OBJECTIVES: We investigated the Ozaki procedure using a single interrupted suture technique (SST) and compared this with the standard continuous suture technique (CST) with regard to hydrodynamic valve performance. In addition, both techniques were compared with the native aortic valve (NAV).
METHODS: Effective orifice area, mean pressure gradient and leakage volume were evaluated in the NAV as well as after an Ozaki procedure using SST or CST in fresh swine aortic roots using a mock circulation loop. The NAV, SST and CST were evaluated under 4 defined hydrodynamic conditions.
RESULTS: Both suture techniques resulted in a similar effective orifice area under all conditions [for stroke volume of 70 ml: SST: 1.50 (1.35-1.87) vs CST: 1.57 (1.41-1.72) cm2, P = 0.8] and there were no significant differences between both suture techniques and the NAV (P > 0.05). Regarding mean pressure gradient, the Ozaki procedure with SST and CST showed no significant differences [7.23 (5.53-8.91) vs 7.04 (6.65-7.60) mmHg, P = 0.72] and there was no significant difference between both suture techniques and the NAV (P > 0.1). In leakage volume, there was no significant difference between SST and CST [4.49 (3.91-4.99) vs CST: 4.23 (3.58-4.87) ml/stroke, P = 0.34].
CONCLUSIONS: The Ozaki procedure with SST performed similarly to that with CST with regard to hydrodynamic performance. Our results suggest that the Ozaki procedure can be performed with SST instead of CST, which may be useful in patients with limited surgical exposure, such as a small annulus.
© The Author(s) 2021. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  Aortic valve replacement; Hydrodynamic performance; Ozaki procedure; Single suture technique; Small aortic annulus

Mesh:

Year:  2021        PMID: 34089604      PMCID: PMC8923410          DOI: 10.1093/icvts/ivab138

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  19 in total

1.  Aortic valve reconstruction using self-developed aortic valve plasty system in aortic valve disease.

Authors:  Shigeyuki Ozaki; Isamu Kawase; Hiromasa Yamashita; Shin Uchida; Yukinari Nozawa; Takayoshi Matsuyama; Mikio Takatoh; So Hagiwara
Journal:  Interact Cardiovasc Thorac Surg       Date:  2011-01-27

2.  Effect of valve suture technique on incidence of paraprosthetic regurgitation and 10-year survival.

Authors:  Sukumaran K Nair; Gauraang Bhatnagar; Oswaldo Valencia; Venkatachalam Chandrasekaran
Journal:  Ann Thorac Surg       Date:  2010-04       Impact factor: 4.330

Review 3.  Prosthesis-patient mismatch: definition, clinical impact, and prevention.

Authors:  P Pibarot; J G Dumesnil
Journal:  Heart       Date:  2005-10-26       Impact factor: 5.994

Review 4.  Minimally invasive surgical aortic valve replacement.

Authors:  Marjan Jahangiri; Azhar Hussain; Enoch Akowuah
Journal:  Heart       Date:  2019-03       Impact factor: 5.994

Review 5.  Aortic Stenosis and Small Aortic Annulus.

Authors:  Afonso B Freitas-Ferraz; Gabriela Tirado-Conte; Francois Dagenais; Marc Ruel; Talal Al-Atassi; Eric Dumont; Siamak Mohammadi; Mathieu Bernier; Philippe Pibarot; Josep Rodés-Cabau
Journal:  Circulation       Date:  2019-06-03       Impact factor: 29.690

6.  Very long-term outcomes of the Carpentier-Edwards Perimount valve in aortic position.

Authors:  Thierry Bourguignon; Anne-Lorraine Bouquiaux-Stablo; Pascal Candolfi; Alain Mirza; Claudia Loardi; Marc-Antoine May; Rym El-Khoury; Michel Marchand; Michel Aupart
Journal:  Ann Thorac Surg       Date:  2015-01-09       Impact factor: 4.330

7.  Influence of stentless versus stented valves on ventricular remodeling assessed at 6 months by magnetic resonance imaging and long-term follow-up.

Authors:  Olivier Fouquet; Christophe Baufreton; Aude Tassin; Frédéric Pinaud; Jean-Patrice Binuani; Simon DangVan; Fabrice Prunier; Frédéric Rouleau; Serge Willoteaux; Jean-Louis De Brux; Alain Furber
Journal:  J Cardiol       Date:  2016-06-16       Impact factor: 3.159

8.  Aortic valve replacement: continuous-suture technique.

Authors:  D B Doty; R M Nelson
Journal:  J Card Surg       Date:  1986-12       Impact factor: 1.620

9.  Rapid Deployment Versus Conventional Bioprosthetic Valve Replacement for Aortic Stenosis.

Authors:  Stephan Ensminger; Buntaro Fujita; Timm Bauer; Helge Möllmann; Andreas Beckmann; Raffi Bekeredjian; Sabine Bleiziffer; Sandra Landwehr; Christian W Hamm; Friedrich W Mohr; Hugo A Katus; Wolfgang Harringer; Thomas Walther; Christian Frerker
Journal:  J Am Coll Cardiol       Date:  2018-04-03       Impact factor: 24.094

10.  Thirty-year experience with a bileaflet mechanical valve prosthesis.

Authors:  Scott Johnson; Martha R Stroud; John M Kratz; Scott M Bradley; Fred A Crawford; John S Ikonomidis
Journal:  J Thorac Cardiovasc Surg       Date:  2018-09-15       Impact factor: 5.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.