F Riedel1, J Golda2,3, J Held3, H L Davies1,4, M W van der Woude4, J Bredin1, K Niemi1,5, T Gans1, V Schulz-von der Gathen3, D O'Connell1. 1. York Plasma Institute, Department of Physics, University of York, York YO10 5DD, United Kingdom. 2. Institute of Experimental and Applied Physics, Kiel University, 24098 Kiel, Germany. 3. Experimental Physics II, Ruhr-Universität Bochum, 44801 Bochum, Germany. 4. York Biomedical Research Institute, Hull York Medical School, University of York, York YO10 5DD, United Kingdom. 5. kari.niemi@york.ac.uk.
Abstract
Atmospheric pressure plasmas have been ground-breaking for plasma science and technologies, due to their significant application potential in many fields, including medicinal, biological, and environmental applications. This is predominantly due to their efficient production and delivery of chemically reactive species under ambient conditions. One of the challenges in progressing the field is comparing plasma sources and results across the community and the literature. To address this a reference plasma source was established during the 'biomedical applications of atmospheric pressure plasmas' EU COST Action MP1101. It is crucial that reference sources are reproducible. Here, we present the reproducibility and variance across multiple sources through examining various characteristics, including: absolute atomic oxygen densities, absolute ozone densities, electrical characteristics, optical emission spectroscopy, temperature measurements, and bactericidal activity. The measurements demonstrate that the tested COST jets are mainly reproducible within the intrinsic uncertainty of each measurement technique.
Atmospheric pressure plasmas have been ground-breaking for plasma science and technologies, due to their significant application potential in many fields, including medicinal, biological, and environmental applications. This is predominantly due to their efficient production and delivery of chemically reactive species under ambient conditions. One of the challenges in progressing the field is comparing plasma sources and results across the community and the literature. To address this a reference plasma source was established during the 'biomedical applications of atmospheric pressure plasmas' EU COST Action MP1101. It is crucial that reference sources are reproducible. Here, we present the reproducibility and variance across multiple sources through examining various characteristics, including: absolute atomic oxygen densities, absolute ozone densities, electrical characteristics, optical emission spectroscopy, temperature measurements, and bactericidal activity. The measurements demonstrate that the tested COST jets are mainly reproducible within the intrinsic uncertainty of each measurement technique.
Atmospheric pressure plasma jets have attracted significant interest due to their
application potential, such as in potential cancer treatments and wound healing
[1-15], plasma chemical [16-18] and material synthesis [19, 20], and surface
modifications like thin film deposition [21, 22], etching of
photoresist [23], and
pre-treatment of plastic surfaces [24]. The main motivation of these plasma sources for technological
applications stems from their ability to generate and deliver reactive atomic and
molecular species (both long- and short-lived), along with other active components
such as UV, charged particles, and electric fields, under ambient conditions to a
target. Low temperature plasmas can stimulate specific biological responses, this is
not only, but at least significantly due to the fact that low temperature plasma
generated reactive oxygennitrogen species (RONS) are the same as the RONS produced
endogenously in the human body [25, 26]. These
mediate many physiological processes, such as cell-cell signaling, immune response,
wound healing and cell death processes, so therefore the plasma produced species are
expected to mimic the functions of their native counterparts. Some of these key
reactive species include atomic oxygen and nitrogen, hydroxyl radicals, nitric
oxide, singlet oxygen, ozone and hydrogen peroxide. The role of other plasma
components, for example electric fields and UV, and more importantly the synergies
between these components are also recognised as key to exploit in order to develop
effective therapeutics. In fact, both RONS and electric fields individually are
already known to play vital roles in existing therapeutics [27-29], and the ability of plasmas to directly generate and simultaneously
deliver these offers significant advantages and potential. In this context, it is
crucial to elucidate mechanisms, and to that end efficient and accurate transfer of
knowledge across the community is critical in order to accelerate the pace of
research.Due to the diverse application potential, the rapidly developing field, and
technological need many atmospheric pressure plasma sources have been developed
world-wide for application and fundamental science. These have proven to be
efficient and successful for various means, however, what remains a significant
challenge is the comparison across plasma sources. This results in inefficient
scientific progress, as each research team either needs to conduct lengthy
characterisation of each source, or has limited access to complex, expensive
diagnostic and/or simulation capabilities to do so. This can, therefore, result in a
lot of redundancy of research, but more critically, without correlation of plasma
parameters, causation mechanisms of plasma-induced biological processes is extremely
difficult—if not impossible.To help overcome these difficulties, within the European Cooperation for Science and
Technology (COST) Action MP1101 ‘biomedical applications of atmospheric pressure
plasmas’ [30], a reference plasma
source for atmospheric pressure plasmas, the COST reference microplasma jet, or the
so-called ‘COST jet’ was introduced [31]. The aim of such a source is to have a well-characterised plasma
where the literature and knowledge from various teams can assist with elucidating
plasma and application mechanisms. Until this little effort was made to establish a
reference atmospheric pressure plasma jet with a freely available design [32], as was conducted for low
pressure plasmas with the gaseous electronics conference RF reference cell [33, 34]The aim of the study presented here is to establish the variability between COST
plasma jets, from the same manufacturing batch, and the reproducibility of each
individual source. The motivation for this is two-fold: a reference source is only
as good as how comparable it is to other sources, and how reproducible it is to
itself. This should also serve as an aid for plasma source developers to help better
understand the origin of variability in various plasma parameters.
Background
The COST jet is driven with a radio frequency waveform, capacitively coupled with
parallel stainless steel electrodes, sandwiched between glass panes to confine the
gas flow between the electrodes. It is typically operated with a noble gas flow e.g.
helium and small admixtures of molecular gases such as oxygen, nitrogen or water
vapour. The geometry has been designed to provide good optical diagnostic access to
the plasma core [35-38] as well as the jet/effluent
region, and its geometry is also well suited for simulation and modelling of the
plasma [39-43]. The 30 mm plasma channel ensures, that for
typical operating gas flows in the order of slm, the various chemical species have
evolved to steady state well before the nozzle [36, 44, 45, 46].The cross-field geometry configuration of the plasma is such that the plasma jet, or
effluent, consists of neutral species, and UV radiation can also escape and be
transported to the target [47].
Since the electric field between the electrodes is perpendicular to the gas flow,
the charged particles and electric field rapidly decay outside the core plasma,
therefore leaving the jet region devoid of charged particles and electric fields
[48]. This means that the
resultant plasma jet, or effluent, are not susceptible to typical surrounding
ambient electrostatics and electrodynamics, while in comparison plasma jets
containing charged particles and exhibiting relatively high electric fields, can be
very susceptible to surrounding ground and target type [49-53].There has been a significant body of research carried out on the COST jet, and its
predecessors, including diagnostics, simulations [54] and modelling [55-57], on the electron dynamics, plasma sustainment mechanisms [58-67], and chemical kinetics [45, 46, 68], including
reactive species production in the bulk plasma and jet regions [69, 70]. The plasma has been applied for various
applications and additionally advanced tailoring concepts [71-79] are been developed and employed for improved control over the
reactive species generation and treatments. Improvements on efficient power delivery
and electrical characterisation have also been performed [80-82]. These detailed characterisations and investigations can help inform
future research on the COST jets and also adapted more generally for other
atmospheric pressure plasma sources.In order to control the production of reactive species within the plasma, molecules
are purposely admixed to the feed gas. This provides improved control over reactive
species generation, compared with relying on ambient oxygen and nitrogen molecules
[45]. Knake
et al [83,
84] reported that the atomic
oxygen production is most efficient with an admixture of 0.5%–0.6% molecular oxygen.
This motivates the molecular admixtures used in the presented work. Various reactive
species have been measured and simulated using different techniques including atomic
oxygen [38, 85–88], atomic hydrogen [46], hydroxyl [68,
89], singlet delta oxygen
[90], ozone [35], hydrogen peroxide [91], atomic nitrogen [37, 86], nitric oxide [92], helium metastables [93-95]. These reactive species can propagate varying distances beyond the
plasma nozzle, with some surviving up to several centimetres [96]. The role of synergies between UV and reactive
species, compared with their individual influence, has been investigated using an
extended X-jet configuration and identified as important [47, 97–100]. In general
for plasma treatments, the flux of both these components is important to consider,
as is the heat and gas dynamics impacting on surfaces [96].Gorbanev et al determined the origin of species in plasma treated
liquids and found that most reactive species detected in the liquid phase originated
in the plasma gas phase and were subsequently transported into the liquid [101]. While Hefny
et al highlight the important role of solvated O atoms in
aqueous solutions [102].
Investigations have also started to elucidate mechanisms in physiological solutions
[103-106]. Treatment of different biological systems
have been conducted for efficacy purposes, but also to clarify mechanisms of plasma
action. Treatments on different cancer cell types [4, 107–109], different
DNA origami nanostructures [100,
110, 111], and antibacterial action, including
resistance mechanisms [112] have
been investigated. The COST jet has also been applied for various other applications
[113], including the
development of new chemical and material synthesis and processing such as
epoxidation [16], etching [23], thin film deposition [22], and surface modifications
[24].The investigation presented here focusses on quantifying the repeatability across
different COST reference microplasma jets. Therefore the plasmas, produced by four
devices, are compared for different parameters. These include power characteristics,
gas and substrate surface temperature, optical emission spectroscopy (OES), ozone
densities, atomic oxygen densities, and bactericidal activity.
Setup and diagnostics
Plasma source
In this study four identically constructed and equipped exemplars of the ‘COST
reference microplasma jet’, as specified in [31], were investigated. Each device consists of
(a) the head, which includes the stainless steel electrode assembly between two
quartz glass windows forming a plasma channel of 30 mm length and 1 mm × 1 mm
cross section, and (b) the housing, which comprises the LC resonance based
radio-frequency power coupling circuit [80], a capacitively coupled voltage probe, and
a resistive current probe. Manufacturing tolerances of the COST jet devices are
stated together with the results of the rf power measurements in section 4.1.A commercial 13.56 MHz radio-frequency generator and external manual matching
network unit (coaxial power systems, RFG50 and MMN150) were used to operate the
COST jet devices, with a 50 Ω BNC coaxial cable of 0.5 m length between matching
unit and jet. The length of the feed gas tube that is exposed to moist ambient
air whenever the jet is not in operation was kept as short as possible, in order
to decrease the time for the jet to reach steady-state operation. The feed gas
for all later experiments was chosen as 1 slm helium flux with 0.5% oxygen
admixture (purity grade N4.6 for helium and N5.0 for oxygen). This ensures a
high flux of generated reactive species to the sample, while keeping evaporation
of wet biological samples tolerable [37, 84].In order to ensure a valid comparison, all four COST reference microplasma jets
were initially cleaned with standard solvent in an ultrasonic bath for the
presented investigation, because each of them had an unknown prior history, e.g.
time and conditions of operation, at different universities and institutes.It is known that the performance of an rf plasma can strongly depend on the rf
components used to power the source. The internal voltage and current probes in
the COST jet design are located as close as possible to the electrodes. This
should allow a rf power measurement independent of the used power equipment, if
the required impedance matching to the internal LC circuit can be achieved. In
order to exclude any uncertainty for the comparison, each jet was operated with
the exact same equipment: radio-frequency generator, matching network, coaxial
cables and connectors, and gas mass flow controllers. The same applies for all
measuring equipment, e.g. digital oscilloscope, spectrometer, ozone monitor,
lasers, external electrical probes.Prior to any experiment, the gas lines were flushed for 30 min, then the plasma
jet was ignited and run for a 30 min warm-up duration. After any change of the
operational parameters, externally applied rf power or gas flow, a stabilisation
time of 10 min was observed before conducting the next measurement. The
laboratory conditions were controlled to within 22 ± 0.5 °C room temperature and
50 ± 10% relative humidity for all measurements.
Measurements
Dissipated electrical power. The COST-jet incorporates
miniaturised electrical probes inside its housing, i.e. a capacitively coupled
voltage probe and a resistive current probe, which allow a precise measurement
of the actual electrical power dissipated in the plasma [82]. The output of both probes was
simultaneously recorded by an oscilloscope (Agilent Technologies, Infiniivision
DSO-X 2004A, 8 bit, 2 GSps, 70 MHz) using 50 Ω coaxial cables (Thorlabs, CA2612)
of equal type and length, as an average over 1024 consecutive recordings. The
data was sent to a computer where the deposited power was calculated according
towith Urms and
Irms the effective values of voltage and
current, ϕ the phase shift between voltage and current, and
ϕref the instrumental reference phase shift
determined from measurements without plasma. Since the internal probes are
located directly at the electrodes, their readings do not need to be corrected
for parasitic power consumption occurring elsewhere in the circuit (not in the
plasma), as e.g. in [63,
114]. A detailed error
analysis of the power measurement method can be found in [82]. The internal voltage probe of each COST
jet needs to be calibrated [31], e.g. here by using a commercial external voltage probe
(Tektronix, P5100A, 1000 Vrms, 500 MHz).The described measuring technique for sinusoidal waveforms is valid as long as
the COST jet is operated in stable homogeneous α-glow mode,
which, as shown in [61, 63], does not exhibit any
constricted nanosecond sparks or streamers as in dielectric barrier
discharges.Effluent gas temperature. The gas temperature of the jet’s
effluent as a crucial parameter for the treatment of biomedical or heat
sensitive samples was measured with a K-type thermocouple that was placed 3 mm
in front of the jet’s nozzle. No evidence was found that the thermocouple
measurement was influenced by the radio-frequency electro-magnetic radiation
from the electrode assembly. Each temperature measurement as an average over 5
min, taken alongside the electrical power measurement, results in a mean value
and a standard deviation, which reflects changes like airflow and room
temperature inside the lab.Surface temperature. The spatially resolved surface temperature
on an artificial sample placed at various distances from the jet’s nozzle was
measured with a thermal camera (Agilent, U5855A). The chosen sample is a
standard microscope slide out of chemically inert quartz glass. The sample’s
front surface was roughened by shot blasting to minimise direct reflections from
other heat sources. The thermal emission coefficient of this surface was
measured with respect to the known emission coefficient of a black infrared
sticker.Optical plasma emission. The optical emission from the centre of
the plasma channel is measured with a fibre coupled spectrometer (ocean optics,
HR4000CG and QP600-2-SR-BX) using the reference fibre spacer of the COST jet as
alignment tool for positioning the fibre tip, see [31]. The fixed-configuration spectrometer
covers the spectral range from 200 nm to 1100 nm with a spectral resolution of
about 0.5 nm. We focus on measuring intensity ratios of the dominant atomic
emission lines, He(706 nm), O(777 nm), and O(844 nm), for comparing the four
different COST jets.Ozone densities. The ozone density in the far-effluent of the
COST jet was measured with a commercial ozone gas detection monitor (2B
Technologies, model 106-L) based on 254 nm UV absorption. The gas output of the
jet was sucked into the ozone monitor via a glass funnel close to the jet’s
nozzle and through a 1 m long PFA plastic tube by the internal pump of the
monitor at nominal flow rate of 1 slm.Atomic oxygen density. Atomic oxygen ground state densities in
the near effluent of the COST jet, here at 1 mm distance from the nozzle and
when operated with standard feed gas of 1 slm helium with 0.5% oxygen admixture,
were measured by means of two-photon absorption laser induced fluorescence
(TALIF). An absolute calibration was carried using the technique detailed in
Niemi et al [115, 116]. In
this case, an evacuated gas cell was filled with a defined pressure of xenon,
with a similar TALIF scheme to atomic oxygen. The challenging aspect of this
established method is determination of the effective lifetime of the laser
excited atomic oxygen state. This lifetime is typically in the order of
picoseconds to a few nanoseconds due to significant and inhomogeneous
collisional quenching within the effluent of the COST jet penetrating into
ambient air.Two different spectrally widely tunable OPO/OPA laser systems with inbuilt
frequency doubling and mixing stages were used in this study. For all presented
TALIF measurements, the fluorescence is detected perpendicular to the laser
beam, but in the same horizontal plane, while the COST jets were mounted
upright, with the effluent directed towards a fume extraction hood.A pico-second laser system (Ekspla, PL2251B, APL2100, and PG411), able to provide
up to 300 μJ energy within 30 ps pulse duration with a
repetition rate of 10 Hz at the required wavelengths around 225 nm, was used as
an excitation for the TALIF schemes. The UV output beam was intentionally
attenuated and focussed with an f = 30 cm silica lens about 3
cm behind the COST jet effluent, to keep the local power density below the onset
of various signal saturation effects. The fluorescence volume was imaged about
1:1 by a doublet of achromatic lenses (1 inch diameter,
ftot = 40 mm) through interference filters of 10
nm bandwidth onto the chip of an intensified charge-coupled device camera
(Stanford computer optics, 4-Picos with S25IR photo-cathode and 780 × 580 pixels
of 8.3 μm square size). This laser system in combination with
the camera’s minimal gate width of 200 ps was used to measure the exponential
decay of the fluorescence radiation from the laser excited states in the COST
jet effluent and in the reference gas cell under low pressure, respectively. Our
measurements result in an effective
O(3p3P)
state lifetime of 4.24 ± 0.07 ns, which is about 8 times shorter than the
corresponding natural lifetime of 35.1 ns.A second TALIF setup includes a more conventional nanosecond OPO/OPA laser system
(Continuum, Surelite EX and Horizon OPO) providing 225 nm pulses of up to 5 mJ
energy in about 4 ns duration at the same repetition rate of 10 Hz, and a
different ICCD camera (Andor, IStar with-73 photocathode and 1024 × 1024 pixels
of 13 μm square size) with a longer minimal gate width of 2
ns, in an otherwise similar detection setup. This second setup, since it was
offering a lower pulse-to-pulse laser energy fluctuation than the first, was
used to measure the temporally and spectrally integrated TALIF signals, for
atomic oxygen in the effluent and for xenon in the reference gas cell, from
which the stated atomic oxygen density values were derived.Bactericidal assay. The bactericidal assay was carried out as
described previously in [10].
Briefly, single colonies of non-pathogenic Escherichia coli
K-12, MG1655, were cultured in Luria–Bertani broth (LB, 10 g L−1),
until the late logarithmic growth phase. The optical density (OD) of the
bacteria was then adjusted to OD = 0.02 at 600 nm. Approximately 8 ×
105 colony forming units (cfu) were transferred to LB agar petri
dishes (LB, supplemented with 17.5 g L−1 agarose), and spread using
glass beads to ensure even coverage of the plates with E. coli.
The bacterial plates were allowed to dry.For plasma treatment, the COST jets were operated in downwards orientation inside
a reasonably large Perspex plastic box with constant fume extraction at the top.
The box shields the experiment from changing air movements within the lab, while
the fume extraction prevents a build-up of reactive species inside the box.
Before the 2 min long plasma treatment, the lid of the petri dish was removed,
plate was placed so that the top of the agar was at a distance 5 mm below the
plasma jet nozzle. Afterwards, the plate was immediately removed, the lid put
back on to prevent contamination, and the plate returned into the incubator (at
37 °C) for overnight. On the following day, the plates were imaged using a
scanner (Epson, Perfection V750 Pro) to measure the area of inhibition (AOI) and
to count the number of survivor colonies using the ImageJ
software [117].Care was taken to follow the exact same experimental protocol: autoclaved LB agar
was always allowed to cool-down to 55 °C in a water bath, then 20 ml was poured
into each petri dish, and the plates were allowed to dry in a laminar flow hood
for the same duration. Also, all bacterial cultures were derived from single
colonies, cultured over same duration and under same conditions, and the
bacterial concentration was kept constant for all experiments.
Results
Power characteristics
Figure 1(a) shows a measured
characteristic of dissipated electrical power versus effective voltage as
obtained from a subsequent forward/backward sweep of the external generator
power. The black dots are the results obtained from one of the COST jets. The
standard deviation found among the four COST jets is shown as coloured areas.
Blue indicates the forward sweep, and red the backwards sweep, while the overlap
is in purple. Hysteresis effects are visible, e.g. the plasma ignites into a
stable homogeneous operational mode at about 190
Vrms, which can be sustained down to about 160
Vrms after ignition. At a voltage of about 355
Vrms, the jump to higher power and lower voltage
marks the spontaneous transition into an operational mode with one constricted
filament between the electrodes, which is clearly distinguishable by naked eye
observation. This mode is not desired as the filament is unstable and the power
is constricted to hot-spots on the electrode surfaces leading to damage as well
as excessive gas heating on short time scales. For extinguishing the filament,
the voltage/power needs to be reduced well below the onset of this mode
transition. The four COST jets showed a stable homogeneous operation from 180 to
350 Vrms and corresponding plasma power from 0.18 W
to 2.5 W. The relative standard deviation among the four COST jets in terms of
plasma power increase with effective voltage, but stays below 15% as indicated
by the blue and purple overlap area in figure 1(a), when the constricted mode is avoided. A
detailed error analysis of electrical power measurements on COST jet devices has
been presented in [82]. Our
investigation falls into the second of the three different scenarios that were
considered in this reference, i.e. comparing different devices but using the
same rf power equipment, resulting in a predicted overall relative error of
about 10%. The observed deviation of 15% between the investigated four COST jet
devices is slightly larger than the predicted uncertainty of 10% for the
electrical power measurement. The cause most likely reflects the result of small
manufacturing tolerances for the electrode gap distance and alignment, which lie
within ±0.1 mm as measured with a microscope, and the actual electrode surface
conditions prone to physical roughness and surface coverage (humidity and
oxides), which depend on previous operational conditions (feed gases), or even
possible damage from operation in filamentary mode.
Figure 1.
(a) Measured characteristics of plasma power versus effective voltage,
and (b) measured effluent gas temperature at 3 mm distance from the
nozzle versus plasma power, both for standard feed gas of 1 slm He with
0.5% oxygen admixture. Black dots indicate the result obtained from one
of the COST jets. The coloured areas are the standard deviation between
the four COST jets, explanation provided in text. The red horizontal
line in (b) indicates the maximum permissible temperature for regular
treatment of biological samples.
(a) Measured characteristics of plasma power versus effective voltage,
and (b) measured effluent gas temperature at 3 mm distance from the
nozzle versus plasma power, both for standard feed gas of 1 slm He with
0.5% oxygen admixture. Black dots indicate the result obtained from one
of the COST jets. The coloured areas are the standard deviation between
the four COST jets, explanation provided in text. The red horizontal
line in (b) indicates the maximum permissible temperature for regular
treatment of biological samples.
Effluent gas temperature
Figure 1(b) shows the measured gas
temperature of the free flowing effluent at 3 mm distance from the nozzle as a
function of the plasma power for the jets investigated in this study. The blue
area indicates the standard deviation among the four COST jets, staying below 3%
with respect to room temperature over the whole operational range. The
corresponding error margins in y-axis stay within the indicated
x-axis error margins for the derived plasma power. The
horizontal red line indicates the maximum permissible temperature of 37 °C for
regular treatment of biological samples, which implies that the plasma power
should be kept below 0.3 W for this particular distance.
Surface temperature
Figure 2 (left) shows an example
thermal image taken by the thermal camera, and (right) the schematic of the
measuring setup. The COST jet is mounted horizontally and pointing towards the
microscope slide. The thermal camera is mounted on a tripod at a distance of
about 20 cm and at an angle of about 28° with respect to the shot-blasted front
surface of the microscope quartz slide. As expected, the thermal image shows
that the spatial profile of the surface temperature exhibits a central maximum
at the location where the jet axis hits the surface.
Figure 2.
(Left) thermal image taken for 16 mm distance between nozzle and sample
surface, standard feed gas (1 slm He with 0.5% oxygen admixture), and
reduced plasma power (0.3 W). (Right) schematic of the corresponding
experimental setup.
(Left) thermal image taken for 16 mm distance between nozzle and sample
surface, standard feed gas (1 slm He with 0.5% oxygen admixture), and
reduced plasma power (0.3 W). (Right) schematic of the corresponding
experimental setup.Figure 3(a) shows the measured
maximum central surface temperature as a function of the distance between jet
nozzle and sample surface, for a plasma power of 0.300 ± 0.015 W. At each
distance, three thermal images of the sample surface are taken for each COST
jet, respectively. The black dots represent the mean values, and the blue area
the standard deviation among the four COST jets. The measurement shows that the
maximum surface temperature is decreasing only insignificantly, from about 36 °C
to about one degree less, over a distance of up to 30 mm distance from the jet
nozzle. The standard deviation of 1 °C surface temperature is the uncertainty of
the thermal imaging measurement, given air movements in the lab, from which we
conclude that the four COST jets are fully comparable in this respect. In
addition, the surface temperature and the gas temperature in the free flowing
effluent measured at the same plasma power agree within the 1 °C
uncertainty.
Figure 3.
(a) Measured maximum central surface temperature as a function of the
distance between the jet nozzle and sample. (b) Measured lateral surface
temperature profiles at different nozzle to sample distances. Both for
standard feed gas (1 slm He with 0.5% oxygen admixture and at reduced
plasma power (0.3 W).
(a) Measured maximum central surface temperature as a function of the
distance between the jet nozzle and sample. (b) Measured lateral surface
temperature profiles at different nozzle to sample distances. Both for
standard feed gas (1 slm He with 0.5% oxygen admixture and at reduced
plasma power (0.3 W).Figure 3(b) shows the measured
lateral surface temperature profiles. The black profile with grey margins
represents the mean values and corresponding standard deviations among the four
COST jets from a measurement at near distance of 4 mm, while the red curve with
rose margin indicate the corresponding quantities from a measurement at far
distance of 31 mm. The measurement shows that the spatial temperature
distribution does not spread significantly with increasing distance up to 31 mm
distance, as expected from the measured marginal decrease of the on-axis surface
temperature. The standard deviation among the four COST jets however increases
with distance, because of the mentioned environmental influences. Our findings
that the free flowing jet effluent in terms of temperature stays constricted
over a distance of 30 mm and produces a lateral sample surface temperature
profile of about 16 mm full-width at half-maximum are supported by the spatially
resolved thermocouple measurements in [83] and Schlieren imaging in [96].
Optical plasma emission
Optical plasma emission is a very useful external diagnostics to monitor plasma
stability and reproducibility. It depends on various internal plasma parameters,
in particular species composition, electron density and electron energy
distribution function. While it is challenging to quantitatively distinguish
between details of the origin of changes, it is highly sensitive to overall
changes.The results of the OES measurements taken in the centre of the plasma channel are
shown in figure 4 as intensity
ratios for the most prominent atomic lines, He(706 nm), O(777 nm) and O(844 nm),
as a function of the dissipated plasma power. The solid squares represent the
mean values and the shaded areas the standard deviations, respectively, found
among the four COST jets.
Figure 4.
Intensity ratios from optical emission measurements as a function of
plasma power for standard feed gas of 1 slm He with 0.5% oxygen
admixture. The considered atomic lines, He 706 nm, O 777 nm, and O 844
nm, are labelled according to their wavelength.
Intensity ratios from optical emission measurements as a function of
plasma power for standard feed gas of 1 slm He with 0.5% oxygen
admixture. The considered atomic lines, He 706 nm, O 777 nm, and O 844
nm, are labelled according to their wavelength.The O(844 nm)/O(777 nm) intensity ratio stays almost constant over the
investigated power range. Both intensity ratios, He(706 nm)/O(777 nm) and He(706
nm)/O(844 nm), show an increase with increasing plasma power. The standard
deviations for the He(706 nm)/O(777 nm) and He(706 nm)/O (844 nm) line ratios
are below 8%. This indicates a very good agreement among the four COST jets,
since this deviation is less significant than the uncertainty in the plasma
power measurement.
Ozone density
Figure 5 displays the measured
ozone density in the far effluent of the COST jet as a function of plasma power
for a standard feed gas of 1 slm helium and 5 sccm oxygen. The black dots
represent the mean values and the shaded area the standard deviation of the four
COST jets. The ozone density increases under-linear with increasing plasma power
from 1 × 1021 m−3 at 0.2 W to 2.6 ×
1021 m−3 at 1 W.
Figure 5.
Measured ozone density in the far effluent of the COST jets as a function
of the plasma power for standard feed gas of 1 slm He and 0.5% oxygen
admixture.
Measured ozone density in the far effluent of the COST jets as a function
of the plasma power for standard feed gas of 1 slm He and 0.5% oxygen
admixture.The formation of ozone in this type of plasma source has previously been
investigated in detail using two-beam UV-LED absorption spectroscopy and
numerical simulations [35].
This revealed that the ozone density inside the bulk plasma source slightly
decreases with increasing power. Here we use a simple ozone monitor in the far
effluent where the plasma-produced atomic oxygen is already converted into
additional ozone [118, 119] through three-body
recombination with molecular oxygen and helium [35].The observed increase in ozone density with increasing power in the far effluent
can be explained by the increased production of atomic oxygen at elevated powers
(see sub-section on atomic oxygen density below). The results for 0.5 W are also
in good agreement with previous measurements in the far effluent [118, 119]. The ozone density is expected to be
lower at shorter distances from the jet nozzle, due to less conversion of O and
O2 into ozone. The deviation of the ozone density between the
jets stays below 3% and is, therefore, less significant than the uncertainty of
the plasma power measurement.
Atomic oxygen density
Figure 6 shows the absolute atomic
oxygen ground state density measured at a distance of 1 mm from the COST jet
nozzle as a function of the plasma power. The black dots represent the mean
values and the shaded area the standard deviation among the four COST jets. The
atomic oxygen density increases under-linear with increasing plasma power from 3
× 1020 m−3 at 0.2 W to 1 × 1021 m−3
at 1 W. Our results are in good agreement with previous findings [70] from molecular mass beam
spectrometry and nanosecond TALIF measurements in the early effluent. The
observed increase of atomic oxygen density with plasma power in the near
effluent is consistent with the measured increase of ozone in the far effluent,
see figure 5, in view of the
efficient chemical conversion of O and O2 into ozone with increasing
reaction time/distance from the nozzle. The standard deviation between the four
COST jets for the atomic oxygen density at the nozzle increases with increasing
plasma power, e.g. smaller than 5% below 0.5 W, while increasing up to 13% at 1
W.
Figure 6.
Measured atomic oxygen density at 1 mm distance from the COST jet nozzle
as a function of the plasma powers for standard feed gas of 1 slm He
with 0.5% oxygen admixture.
Measured atomic oxygen density at 1 mm distance from the COST jet nozzle
as a function of the plasma powers for standard feed gas of 1 slm He
with 0.5% oxygen admixture.It should be noted that the stated absolute atomic oxygen densities are prone to
an additional systematic uncertainty of more than 20% due to the uncertainty of
the two-photon excitation cross section ratio that is required for the TALIF
calibration [120]. This
absolute error is not shown in figure 6, since it does not affect the
relative comparison of the investigated COST jets as
presented here.The atomic oxygen density measured here at 1 mm distance from the nozzle is
expected to drop by not more than 15% over a distance of 5 mm according to the
measurements reported in [70]. 5 mm is the distance at which the agar plates for the following
bacterial activity investigation were located at.
Bactericidal activity
Surface decontamination is one application of atmospheric pressure low
temperature plasmas (LTPs). Due to their low temperatures, thermally sensitive
objects, including skin and other biological tissues can be treated with LTPs to
remove bacterial burdens [15,
121, 122]. As the development of
the COST reference microplasma jet is intended to aid the advancement of the
field of plasma medicine, it is important to understand how similarly the jets
perform to one another, in a biological assay. To do this, a basic bacterial
killing assay was used, to determine (a) the efficacy of bacterial killing by
the COST jet and (b) how similar the killing ability is between different COST
jets. For this, the non-pathogenic E. coli MG1655 strain was
plated onto LB agar plates and subjected to two minute treatment with the COST
jets, using the protocol outlined above. The treatment effects were quantified
in two ways. First, the AOI was measured, and secondly, surviving colonies were
counted to calculate the log reduction of bacteria in the AOI. Experiments were
repeated in triplicate for each jet, and the mean and standard deviation
calculated.The experimental process was to allow the jet to warm up an extrac for 30 min,
carry out treatments, then repeat the process for another jet. As this was a
fairly lengthy process, treatments with the first jet were then repeated to
check that the bacteria had not changed over time due to being left in culture
for longer on the bench. These checks showed that there was no difference in the
AOI or log reduction in bacteria between the first time the jet was used for
treatments, and the second time approximately an hour later (data not shown).
The chosen treatment time was 2 min as this was long enough to give a consistent
sized AOI, without making the treatment process too long.Representative images of treated and control plates are shown in figure 7. Treatment with each of the four
jets gives circular AOI which appear similar across all the jets. The position
of the AOI differs between jets as a result of the each plate not being placed
exactly centrally below the plasma nozzle each time. As well as showing similar
AOI across all jets, the number of survivor colonies across all of the jets also
appears similar in figure 7.
Surviving colonies are thought to occur due to imperfections in the plated
bacterial monolayer resulting in some cells being in multiple layers. As a
result, cells in upper layers could shield bacteria in lower layers, and prevent
their killing by plasma treatment. A gas-only treatment control was also
included to confirm the AOI seen in plasma-treated plates was due to plasma, not
just the gas flow. The representative image of the gas-only control in the lower
left image of figure 7 confirms
that the gas flow does not induce an AOI. As expected, the untreated control in
the lower right image of figure 7
shows an even coverage of bacteria, with no clear regions. As well as acting as
an experimental control, the untreated control plates were also checked to
confirm that the bacterial plating methods were good.
Figure 7.
Representative images of bacterial plates treated by COST jets. 100
μl of E. coli MG1655 at
approximately 8 × 106 cfu ml−1 were plated onto LB
agar plates and exposed to a 2 min treatment by COST jet. The plasma
power was kept at 0.3 W, and the feed gas was 1 slm helium with 0.5%
oxygen admixture. The top panel shows a representative plate for each
jet, plates following 2 min COST jet treatment and overnight incubation.
Representative control plates are shown in the bottom panel. The
gas-only control was also treated for 2 min but the plasma power was
turned off (therefore only the helium/oxygen gas was incident on the
sample), and the untreated control was plated identically to all the
other treatment plates, but did not receive a COST jet treatment. For
each jet, treatments were carried out in triplicate.
Representative images of bacterial plates treated by COST jets. 100
μl of E. coli MG1655 at
approximately 8 × 106 cfu ml−1 were plated onto LB
agar plates and exposed to a 2 min treatment by COST jet. The plasma
power was kept at 0.3 W, and the feed gas was 1 slm helium with 0.5%
oxygen admixture. The top panel shows a representative plate for each
jet, plates following 2 min COST jet treatment and overnight incubation.
Representative control plates are shown in the bottom panel. The
gas-only control was also treated for 2 min but the plasma power was
turned off (therefore only the helium/oxygen gas was incident on the
sample), and the untreated control was plated identically to all the
other treatment plates, but did not receive a COST jet treatment. For
each jet, treatments were carried out in triplicate.To quantify the comparability between jets in the bacterial killing assay, the
AOIs and number of survivor colonies were compared. The mean and standard
deviation of the AOIs across all four jets, following three repeats for each jet
is shown in figure 8 by the red
points, axis and shaded area. The AOIs for each of the jets were very similar,
with means ranging from 5.3–5.7 cm2. The log reduction of bacteria in
the AOIs for each jet were also calculated, and is shown in figure 8 by the blue data points, right
axis and blue shaded area. Similar to the AOIs, the log reduction in bacteria
due to treatment with each jet also appears to be consistent across all the
jets, showing approximately 2.5–3 log reduction by each jet. There is some
variation between jets, however, this variation is generally smaller than the
variation seen within each jet. In general, the four jets have similar abilities
to kill our E. coli model bacteria.
Figure 8.
Figure showing AOI and bacterial log reduction following treatment with
the different COST jets. The red points show the average AOI induced by
each jet, with the error bars showing the standard deviation. The blue
points show the mean log reduction in E. coli MG1655
cfu following treatment with each jet, with the error bars showing
standard deviation. For each jet, treatments were carried out in
triplicate. Conditions were as stated in figure 7.
Figure showing AOI and bacterial log reduction following treatment with
the different COST jets. The red points show the average AOI induced by
each jet, with the error bars showing the standard deviation. The blue
points show the mean log reduction in E. coli MG1655
cfu following treatment with each jet, with the error bars showing
standard deviation. For each jet, treatments were carried out in
triplicate. Conditions were as stated in figure 7.The biological effects induced by LTP treatment are expected to be due to the
synergies between the relatively high fluxes of reactive oxygen and nitrogen
species, and UV delivered to the biological target, as investigated in detail by
Schneider et al and Lackmann et al [97-99] and discussed above. The treatments were
carried out in a perspex box to reduce an extrac any effects of air flow in the
laboratory interfering with the RONS delivery to the treated bacteria. To
prevent excessive build up of long-lived, toxic species, such as
NO and O3,tor fan was attached
to the box, and the NO and O3
concentrations in the box were monitored using commercially available monitors
(2B Technologies: Model 106 L O3 and Model 405 nm
NO2/NO/NO). This monitoring
showed that these species did not increase appreciably over the treatment time,
suggesting that the bacterial killing effects were due to the actual plasma
treatment rather than any build up of species in the box. This is further
confirmed by the definition of the AOI, which suggests local effects are due to
the direct plasma treatment, rather than as a result of the box environment.
Summary and conclusions
In this study, we compared four COST jet devices constructed to the same nominal
specifications, with regard to their actual performance in terms of the internal
dissipated plasma power and the resulting external quantities, like effluent gas
temperature, sample surface temperature, optical plasma emission, ozone output in
the far effluent, generated atomic oxygen density in the near effluent, as well as
their bactericidal activity.The standard deviations found, for each measured external quantity among the four
COST jets, respectively, were below the stated standard deviation of 15% for the
internal plasma power. The uncertainty of the power measurement itself, rather than
actual differences between the individual COST jets, contributes the greatest to
this 15% value, see [82], while
the relevant manufacturing tolerance for the COST jets relates to the electrode gap
distance and alignment to an accuracy of 5%.The effluent gas temperature and the sample surface temperature are critical
parameters for the treatment of heat-sensitive material such as biological tissue.
Both were found to agree well within a narrow standard deviation of about 3%. It was
found that restricting the plasma power to 0.3 W limits the temperature to the
critical value of 37 °C. In that sense the COST jet can be used safely for the
treatment of biological tissue without the necessity of monitoring the temperature
of gas effluent or sample surface.Absolute densities of reactive oxygen species, known to play a key-role in many
surface and biological sample treatment processes, were measured. At the reduced
plasma power of 0.3 W, an atomic oxygen density of about 4 ×
1020 m−3 at 1 mm distance from the jet nozzle and an ozone
density of 1.3 × 1021 m−3 in the far effluent were found. The
contributions of measurement accuracy and difference between the four COST jets to
the observed overall standard deviations, about 10% for the atomic oxygen density
and about 3% for the ozone density, is about equal.A performance study of the four COST jets using the bactericidal assay was conducted.
It was found that the achieved bacterial log reduction differed less between the
individual COST jets, than between different experiments with one COST jet.In conclusion, the COST reference microplasma jet is a simple, inexpensive and robust
plasma source. Results obtained with four exemplar devices show consistently less
than 15% differences, when the internal plasma power is used as the control
parameter. This makes the COST jet design a suitable candidate to act as a reference
source for scientists working in this field to compare their results as directly as
possible.
Authors: Jan-Wilm Lackmann; Simon Schneider; Eugen Edengeiser; Fabian Jarzina; Steffen Brinckmann; Elena Steinborn; Martina Havenith; Jan Benedikt; Julia E Bandow Journal: J R Soc Interface Date: 2013-09-25 Impact factor: 4.118
Authors: Luca G Campana; Erika Kis; Krisztina Bottyán; Antonio Orlando; Francesca de Terlizzi; Grammatiki Mitsala; Rosanna Careri; Pietro Curatolo; Marko Snoj; Gregor Sersa; Sara Valpione; Pietro Quaglino; David Mowatt; Matteo Brizio; Hadrian Schepler Journal: Int J Surg Date: 2019-10-13 Impact factor: 6.071
Authors: Eugen Edengeiser; Jan-Wilm Lackmann; Erik Bründermann; Simon Schneider; Jan Benedikt; Julia E Bandow; Martina Havenith Journal: J Biophotonics Date: 2015-02-06 Impact factor: 3.207
Authors: J-W Lackmann; K Wende; C Verlackt; J Golda; J Volzke; F Kogelheide; J Held; S Bekeschus; A Bogaerts; V Schulz-von der Gathen; K Stapelmann Journal: Sci Rep Date: 2018-05-16 Impact factor: 4.379