Ruili Feng1,2,3, Dongyue Wang1,2,3, Yanlu Chen1,2,3, Jing Lu1,2,3, Liang Gao1,2,3.
Abstract
Here, we describe a protocol to construct, calibrate, and operate a versatile tiling light sheet microscope for imaging cleared tissues. The microscope uses adjustable tiling light sheets to achieve higher spatial resolution and better optical sectioning ability than conventional light sheet microscopes and to image cleared tissues with the cellular to the subcellular spatial resolution. It is compatible with all tissue clearing methods and aligned semiautomatically through the phase modulation of the illumination light. For complete details on the use and execution of this protocol, please refer to Chen et al. (2020).Entities:
Keywords: Microscopy; Neuroscience
Mesh:
Year: 2021 PMID: 34095860 PMCID: PMC8164092 DOI: 10.1016/j.xpro.2021.100546
Source DB: PubMed Journal: STAR Protoc ISSN: 2666-1667
Figure 1The design and construction of the presented tiling light sheet microscope
(A) The control system diagram of the microscope.
(B) The optical system diagram of the microscope.
(C) The optomechanical design of the microscope.
(D) The excitation path of the microscope and the associated sub-assemblies.
(E and F) The construction steps of the microscopes. 1, Laser combiner assembly. 2, Laser collimation assembly. 3, PBS assembly. 4, SLM assembly. 5, Optical slit assembly. 6, Galvanometer assembly. 7, Prism mirror assembly. 8, Relay lens assembly. 9, Excitation objective assembly. 10, Imaging chamber assembly. 11, Sample scan assembly. 12, Detection path mounting base.
Figure 2The custom software
The control panel layout and functions of the custom (A) phase map generation, (B) main control, and (C) image reconstruction software.
Figure 3The calibration of the microscope using dye solution or imaging buffers
(A) The excitation beam stays in focus at all X positions across the detection FOV.
(B–D) Zoom-in views of the excitation beam at the selected positions in (A).
(E and F) The left and right tiling positions defined by the user.
(G–J) The interpolated tiling positions according to the calibrated left and right tiling positions and the desired tiling number.
Scale bars, 200 μm (A, E-J), and 100 μm (B–D).
Figure 4Comparisons of the different spatial resolutions obtained with different tiling light sheets
(A) Lateral maximum intensity projection (MIP) of a 2.3×2.3×0.5 mm3 sample volume in a Thy1-eGFP mouse brain cleared with uDISCO. (B-D) Axial MIPs of the selected sample volume in (A) imaged using light sheets with different light sheets and tiling numbers.
(B) Excitation NAOD=0.01, NAID=0, 1 tile.
(C) Excitation NAOD=0.03, NAID=0.01, 3 tiles.
(D) Excitation NAOD=0.05, NAID=0.015, 6 tiles.
(E–G) Zoom-in views of the same selected areas in (B–D).
Scale bars, 200 μm (A–D), and 20 μm (E–J).
Figure 5Cleared tissue imaging with micron-scale to submicron-scale spatial resolution
(A and B) Lateral and axial MIPs of a Thy1-eGFP mouse brain cleared with CUBIC-L imaged at 2×2×5 μm3 spatial resolution using a 0.25 NA air detection objective.
(C) Axial MIP of the selected sample volume in (A).
(D) The same sample volume in (B) imaged at 0.6×0.6×1.5 μm3 spatial resolution using a 0.5 NA immerse detection objective.
(E–H) Lateral and Axial projections of the selected 100×100×100 μm3 regions in (D).
Scale bars, 1 mm (A and B), 200 μm (C and D), and 10 μm (E–H).
| REAGENT or RESOURCE | SOURCE | IDENTIFIER |
|---|---|---|
| Alexa Fluor 488, Hydrazide, sodium salt | Thermo | A10436 |
| Mouse: Thy1-eGFP-M | The Jackson Laboratory | JAX: 007788 |
| LabVIEW 2019 developer suite | National Instruments | |
| MATLAB 2019a | MathWorks | |
| Amira (optional) | Thermo Fisher | |
| Main control software | Westlake University | Lab developed |
| Phase map generation software | Westlake University | Lab developed |
| Image reconstruction software | Westlake University | Lab developed |
| 405 nm CW Laser (150 mW), qty: 1 | Coherent | OBIS 405 |
| 488 nm CW Laser (150 mW), qty: 1 | Coherent | OBIS 488 |
| 561 nm CW Laser (150 mW), qty: 1 | Coherent | OBIS 561 |
| Spatial light modulator and controller, qty: 1 | Fourth Dimension Displays | SXGA-3DM |
| Galvanometer and controller (6215H, 7 mm, Y axis), qty: 1 | Cambridge Technology | 6SD11490 |
| Linear translational stage and controller, 26 mm distance, qty: 3 | Physik Instruments | Q545.240 |
| ORCA-Fusion Digital CMOS camera and frame grabber, qty: 1 | Hamamatsu | C14440-20UP |
| Filter wheel, qty: 1 | Thorlabs | FW102-C |
| Lens, f=30 mm, Ø=25 mm, qty: 1 | Edmund Optics | 47633 |
| Lens, f=175 mm, Ø =25 mm, qty: 1 | Edmund Optics | 47644 |
| Lens, f=200 mm, Ø =25 mm, qty: 1 | Edmund Optics | 47645 |
| Lens, f=300 mm, Ø =25 mm, qty: 1 | Edmund Optics | 47649 |
| Lens, f=150 mm, Ø =50 mm, qty: 2 | Edmund Optics | 49285 |
| Lens, f=250 mm, Ø =50 mm, qty: 2 | Edmund Optics | 49287 |
| Prism mirror, qty: 2 | Edmund Optics | 87390 |
| Polarizing beam splitter, qty: 1 | Thorlabs | CCM1-PBS251/M |
| 1 inch broadband dielectric mirror, qty: 7 | Thorlabs | BB1-E02 |
| 2 inch broadband dielectric mirror, qty: 4 | Thorlabs | BB2-E02 |
| LaserMUX single-edge laser dichroic beamsplitter, qty: 1 | Semrock | LM01-427-25 |
| LaserMUX single-edge laser dichroic beamsplitter, qty: 1 | Semrock | LM01-503-25 |
| 447/60 nm single-band bandpass filter, qty: 1 | Semrock | FF02-447/60-25 |
| 525/50 nm single-band bandpass filter, qty: 1 | Semrock | FF03-525/50-25 |
| 617/73 nm single-band bandpass filter, qty: 1 | Semrock | FF02-617/73-25 |
| Halfwave plate, qty: 2 | Bolder Vision | AHWP3 |
| Excitation objective, qty: 2 | Mitutoyo | MY5X-802 |
| Detection objective, qty: 1 | Olympus | MVPLAPO 1X |
| Excitation objective (optional), qty: 2 | Mitutoyo | MY10X-803 |
| Detection objective (optional), qty: 1 | Olympus | MVPLAPO 2XC |
| Detection objective (optional), qty: 1 | Olympus | XLSLPLN25XGMP |
| Metric optical table (>0.9 m×0.9 m), qty: 1 | Thorlabs | B9090B |
| Precision kinematic mirror mounts, qty: 3 | Thorlabs | KS1 |
| Threaded precision kinematic mirror mounts, qty: 2 | Thorlabs | KS1T |
| Right-angle kinematic mirror mount 30 mm, qty: 5 | Thorlabs | KCB1/M |
| Right-angle kinematic mirror mount 60 mm, qty: 4 | Thorlabs | KCB2/M |
| Kinematic platform mount, qty: 1 | Thorlabs | KM200B/M |
| Compact kinematic mirror mount, qty: 2 | Thorlabs | KMS/M |
| 1 inch translating lens mount, qty: 4 | Thorlabs | CXY1 |
| 2 inch translating lens mount, qty: 4 | Thorlabs | CXY2 |
| Cage rotation mount, qty: 2 | Thorlabs | CRM1/M |
| Adjustable slit, qty: 1 | Thorlabs | VA100C/M |
| XY dovetail translation stage, qty: 1 | Thorlabs | DT12XY/M |
| XZ dovetail translation stage, qty: 1 | Thorlabs | DT12XZ/M |
| 1 inch translation stage, qty: 2 | Thorlabs | PT1B/M |
| Objective XY translator, qty: 2 | Thorlabs | ST1XY-A/M |
| Complete kinematic base, qty: 1 | Thorlabs | KB25/M |
| 50 mm Pillar post, qty: 6 | Thorlabs | RS50/M |
| 30 mm cage mounting bracket, qty: 6 | Thorlabs | CP02B |
| 60 mm cage mounting bracket, qty: 2 | Thorlabs | LCP01B |
| Thread adapter, qty: 2 | Thorlabs | SM1A27 |
| Lens tube, qty: 2 | Thorlabs | SM1L05 |
| 1 inch cage assembly rod, qty: 4 | Thorlabs | ER1 |
| 2 inch cage assembly rod, qty: 4 | Thorlabs | ER2 |
| 3 inch cage assembly rod, qty: 4 | Thorlabs | ER3 |
| 4 inch cage assembly rod, qty: 8 | Thorlabs | ER4 |
| 6 inch cage assembly rod, qty: 4 | Thorlabs | ER6 |
| 25 mm Pillar POST, qty: 4 | Thorlabs | RS50/M |
| Other mechanical components | Westlake University | Lab developed |
| Control unit, qty: 1 | Olympus | SZX-MDHSW |
| Control unit hand switch, qty: 1 | Olympus | U-ACAD4515 |
| 400 mm pillar, qty: 1 | Olympus | SZH-P400 |
| Drop prevention collar, qty: 1 | Olympus | SZX-R |
| Motorized focus unit, qty: 1 | Olympus | SZX2-FOA |
| MVX10 zoom body, qty: 1 | Olympus | MVX-ZB10 |
| Tube lens unit, qty: 1 | Olympus | MVX-TLU |
| Camera port, qty: 1 | Olympus | MVX-TV 1XC |
| Workstation (Optional configuration: Supermicro X11SPA-TF motherboard, Intel Xeon W-3235 processor, >256 GB memory, >12 TB SSD storage space.) | Customized | |
| PCIe-6323 multifunction I/O card, qty: 1 | National Instruments | 781045-01 |
| SCB-68A connector block, qty: 2 | National Instruments | 782536-01 |
| SHC68-68-EPM shielded cable, qty: 2 | National Instruments | 782536-01 |