Jesús Moreno-Fernández1, José Alberto García-Seco1.
Abstract
OBJECTIVE: Hybrid closed-loop (HCL) devices can achieve tight glycemic control but are rarely used in pregnancy, which remains an off-label indication. We present a case of a pregnant patient with type 1 diabetes mellitus (T1DM) who used the Medtronic MiniMed 670G HCL system.Entities:
Keywords: HCL, hybrid closed-loop; HbA1c, glycated hemoglobin; MM670G, Medtronic MiniMed 670G; RT-CGM, real-time continuous glucose monitoring; T1DM, type 1 diabetes mellitus; TIR, time in range; TUR, time under range; closed-loop; pregnancy; type 1 diabetes mellitus
Year: 2021 PMID: 34095482 PMCID: PMC8165106 DOI: 10.1016/j.aace.2020.11.039
Source DB: PubMed Journal: AACE Clin Case Rep ISSN: 2376-0605
Real-Time Continuous Glucose Monitoring Evolution
| 16 wks (before auto mode) | 20 wks | 28 wks | 32 wks | 36 wks | |
|---|---|---|---|---|---|
| RT-CGM adherence, % of time | 96 | 93 | 91 | 97 | 95 |
| Automode adherence, % of time | 0 | 96 | 95 | 96 | 98 |
| MIG, mmol/mol (mg/dL) | 7.6 (137) | 7.7 (139) | 7.4 (133) | 7.5 (135) | 8.0 (144) |
| CV, % | 38 | 31 | 31 | 30 | 24 |
| TAR, % | 50.2 | 47.6 | 47.6 | 46.5 | 35.1 |
| TIR, % | 46.8 | 51.3 | 50.3 | 52.8 | 63.5 |
| TUR, % | 3.0 | 1.1 | 2.1 | 0.7 | 1.4 |
Abbreviations: RT-CGM = real-time continuous glucose monitoring; MIG = mean interstitial glucose; CV = MIG coefficient of variation; TAR = time above range (>7.8 mmol/mol, >140 mg/dL); TIR = time in range (3.7-7.8 mmol/mol, 63-140 mg/dL); TUR = time under range (<3.7 mmol/mol, <63 mg/dL).
The table shows the improvement of glycemic control measured using RT-CGM based on the initiation of the Minimed Medtronic 670G device until the end of pregnancy.