Informed consent should include the possibility of allergic skin reactions occurring even when hypo-allergenic oxinium implants are used in total knee arthroplasty.
Introduction
The knee is commonly affected by osteoarthritis and it is estimated that eleven million adults in the United States have symptomatic arthritis. At present about four million patients have had a total knee replacement (TKR) and utilization is projected to grow by 673% in 2030. [1, 2]Young patients between the ages of 45 to 64 years represent the fastest growing age group requiring TKR. [3] This group is usually very active, demanding more out of their joint replacement. The survivorship for conventional TKR in young active patients is considerably less than in the older, more sedentary patient, paving the way for the development of more robust implants [4].Aseptic loosening due to the development of allergic reactions to the metallic implants has been implicated as a reason for revision surgery. In an attempt to reduce this complication, we have seen the evolution of newer, high-performance hypoallergenic knee replacement systems. These new materials promise better durability and function with a lower risk of allergic reactions.There is a notable relationship between metal sensitivity and allergic reactions to implants. Approximately 10% of the general population is patch test positive for metal sensitivity. It is noted that the incidence of metal sensitivity doubles in patients with well-functioning joint replacements, despite the absence of clinical allergic reactions to their implants. Therefore, one can appreciate that the relationship remains relatively obscure.We report a case of a patient with idiopathic allergic dermatitis following hypoallergenic knee replacements. To the best of our knowledge this is the first case of dermatitis occurring after an oxidized zirconium TKR. This case highlights the difficulties in diagnosing metal hypersensitivity reactions and also reinforces the need for further research to better elucidate the relationship between metal hypersensitivity, allergic reactions and implant survivorship.
Case Report
A 55-year-old woman presented with bilateral knee pain for several years secondary to advanced tri-compartmental osteoarthritis, marginally worse on the right side.The patient enjoyed an active social life, had no known allergies, and exercised regularly. Following a discussion of the surgical options, it was decided that the genesis II total knee replacement (TKR) with Oxinium (Smith and Nephew, Memphis, TN) would be most appropriate for her activity levels.We performed the right TKR first, using a midline skin incision and a medial parapatellar arthrotomy. Intramedullary alignment jigs and a measured resection technique were used to prepare the bone and achieve balance in flexion and extension. The bone was cleaned with pulsatile lavage and the components cemented in place, with the tourniquet used only during cementation. The knee was closed in layers and the patient was given three post-operative doses of intravenous cefuroxime.Following an uncomplicated post-operative period at 10 weeks after surgery, the proximal extent of the incision became erythematous, discharging a serous fluid. This was treated expectantly with local dressings, however, the erythematous area became larger and began to spread distally (Fig. 1).
Figure 1
Eczematous skin reaction after the right total knee replacement.
Eczematous skin reaction after the right total knee replacement.Inflammatory markers were within normal limits (C-reactive protein 1.1 mg/L and erythrocyte sedimentation rate 18 mm/h) and clinically the patient demonstrated an excellent range of pain-free knee movement. A consultation with the dermatology unit suggested that this may be allergic dermatitis and recommended a trial of topical corticosteroids. Dermovate (clobetasol propionate 0.05% w/w) was applied to the affected area and within 48 h there was a marked improvement in the appearance of her skin (Fig. 2).
Figure 2
Improvement in the skin following application of a topical corticosteroid.
Improvement in the skin following application of a topical corticosteroid.Six months later, the patient underwent a left genesis II TKR with oxidized zirconium implants using a similar technique as previously described. Twelve weeks after surgery, the proximal aspect of the incision became inflamed with local blistering. On this occasion, dermovate cream was used as first-line treatment, leading to resolution within 24 h.At the annual clinic visit, we observed hyperpigmented spots on the skin surrounding both knees and on her feet and hands (Fig. 3).
Figure 3
Skin changes at annual clinic follow-up.
Skin changes at annual clinic follow-up.The patient has remained asymptomatic, with excellent knee function and radiographs show no signs of loosening (Fig. 4). Skin patch testing with a commercially available kit for the metals used in the knee replacement was negative and at 18 months from her first TKR, the patient remains satisfied with the clinical outcome. We continue to follow her up with annual clinical and radiological assessments.
Figure 4
Anterior-posterior radiograph at 1 year post- operative.
Anterior-posterior radiograph at 1 year post- operative.
Discussion
Nickel allergy caused by the implantation of a stainless steel plate was first reported in the medical literature in 1966 by Foussereau and Laugier [5]. Nickel is most commonly implicated in metal allergies, but other metals used in reconstructive surgery have also been associated with allergic reactions, including cobalt, chromium, and titanium.Although many authors have noted an association between metal hypersensitivity and the development of allergies to implanted metallic devices, a causal relationship has not been definitively proven [6]. Metal hypersensitivity is fairly common, with a reported incidence of 10–15% in the general population. It increases to 25% in patients with well-functioning implants and up to 60% when there is evidence of aseptic loosening [7,8].Aseptic loosening is a major cause of implant failure leading to revision, and there is concern that patients with metal hypersensitivity will have a higher rate of aseptic loosening. With revision surgery costing 2–3 times the primary operation, this can lead to a significant financial burden on the health system [9].The immune system is responsible for the development of allergies to implanted metallic devices. Therefore, it is important to understand the underlying pathophysiology. Essentially, any metallic implant that is in contact with body fluids will undergo corrosion and release metal ions, which are not allergenic in themselves, but may combine with tissue proteins and then act as antigens. The reaction is a delayed hypersensitivity Type IV allergic reaction which involves the initial phase of sensitization and then the effector phase with the activation of T-cells and macrophages [6]. The primary antigen-presenting cell in the dermis is the Langerhans cell which produces a dermatitis-type reaction. It must be borne in mind that materials other than those commonly implicated in allergic-type reactions can also cause contact dermatitis. Polymethylmethacrylate used in fixing the implants to bone, gentamycin embedded in the cement and even sutures have all been reported to cause contact dermatitis [10, 11, 12].Histological examination of tissues around failed metal-on-metal hip arthroplasties has revealed perivascular and diffuse lymphocyte infiltration, both of which are associated with delayed-type hypersensitivity reactions. These findings support the theory that delayed-type hypersensitivity to metals plays an important role in implant failure [13].Skin patch testing remains the most common method used to test for delayed hypersensitivity. It is a relatively inexpensive method that can be readily performed in the office with commercially available kits. Some drawbacks include false-positive results due to contact irritation and false-negative results due to the lack of ions produced from corrosion or dissolution.In this case, the patient presented on two occasions following oxidized zirconium TKRs with a temporally related eczematous dermatitis reaction on the skin surrounding the incision. The clinical features were in keeping with the description by Verma et al. in their report of the typical allergic dermatitis-type reactions [14]. Determining whether the allergic reaction is due to an implanted metal is difficult. Some authors have proposed management algorithms to diagnose and treat clinically suspected implant-related allergies, although none are universally accepted [15].Common among the many reports is the need to first exclude infection as the cause of a skin reaction following TKR [16]. Using the 2018 definition of periprosthetic hip and knee infection as proposed by the musculoskeletal infection society, our patient’s aggregate score did not meet the threshold for periprosthetic infection [17].Patch testing using the implant manufacturer-derived testing kit revealed no evidence of skin allergy to the implanted metals. Based on the clinical features and there being no evidence of infection or metal allergy, we suspected that the skin reaction was due to contact dermatitis caused by an unknown allergen. Verma et al. reported on 30 patients who developed post-operative contact dermatitis following TKR, in 23 patients, no allergen was found with only 7 cases showing any signs of metal hypersensitivity but all patients were treated successfully with topical corticosteroids [14]. In keeping with the report by Verma et al., our patient responded well to topical corticosteroids.The well-publicized failure of metal-on-metal hip replacement has no doubt heightened public awareness of the potential problems related to metal implants. Manufacturers have been quick to capitalize on this and several companies now produce hypoallergenic implants. Oxidized zirconium (Oxinium®, Smith and Nephew) has been proven efficacious at preventing post-operative allergic reactions while being less expensive than the alumina ceramic implant [18]. In our patient, we chose an oxidized zirconium implant, because of its reported durability in active patients, an added benefit was its hypoallergenicity. Although we have excluded a metal allergy as the cause of the skin reaction, we were unable to identify the allergen in this patient.
Conclusion
To the best of our knowledge, this is the first case of dermatitis occurring after an oxidized zirconium TKR. This highlights the possibility than in susceptible individuals undergoing joint replacement, allergic skin reactions may occur even with the use of hypoallergenic implants. This is important since many patients choose to use hypoallergenic implants with the expectation that allergic reactions will not occur. Satisfaction is directly related to surgical expectations, and the occurrence of an allergenic skin reaction in this group of patients is likely to lead to increased dissatisfaction rates.Allergic skin reactions can occur in patients following hypoallergenichypo allergenic knee implants. A trial of topical corticosteroids may be considered but early peri-prosthetic infection must be excluded because this will likely worsen the infection.
Authors: Panayiotis D Megaloikonomos; Evanthia Mitsiokapa; Dimitrios A Flevas; Georgios Kakouratos; Georgios Kyrou; Panayiotis Koulouvaris; Andreas F Mavrogenis Journal: J Long Term Eff Med Implants Date: 2017