Effect of Manual Upper Eyelid Elevation on Intraocular Pressure Measurement by Four Different Tonometers.

Shunsuke Nakakura, Etsuko Mori1, Yuki Fujio1, Yasuko Fujisawa1, Kanae Matsuya1, Yui Kobayashi1, Satomi Oogi1, Miku Shiraishi1, Hitoshi Tabuchi1, Ryo Asaoka2, Yoshiaki Kiuchi3.   

Abstract

SIGNIFICANCE: This study is the first to show that the manual upper eyelid elevation (manual UEE) that is commonly used to prevent disruption of the IOP measurement due to blinking or upper eyelid contact with the tip of the tonometer does not affect the IOP values.
PURPOSE: We investigated whether manual UEE affects the IOP readings using three rebound tonometers (Icare TA01i, Icare PRO, and Icare ic100) and Goldmann applanation tonometry (GAT).
METHODS: One eye was measured for 101 patients (56 eyes of primary open-angle glaucoma patients and 45 healthy subjects). The IOPs were measured without and with manual UEE. Each IOP was measured twice; the measurement order using the tonometers was randomly selected. In addition, palpebral fissure height (distance between the upper and lower eyelids) was measured.
RESULTS: The IOPs without manual UEE were 12.1 ± 2.9, 13.3 ± 2.7, 11.7 ± 2.9, and 16.0 ± 3.2 mmHg (Icare TA01i, Icare PRO, Icare ic100, and GAT), and those with manual UEE were 12.3 ± 3.0, 13.3 ± 2.8, 11.7 ± 2.9, and 16.0 ± 3.3, respectively. No significant difference was found between the IOP without and with manual UEE (IOP difference; all, P > .50; paired t test). Multiple linear regression analyses revealed that palpebral fissure height did not affect IOP difference for any of the tonometers.
CONCLUSIONS: Simple manual UEE when measuring the IOP has little effect on the IOP obtained using all current rebound tonometers and GAT.

Entities:  

Mesh:

Year:  2020        PMID: 32011586     DOI: 10.1097/OPX.0000000000001472

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  1 in total

1.  Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma.

Authors:  Shunsuke Nakakura; Ryo Asaoka; Etsuko Terao; Yuki Nagata; Yasuko Fukuma; Satomi Oogi; Miku Shiraishi; Yoshiaki Kiuchi
Journal:  Eye Vis (Lond)       Date:  2021-07-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.