Alan-Shaun Wilkinson1, Michael Charles Allwood1, Colin Patrick Morris1, Andrew Wallace1, Rebecca Finnis1, Ewelina Kaminska1, Donata Stonkute1, Maja Szramowska1, Joe Miller1, Ian Pengelly2, Michael Hemingway2.
Abstract
OBJECTIVES: The United States National Institute for Occupational Safety and Health (NIOSH) is developing a protocol to assess the containment performance of closed system transfer devices (CSTDs) when used for drug preparation (task 1) and administration (task 2) and published a draft protocol in September 2016. Nine possible surrogates were proposed by NIOSH for use in the testing. The objectives of this study were to: (A) select the most appropriate surrogate; (B) validate the NIOSH protocol using this surrogate; and (C) determine the containment performance of four commercial CSTDs as compared with an open system of needle and syringe using the validated NIOSH protocol.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30379831 PMCID: PMC6209153 DOI: 10.1371/journal.pone.0205263
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Nine potential challenge agents suggested by NIOSH in the draft protocol [11].
| Compound | CAS | Vapor pressure | Henry’s constant H (atm m3/mole) |
|---|---|---|---|
| Dimethyl sulfoxide | 67-68-5 | 0.61 | 1.51 × 10−9 |
| Trimethyl phosphate | 512-56-1 | 0.85 | 7.20 × 10−9 |
| Tetramethylurea | 632-22-4 | 1.16 | 8.48 × 10−9 |
| Triacetin | 102-76-1 | 0.002 | 1.23 × 10−8 |
| Propylene glycol | 57-55-6 | 0.129 | 1.28 × 10−8 |
| Tetraethylurea | 1187-03-7 | 0.208 | 2.80 × 10−8 |
| Triethyl phosphate | 78-40-0 | 0.393 | 3.60 × 10−8 |
| 2-Phenoxyethanol | 122-99-6 | 0.007 | 4.72 × 10−8 |
| Tripropyl phosphate | 513-08-6 | 0.004 | 6.80 × 10−7 |
| Isopropyl alcohol | 67-63-0 | 100 | 7.90 × 10−6 |
aEstimated values used where experimental data were unavailable generated using the Environmental Protection Agency (EPA) EPI SUITE software [21].
bChemical Abstracts Service Registry.
cVapor pressure refers to the active pharmaceutical or active ingredient in its pure state.
d Vapour pressure data of Isopropyl alcohol shown for comparison purposes only.
Physicochemical data for some commonly used hazardous drugs.
| Compound | CAS | Vapor pressure | Henry’s Constant H (atm m3/mole) |
|---|---|---|---|
| Cisplatin | 015663-27-1 | 0.0000135 | 2.091 × 10−26 |
| Carboplatin | 041575-94-4 | 1.33 × 10−9 | 5.51 × 10−14 |
| Thiotepa | 52-24-4 | 0.00941 | 2.83 × 10−10 |
| Carmustine | 154-93-8 | 0.000142 | 4.76 × 10−11 |
| Cyclophosphamide | 50-18-0 | 0.0000247 | 1.36 × 10−11 |
| Etoposide | 33419-42-0 | 0.0000195 | 1.75 × 10−30 |
| 5-Fluorouracil | 51-21-8 | 0.0000105 | 1.66 × 10−10 |
aChemical Abstracts Service Registry.
b(Kiffmeyer et al., 2002) [20].
cCalculated using the modified grain method using Environmental Protection Agency (EPA) EPI SUITE software [21].
Fig 1Calibration plot for 2-phenoxyethanol by automated thermal desorption gas chromatography mass spectrometry.
Fig 2Biopharma Stability Testing Laboratory modified National Institute for Occupational Safety and Health test chamber with aluminum extension piece.
CSTD containment performance data using 2.5% v/v 2-phenoxyethanol (2-POE) as challenge agent according to draft NIOSH protocol and defined tasks: task 1 (n = 5) and task 2 (n = 5) performed under IFU conditions.
| CSTD product | NIOSH task | Mean concentration of 2-POE (ppb) |
|---|---|---|
| Tevadaptor (marketed as OnGuard In USA, Teva) | 1 | ≤ 0.88 |
| 2 | ≤ 0.88 | |
| PhaSeal (BD) | 1 | ≤ 0.88 |
| 2 | ≤ 0.88 | |
| ChemoClave vial shield and Spiros (ICU Medical) | 1 | 2.90 ± 2.25 |
| 2 | 7.50 ± 17.86 | |
| Equashield (Equashield) | 1 | ≤ 0.88 |
| 2 | ≤ 0.88 | |
| Needle and syringe | 1 | 4.20 ± 2.23 |
| 2 | 5.10 ± 1.71 | |
| Blank (N = 85) | — | 0.12 ± 0.02 |
CSTD: closed system transfer device; IFU: instructions for use; NIOSH: National Institute for Occupational Safety and Health; ppb: parts per billion.
* Limit of detection was determined to be 0.35 ppb and the limit of quantitation was determined to be 0.88 ppb.
† range calculated at the 95% confidence interval (CI).
Fig 3Typical gas chromatogram for 2-phenoxyethanol with mass spectrometry (selected M/Z values: 77, 94, and 138) detection.
Fig 4Liquid droplets observed on the membrane of ChemoClave CSTD following task 2.