| Literature DB >> 3409192 |
Y Higashida1, N Moribe, Y Hirata, K Morita, S Doudanuki, Y Sonoda, N Katsuda, Y Hiai, W Misumi, M Matsumoto.
Abstract
Threshold contrasts of low-contrast objects with computed radiography (CR) images were compared with those of blue and green emitting screen-film systems by employing the 18-alternative forced choice (18-AFC) procedure. The dependence of the threshold contrast on the incident X-ray exposure and also the object size was studied. The results indicated that the threshold contrasts of CR system were comparable to those of blue and green screen-film systems and decreased with increasing object size, and increased with decreasing incident X-ray exposure. The increase in threshold contrasts was small when the relative incident exposure decreased from 1 to 1/4, and was large when incident exposure was decreased further.Entities:
Mesh:
Year: 1988 PMID: 3409192 DOI: 10.1016/0895-6111(88)90025-0
Source DB: PubMed Journal: Comput Med Imaging Graph ISSN: 0895-6111 Impact factor: 4.790