Galit Levin1, Paul Jm Sessink2.
Abstract
PURPOSE: The purpose of this study was to test the efficacy of ChemfortTM, an air filtration closed-system drug transfer device to prevent release of chemotherapy drug vapors and aerosols under extreme conditions. The air cleaning system is based on the adsorption of drug vapors by an activated carbon filter in the Vial Adaptor before the air is released out of the drug vial. The functionality of the carbon filter was also tested at the end of device's shelf life, and after a contact period with drug vapors for 7 days. Cyclophosphamide and 5-fluorouracil were the chemotherapy drugs tested.Entities:
Keywords: Antineoplastic agents; activated carbon filter; air cleaning technology; closed-system drug transfer device; drug vapors
Mesh:
Substances:
Year: 2021 PMID: 34229499 PMCID: PMC9465531 DOI: 10.1177/10781552211030682
Source DB: PubMed Journal: J Oncol Pharm Pract ISSN: 1078-1552 Impact factor: 1.416
Figure 1.Schematic set up of the test equipment with path of airflow indicated.
Cyclophosphamide (ng) on surfaces in the vapor trap system after 5 hours of vapor/aerosol trapping.
| Tested sample | Vial Adaptor without activated carbon filter | Vial Adaptor with activated carbon filter | ||||||
|---|---|---|---|---|---|---|---|---|
| New | Aged | Aged and 7 days exposure to drug vapors | ||||||
| Replicate | 1 | 2 | 3 | 4 | 5 | 6–10 | 11–12 | 13–17 |
|
| ||||||||
| Rinse sample | ||||||||
| Vessel cold trap | ND | ND | ND | ND | ND | ND | ND | ND |
| Tubing | 1.2 | 1.0 | 12.8 | 8.5 | 26.0 | ND | ND | ND |
| Wipe sample | ||||||||
| Vessel top | ND | 0.6 | 1.7 | 4.0 | 15.5 | ND | ND | ND |
| Vessel bottom | ND | ND | 0.2 | 3.8 | 0.1 | ND | ND | ND |
| Drug vial and Vial Adaptor | 20.8 | 59.1 | 74.8 | 43.9 | 70.6 | ND | ND | ND |
| Total | 22.0 | 60.7 | 89.5 | 60.2 | 112.2 | ND | ND | ND |
| Mean ± SD | 68.9 ± 34.1 | ND | ND | ND | ||||
ND: not detected (< 0.1 ng in solution or on wipe).
5-Fluorouracil (ng) on surfaces in the vapor trap system after 5 hours of vapor/aerosol trapping.
Tested sample | Vial Adaptor without activated carbon filter and without membrane | Vial Adaptor without activated carbon filter | Vial Adaptor with activated carbon filter | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Replicate | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11–15 |
| Rinse sample | |||||||||||
| Vessel cold trap | ND | ND | ND | ND | ND | ND | ND | ND | ND | ND | ND |
| Tubing | ND | ND | ND | ND | ND | ND | ND | ND | ND | ND | ND |
| Wipe sample | |||||||||||
| Vessel top | 3.7 | 4.2 | 5.2 | ND | ND | ND | ND | 4.3 | 5.6 | ND | ND |
| Vessel bottom | 4.5 | 7.9 | 6.7 | 6.7 | 43.8 | 5.0 | 3.6 | 5.3 | 6.8 | 10.2 | ND |
| Drug vial and Vial Adaptor | 102.3 | 17.5 | 136.0 | 4.5 | 6.0 | 39.9 | 50.4 | 5.5 | 25.6 | 37.6 | ND |
| Total | 110.5 | 29.6 | 147.9 | 11.2 | 49.8 | 44.9 | 54.0 | 15.1 | 38.0 | 47.8 | ND |
| Mean ± SD | 96.0 ± 60.5 | 37.3 ± 17.2 | ND | ||||||||
ND: not detected (< 2 ng in solution or on wipe).