioventricular block; PACs, premature atrial contraction; PVCs, premature ventricular contraction; ECG, electrocardiography.
Figure 4Bland-Altman plots of the heart ratio between the ECG and AMAZFIT®.