| Literature DB >> 34121320 |
Kai Siang Chan1, Shanying Liang1, Yuan Teng Cho1, Yam Meng Chan1, Audrey Hui Min Tan2, Sivakami Muthuveerappa2, Tina Peiting Lai2, Cheng Cheng Goh2, Annie Joseph3, Qiantai Hong1, Enming Yong1, Li Zhang1, Lester Rhan Chaen Chong1, Glenn Wei Leong Tan1, Sadhana Chandrasekar1, Zhiwen Joseph Lo1,3,4.
Abstract
Chronic venous insufficiency is a chronic disease of the venous system with a prevalence of 25% to 40% in females and 10% to 20% in males. Venous leg ulcers (VLUs) result from venous insufficiency. VLUs have a prevalence of 0.18% to 1% with a 1-year recurrence of 25% to 50%, bearing significant socioeconomic burden. It is therefore important for regular assessment and monitoring of VLUs to prevent worsening. Our study aims to assess the intra- and inter-rater reliability of a machine learning-based handheld 3-dimensional infrared wound imaging device (WoundAide [WA] imaging system, Konica Minolta Inc, Tokyo, Japan) compared with traditional measurements by trained wound nurse. This is a prospective cross-sectional study on 52 patients with VLUs from September 2019 to January 2021 using three WA imaging systems. Baseline patient profile and clinical demographics were collected. Basic wound parameters (length, width and area) were collected for both traditional measurements and measurements taken by the WA imaging systems. Intra- and inter-rater reliability was analysed using intra-class correlation statistics. A total of 222 wound images from 52 patients were assessed. There is excellent intra-rater reliability of the WA imaging system on three different image captures of the same wound (intra-rater reliability ranging 0.978-0.992). In addition, there is excellent inter-rater reliability between the three WA imaging systems for length (0.987), width (0.990) and area (0.995). Good inter-rater reliability for length and width (range 0.875-0.900) and excellent inter-rater reliability (range 0.932-0.950) were obtained between wound nurse measurement and each of the WA imaging system. In conclusion, high intra- and inter-rater reliability was obtained for the WA imaging systems. We also obtained high inter-rater reliability of WA measurements against traditional wound measurement. The WA imaging system is a useful clinical adjunct in the monitoring of VLU wound documentation.Entities:
Keywords: chronic venous insufficiency; machine learning; venous leg ulcer; wound healing; wound imaging
Mesh:
Year: 2021 PMID: 34121320 PMCID: PMC8762571 DOI: 10.1111/iwj.13644
Source DB: PubMed Journal: Int Wound J ISSN: 1742-4801 Impact factor: 3.315
FIGURE 1Study protocol for participation recruitment and standardisation of process for wound measurement
FIGURE 2Schematic diagram for the determination of length and width using traditional measurements by a trained wound nurse
FIGURE 3Usage of the WoundAide imaging system for the measurement of venous leg ulcer at approximately 40 cm from the wound with optical zoom of ×1.0
FIGURE 4Application interface of the WoundAide imaging system (Version 1.5.0.10) on an iPad (iPad Mini 4, iOS version 12.4.1)
Clinical profile of patients included in the study
| Number of patients (n = 52) | |
|---|---|
| Age, median (IQR) | 67 (60.3‐75.8) |
| Gender, male, n (%) | 30 (57.7) |
| Co‐morbidities, n (%) | |
| Diabetes mellitus | 18 (34.6) |
| Hypertension | 31 (59.6) |
| Heart disease | 9 (17.3) |
| Kidney disease | 7 (13.5) |
| Cerebrovascular accident | 5 (9.6) |
| Peripheral vascular disease | 50 (96.2) |
| Varicose veins | 25 (48.1) |
| Previous venous surgery | 16 (30.8) |
| Previous skin/leg ulcers | 46 (88.5) |
| Location of ulcer, n (%) | |
| Calf | 3 (5.8) |
| Shin | 12 (23.1) |
| Medial gaiter | 2 (3.8) |
| Medial malleolus | 17 (32.7) |
| Lateral malleolus | 15 (28.8) |
| Dorsum of foot | 3 (5.8) |
| Wound parameters, median (IQR) | |
| Length, cm | 3.10 (2.00‐5.88) |
| Width, cm | 2.35 (1.28‐4.00) |
| Area, cm2 | 6.65 (2.10‐11.50) |
Abbreviations: IQR, interquartile range.
Baseline characteristics of all included wound episodes
| Number of wound episodes (n = 222) | |
|---|---|
| Age, median (IQR) | 67 (60.8‐74) |
| Gender, male, n (%) | 137 (61.7) |
| Co‐morbidities | |
| Diabetes mellitus | 76 (34.2) |
| Hypertension | 126 (56.8) |
| Heart disease | 39 (17.6) |
| Kidney disease | 27 (12.2) |
| Cerebrovascular accident | 24 (10.8) |
| Peripheral vascular disease | 214 (96.4) |
| Varicose veins | 104 (46.8) |
| Previous venous surgery | 68 (30.6) |
| Previous skin/leg ulcers | 199 (89.6) |
| Location of ulcer | |
| Calf | 13 (5.9) |
| Shin | 51 (23.0) |
| Medial gaiter | 8 (3.6) |
| Medial malleolus | 72 (32.4) |
| Lateral malleolus | 64 (28.8) |
| Dorsum of foot | 14 (6.3) |
| Wound parameters, median (IQR) | |
| Length, cm | 3.00 (1.90‐5.03) |
| Width, cm | 2.20 (1.10‐3.50) |
| Area, cm2 | 5.00 (1.68‐11.00) |
Abbreviations: IQR, interquartile range.
Intra‐rater reliability of the same WoundAide device on three different images obtained from the same wound
| Measurements, mean ± SD | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Image 1 | Image 2 | Image 3 | Intra‐rater reliability (95% CI) |
| |||||||||||
| Length | Width | Area | Length | Width | Area | Length | Width | Area | Length | Width | Area | Length | Width | Area | |
| Device 1 | 3.43 ± 2.55 | 2.49 ± 1.73 | 9.06 ± 12.69 | 3.45 ± 2.49 | 2.49 ± 1.69 | 9.00 ± 12.27 | 3.44 ± 2.52 | 2.49 ± 1.74 | 9.17 ± 12.80 | 0.979 (0.974, 0.983) | 0.978 (0.972, 0.982) | 0.990 (0.987, 0.992) | <.001 | <.001 | <.001 |
| Device 2 | 3.34 ± 2.43 | 2.47 ± 1.69 | 8.88 ± 12.55 | 3.34 ± 2.43 | 2.45 ± 1.69 | 8.86 ± 12.30 | 3.30 ± 2.40 | 2.46 ± 1.76 | 8.83 ± 12.53 | 0.978 (0.973, 0.983) | 0.979 (0.973, 0.983) | 0.990 (0.988, 0.992) | <.001 | <.001 | <.001 |
| Device 3 | 3.45 ± 2.52 | 2.48 ± 1.70 | 9.18 ± 12.90 | 3.42 ± 2.54 | 2.49 ± 1.75 | 9.32 ± 13.08 | 3.42 ± 2.52 | 2.48 ± 1.74 | 9.19 ± 12.76 | 0.989 (0.986, 0.991) | 0.980 (0.975, 0.984) | 0.992 (0.990, 0.994) | <.001 | <.001 | <.001 |
Note: Length and breadth are expressed in cm. Area was expressed in cm2.
Inter‐rater reliability of the three different WoundAide device on the first image, second image and third image and the average measurements across all three images taken on each device
| Measurements, mean ± SD | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Device 1 | Device 2 | Device 3 | Inter‐rater reliability |
| |||||||||||
| Length | Width | Area | Length | Width | Area | Length | Width | Area | Length | Width | Area | Length | Width | Area | |
| Image | 3.45 ± 2.50 | 2.50 ± 1.71 | 9.13 ± 12.56 | 3.36 ± 2.44 | 2.47 ± 1.69 | 8.99 ± 12.48 | 3.49 ± 2.57 | 2.52 ± 1.74 | 9.50 ± 13.23 | 0.987 (0.983, 0.990) | 0.990 (0.988, 0.992) | 0.995 (0.994, 0.996) | <.001 | <.001 | <.001 |
Note: Length and breadth are expressed in cm. Area was expressed in cm2.
Inter‐rater reliability between wound nurse and WoundAide device 1 for the corresponding image
| Measurements, mean ± SD | Inter‐rater reliability |
| |||||||
|---|---|---|---|---|---|---|---|---|---|
| Length | Width | Area | Length | Width | Area | Length | Width | Area | |
| Device 1 | 3.43 ± 2.49 | 2.49 ± 1.70 | 9.03 ± 12.47 | 0.882 (0.829, 0.916) | 0.893 (0.860, 0.918) | 0.936 (0.899, 0.957) | <.001 | <.001 | <.001 |
| Wound nurse | 3.86 ± 2.82 | 2.67 ± 1.94 | 10.95 ± 16.36 | ||||||
Note: Length and breadth are expressed in cm. Area was expressed in cm2.
Inter‐rater reliability between wound nurse and WoundAide device 2 for the corresponding image
| Measurements, mean ± SD | Inter‐rater reliability | p value | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Length | Width | Area | Length | Width | Area | Length | Width | Area | |
| Device 2 | 3.36 ± 2.44 | 2.47 ± 1.69 | 8.99 ± 12.48 | 0.875 (0.802, 0.917) | 0.891 (0.854, 0.919) | 0.932 (0.890, 0.956) | <.001 | <.001 | <.001 |
| Wound nurse | 3.88 ± 2.83 | 2.69 ± 1.95 | 11.07 ± 16.48 | ||||||
Note: Length and breadth are expressed in cm. Area was expressed in cm2.
Inter‐rater reliability between wound nurse and WoundAide device 3 for the corresponding image
| Measurements, mean ± SD | Inter‐rater reliability | p value | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Length | Width | Area | Length | Width | Area | Length | Width | Area | |
| Device 3 | 3.47 ± 2.55 | 2.51 ± 1.73 | 9.39 ± 13.14 | 0.889 (0.844, 0.920) | 0.900 (0.870, 0.923) | 0.950 (0.926, 0.965) | <.001 | <.001 | <.001 |
| Wound nurse | 3.86 ± 2.82 | 2.67 ± 1.94 | 10.95 ± 16.36 | ||||||
Note: Length and breadth are expressed in cm. Area was expressed in cm2.