Shunsuke Nakakura1, Ryo Asaoka2,3, Etsuko Terao4, Yuki Nagata4, Yasuko Fukuma4, Satomi Oogi4, Miku Shiraishi4, Yoshiaki Kiuchi5.
Abstract
BACKGROUND: This study investigated the agreement between a new rebound tonometer, IC200, and IcarePRO and Goldmann applanation tonometry (GAT).Entities:
Keywords: Glaucoma; Goldmann applanation tonometer; IC200; IcarePRO; Rebound tonometer
Year: 2021 PMID: 34193284 PMCID: PMC8247177 DOI: 10.1186/s40662-021-00249-z
Source DB: PubMed Journal: Eye Vis (Lond) ISSN: 2326-0254
Fig. 1Rebound tonometers of IcarePRO and IC200. a Overview and b display of IcarePRO. c Overview and d display of the new rebound tonometer IC200
Patient demographics
| Variable | Value |
|---|---|
| Eyes, R/L | 122/23 |
| Subjects, female/male | 82/63 |
| Age, mean ± SD (range), years | 65.6 ± 11.6 (35 to 86) |
| Axial length, mean ± SD (range), mm | 25.1 ± 1.8 (20.2 to 30.9) |
| Corneal curvature, mean ± SD (range), mm | 7.61 ± 0.29 (6.89 to 8.45) |
| Central corneal thickness, mean ± SD (range), mm | 0.522 ± 0.030 (0.412 to 0.606) |
SD standard deviation
IOP value measured with each tonometer in the sitting and supine positions
| Tonometer/position | Mean IOP, mean ± SD (range), mmHg |
|---|---|
| Sitting | |
| GAT | 14.5 ± 2.9 (5 to 22) |
| IcarePRO | 13.3 ± 3.2 (6.3 to 23.6)* |
| IC200-single | 11.6 ± 3.2 (3.6 to 21.2)*† |
| I200-continuous | 11.5 ± 3.2 (6.5 to 25.3)*† |
| Supine | |
| IcarePRO | 14.5 ± 3.8 (6.5 to 25.3) |
| IC200-single | 14.2 ± 4.4 (3.2 to 26.7) |
| IC200-continuous | 14.3 ± 4.5 (3.4 to 29.5) |
Tukey’s multiple-comparison test
GAT Goldmann applanation tonometry; IOP intraocular pressure; SD standard deviation
*P < 0.05 as compared with GAT
†P < 0.05 as compared with IcarePRO
Intraclass coefficient and coefficient of variance of each intraocular pressure in the sitting and supine positions
| Tonometer/position | ICC (95% CI) | CoV (%) |
|---|---|---|
| Sitting | ||
| GAT | 0.984 (0.979–0.988) | 1.6 ± 2.2 (0.0–7.1) |
| IcarePRO | 0.945 (0.929–0.959) | 5.0 ± 3.2 (0.0–19.4)* |
| IC200-single | 0.967 (0.957–0.975) | 4.4 ± 2.2 (0.0–16.9)* |
| IC200-continuous | 0.971 (0.962–0.978) | 4.1 ± 2.8 (0.0–14.9)*† |
| Supine | ||
| IcarePRO | 0.963 (0.952–0.973) | 4.6 ± 3.1 (0.0–17.2) |
| IC200-single | 0.987 (0.984–0.991) | 3.9 ± 2.2 (0.0–10.5)† |
| IC200-continuous | 0.984 (0.979–0.984) | 3.5 ± 2.3 (0.0–13.7)† |
Tukey’s multiple-comparison test
CI confidence interval; CoV coefficient of variance; ICC intraclass correlation coefficient
*P < 0.05 as compared with GAT
†P < 0.05 as compared with IcarePRO
Fig. 2Bland–Altman plots of the measured IOPs between GAT and IcarePRO, between GAT and IC200-single, and between GAT and IC200-continuous at the sitting position. a GAT and IcarePRO, b GAT and IC200-single, c GAT and IC200-continuous. GAT, Goldmann applanation tonometry; LOA, limit of agreement; SD, standard deviation
Fig. 3IOP elevation of IcarePRO, IC200-single, and IC200-continuous. The elevation of IOP was significantly different between IcarePRO and IC200. IcarePRO has less IOP elevation. *Significant pair by Tukey’s multiple comparison test
Fig. 4IOP elevation of each tonometer with and without bleb. The elevation of IOP was significantly different with and without bleb in IC200 but not in IcarePRO. *Significant pair by nonpaired Wilcoxon signed-rank test
Effects of age, axial length, corneal curvature, and central corneal thickness on IOP measurement with each tonometera
| Parameter | GAT | IcarePRO | IC200-single | IC200-continuous | ||||
|---|---|---|---|---|---|---|---|---|
| Coefficient | Coefficient | Coefficient | Coefficient | |||||
| Sitting | ||||||||
| Age | 0.022 | 0.34 | −0.022 | 0.39 | −0.0028 | 0.91 | 0.029 | 0.91 |
| Central corneal thickness | ||||||||
| Corneal curvature | 0.29 | 0.73 | 0.15 | 0.87 | 1.10 | 0.19 | 1.21 | 0.19 |
| Axial length | 0.0031 | 0.98 | 0.20 | 0.21 | 0.13 | 0.42 | 0.13 | 0.41 |
| Supine | ||||||||
| Age | – | – | 0.016 | 0.60 | 0.0011 | 0.97 | 0.015 | 0.67 |
| Central corneal thickness | – | – | ||||||
| Corneal curvature | – | – | −0.29 | 0.79 | 1.21 | 0.36 | 1.38 | 0.30 |
| Axial length | – | – | 0.28 | 0.15 | 0.072 | 0.75 | 0.16 | 0.48 |
IOP intraocular pressure
aBoldface type indicates P < 0.05