| Literature DB >> 34125060 |
Linda Sabbadini1, Rocco Germano1, Emily Hopkins2, Jason S Haukoos2,3,4, John L Kendall2,3.
Abstract
INTRODUCTION: Ultrasound hypotension protocols (UHP) involve imaging multiple body areas, each with different transducers and imaging presets. The time for task switching between presets and transducers to perform an UHP has not been previously studied. A novel hand-carried ultrasound (HCU) has been developed that uses a multifrequency single transducer to image areas of the body (lung, heart, abdomen, superficial) that would typically require three transducers using a traditional cart-based ultrasound (CBU) system. Our primary aim was to compare the time to complete UHPs with a single transducer HCU to a multiple transducer CBU.Entities:
Mesh:
Year: 2021 PMID: 34125060 PMCID: PMC8202981 DOI: 10.5811/westjem.2020.12.47862
Source DB: PubMed Journal: West J Emerg Med ISSN: 1936-900X
Total time to complete ultrasound hypotension protocol, using handheld Butterfly vs cart-based GE devices.
| Time | Butterfly | GE | Difference | 95% CI | P-value |
|---|---|---|---|---|---|
| Total time 1 | |||||
| Mean, (SD) | 5.55 (3.51) | 9.57 (3.73) | −4.02 | (−5.93 – −2.12) | <0.0001 |
| Median, (IQR) | 4.28 (3.63 – 5.62) | 8.52 (7.17 – 9.57) | −4.23 | (−5.48 – −2.98) | <0.0001 |
| Total time 2 | |||||
| Mean, (SD) | 5.32 (3.54) | 7.13 (3.51) | −1.81 | (−3.67 – 0.04) | 0.0556 |
| Median, (IQR) | 4.08 (3.43 – 5.35) | 5.8 (4.73 – 7.67) | −1.72 | (−2.78 – −0.65) | <0.0001 |
Total time (minutes) includes transport and plug-in.
Total time (minutes) excludes transport and plug-in.
UHP, ultrasound hypotension protocols; CI, confidence interval; SD, standard deviation; IQR, interquartile range.
Number of diagnostic-quality images, handheld Butterfly vs cart-based GE ultrasound device.
| Component of UHP exam | Butterfly N = 29 | GE N = 29 | Absolute difference 95% CI | P-value* |
|---|---|---|---|---|
| PLAX | 27 (93, 77 – 99) | 28 (97, 82 – 100) | −3 (−15 – 8) | 1.00 |
| 4C View | 26 (90, 73 – 98) | 25 (86, 68 – 96) | 3 (−13 – 20) | 1.00 |
| IVC | 23 (79, 60 – 92) | 24 (82, 64 – 94) | −3 (−23 – 17) | 1.00 |
| RUQ | 29 (100, 88 – 100) | 29 (100, 88 – 100) | 0 (0 – 0) | – |
| LUQ | 29 (100, 88 – 100) | 29 (100, 88 – 100) | 0 (0 – 0) | – |
| Bladder | 29 (100, 88 – 100) | 29 (100, 88 – 100) | 0 (0 – 0) | – |
| Aorta | 25 (86, 68 – 96) | 23 (79, 68 – 96) | 7 (−12 – 26) | 0.73 |
| RL | 29 (100, 88 – 100) | 29 (100, 88 – 100) | 0 (0 – 0) | – |
| LL | 29 (100, 88 – 100) | 29 (100, 88 – 100) | 0 (0 – 0) | – |
UHP, ultrasound hypotension protocols; CI, confidence interval; PLAX, parasternal long axis; 4C, four chamber; IVC, inferior vena cava; RUQ, right upper quadrant; LUQ, left upper quadrant; RL, right leg; LL, left leg.
Comparison of task-switching time, handheld Butterfly vs GE cart-based ultrasound device.
| Task switching | Butterfly | GE | Difference | 95% CI | P-value |
|---|---|---|---|---|---|
| Total time 1 | |||||
| Mean, (SD) | 5.55 (3.51) | 9.57 (3.73) | −4.02 | (−5.93 – −2.12) | <0.0001 |
| Median, (IQR) | 4.28 (3.63 – 5.62) | 8.52 (7.17 – 9.57) | −4.23 | (−5.48 – −2.98) | <0.0001 |
| Total time 2 | |||||
| Mean, (SD) | 5.32 (3.54) | 7.13 (3.51) | −1.81 | (−3.67 – 0.04) | 0.0556 |
| Median, (IQR) | 4.08 (3.43 – 5.35) | 5.8 (4.73 – 7.67) | −1.72 | (−2.78 – −0.65) | <0.0001 |
| Switching 1 | |||||
| Mean, (SD) | 0.13 (0.06) | 0.30 (0.08) | −0.17 | (−0.21 – −0.13) | <0.0001 |
| Median, (IQR) | 0.13 (0.08 – 0.17) | 0.30 (0.23 – 0.35) | −0.17 | −0.22 – −0.12 | <0.0001 |
| Switching 2 | |||||
| Mean, (SD) | 0.09 (0.08) | 0.08 (0.10) | 0.01 | (−0.04 – 0.06) | 0.6893 |
| Median, (IQR) | 0.08 (0.00 – 0.13) | 0.00 (0.0 – 0.2) | 0.08 | (−0.02 – 0.18) | 0.0825 |
| Switching 3 | |||||
| Mean, (SD) | 0.12 (0.05) | 0.33 (0.23) | −0.20 | (−0.30 – −0.11) | <0.0001 |
| Median, (IQR) | 0.10 (0.10 – 0.13) | 0.28 (0.23 – 0.38) | −0.18 | (−0.23 – −0.14) | <0.0001 |
| Total switching | |||||
| Mean, (SD) | 0.34 (0.14) | 0.71 (0.31) | −0.37 | (−0.37 – −0.50) | <0.0001 |
| Median, (IQR) | 0.33 (0.23 – 0.42) | 0.67 (0.48 – 0.82) | −0.33 | (−0.45 – −0.21) | <0.0001 |
Total time 1 (minutes): includes transport and plug-in.
Total time 2 (minutes): excludes transport and plug-in.
Switching 1 (minutes): cardiac to abdomen preset.
Switching 2 (minutes): abdomen to abdomen deep preset.
Switching 3 (minutes): abdomen deep to lung preset.
CI, confidence interval; SD, standard deviation; IQR, interquartile range.