| Literature DB >> 34096199 |
Charlotte Kelly1, Ioannis Delakis1,2.
Abstract
We describe the technical evaluation of the commercially available, clinical, bi-planar, low dose, digital X-ray system (EOS System, EOS imaging, France). The unit is used for upright, weight-bearing musculoskeletal pathologies, in particular, in the spine and lower limbs. The evaluation incorporated tests on the X-ray generator performance, radiation/imaging field alignment, dose area product accuracy and image quality. The assessment methodology was based on objective parameters and required equipment readily available for technical evaluation of other radiological equipment. Results demonstrated that the system performs well within acceptable performance criteria with regard to X-ray generator performance, radiation/imaging field alignment and dose area product accuracy. In addition, results from the image-quality assessment were aligned with previously published work. The work presented in this article can be used for the technical evaluation of the EOS System at other clinical sites.Entities:
Keywords: data collection; imaging; phantoms; quality control; radiation protection; radiologic; technology
Mesh:
Year: 2021 PMID: 34096199 PMCID: PMC8655884 DOI: 10.1002/jmrs.519
Source DB: PubMed Journal: J Med Radiat Sci ISSN: 2051-3895
Figure 1EOS System (EOS imaging, France) installed in our hospital. Patients step onto the central platform of the system where they stand whilst the image is acquired.
Figure 2Geometry of the EOS System (view from above) indicating the Source to Image Distances (SID) for each X‐ray tube/detector combination (frontal and lateral). The patient stands in the middle of the central platform of the system. The distance between each X‐ray tube and the EOS system covers nearest to the tube is indicated in blue.
Figure 3(A) Dosemeter (R/F chamber) position for generator performance tests of the frontal X‐ray tube. (B): Dosemeter (pencil chamber) position for dose area product (DAP) indicator accuracy assessment.
Dose area product (DAP) results.
| Patient morphotype | kV | mAs | Speed setting | Indicated DAP (mGycm2) | Measured DAP (mGycm2) |
|---|---|---|---|---|---|
| Frontal tube | |||||
| Thin | 83 | 200 | 4 | 1.48 | 1.6 ± 0.2 |
| Normal | 90 | 250 | 4 | 2.20 | 2.3 ± 0.3 |
| Large | 100 | 250 | 4 | 2.74 | 2.8 ± 0.3 |
| Lateral tube | |||||
| Thin | 102 | 200 | 4 | 2.31 | 2.5 ± 0.3 |
| Normal | 105 | 250 | 4 | 3.07 | 3.3 ± 0.3 |
| Large | 110 | 320 | 4 | 6.02 | 5.9 ± 0.4 |
X‐ray generator performance results
| Parameter | Frontal tube | Lateral tube |
Remedial level | Suspension level |
|---|---|---|---|---|
| kV error (kV) | −3.4 | 0.9 | ±5 | ±10 |
| kV accuracy (%) | 0.03 | 0.01 | ±5 | ±10 |
| Output repeatability (%) | 0.02 | 0.05 | Mean ±10% | Mean ±20% |
| Specific radiation output (µGy/mAs) | 29.4 | 29.2 | Baseline ±20% | Baseline ±50% |
| Inferred filtration (mmAl) at 80 kV | 6.4 | 6.7 | ||
| Half‐Value Layer (mm Al) at 80 kV | 5.11 | 5.16 |
Figure 4(A) GMTF, (B) GNNPS and (C) GDQE for the frontal X‐ray tube and detector pair.
Figure 5(A) GMTF, (B) GNNPS and (C) GDQE for the lateral X‐ray tube and detector pair.