| Literature DB >> 34150714 |
Abiola Adesina1, Philani Mashazi1,2.
Abstract
The detection and monitoring of biological markers as disease indicators in a simple manner is a subject of international interest. In this work, we report two simple and sensitive label-free impedimetric immunoassays for the detection ofEntities:
Keywords: C-reactive protein; boronate ester; immunosensor; impedance immunoassay; oriented antibody
Year: 2021 PMID: 34150714 PMCID: PMC8207519 DOI: 10.3389/fchem.2021.587142
Source DB: PubMed Journal: Front Chem ISSN: 2296-2646 Impact factor: 5.221
SCHEME 1Step-by-step modification of gold surface to form MBA-APBA SAM and the immobilization of monoclonal antihuman CRP antibody.
FIGURE 1(A) Cyclic voltammograms and (B) Nyquist plot of (i) bare Au, (ii) Au-MBA-APBA SAM, and (iii) Au-MBA-APBA-mAb/glucose in (1:1) 2 mM K3Fe(CN)6: K4Fe(CN)6 solution containing 0.10 M KCl.
FIGURE 2The survey spectra of (A) bare Au, (B) Au-MBA SAM, (C) Au-MBA-APBA SAM, and (D) Au-MBA-APBA-mAb/glucose.
FIGURE 3High-resolution spectra of C 1s (A) Au-MBA SAM, (B) Au-MBA-APBA SAM, and (C) Au-MBA-APBA-mAb/glucose and O 1s (A’) Au-MBA SAM, (B’) Au-MBA-APBA SAM, and (C’) Au-MBA-APBA-mAb/glucose.
FIGURE 4High-resolution spectra of N 1s for (A) Au-MBA-APBA SAM and (B) Au-MBA-APBA-mAb/glucose and (C) the high-resolution spectrum of B 1s of Au-MBA-APBA SAM.
FIGURE 5(A,B) Nyquist plots and (A’,B’) the corresponding calibration curves at varied CRP antigen concentrations (i) 10 ng ml−1, (ii) 25 ng ml−1, (iii) 50 ng ml−1, (iv) 75 ng ml−1, and (v) 100 ng ml−1 for the (A) direct and (B) sandwich immunoassays. (n = 3, SD ≤ 9.0%).
Analytical parameters for the electrochemical immunosensor for CRP antigen detection compared with reported methods.
| Detection method | Assay linear range (ng ml−1) | Sens. (kΩ.ng−1. mL.cm−2) |
| LoD (ng ml−1) | LoQ (ng ml−1) | Ref |
|---|---|---|---|---|---|---|
| Impedance spectroscopy | 10–100 | 0.585 | 0.998 | 2.90 | 9.66 | This work |
| 10–100 | 0.885 | 0.999 | 1.20 | 3.97 | ||
| Electrochemistry | 50–5,000 | — | — | 11.0 | — |
|
| Electrochemistry | 8.5–9,120 | — | — | 3.50 | — |
|
| Electrochemistry | 1.56–400 | — | — | 0.46 | — |
|
| Electrochemistry | 0–100 | — | — | 2.2 | — |
|
| Electrochemistry | 0.5–200 | — | — | 1.2 | — |
|
| Electrochemistry | 0.2–80 | — | — | 0.040 | — |
|
is the measurement using the direct immunoassay.
is the measurement using sandwich immunoassays.
Spike and recovery results obtained from the fabricated CRP immunosensor in serum samples.
| Sample number | Spiked (ng ml−1) | Found (ng ml−1) | Recovery (%) |
|---|---|---|---|
| 1 | 25 (direct) | 27.30 | 95.87 |
| 2 | 25 (sandwich) | 26.50 | 96.47 |
| 3 | 50 (direct) | 49.90 | 94.94 |
| 4 | 50 (sandwich) | 49.00 | 91.18 |