Ali S Alyami1,2,3, Hannah G Williams2,3, Konstantinos Argyriou4, David Gunn2,4, Victoria Wilkinson-Smith2,4, Jonathan R White2,4, Jaber Alyami5, Penny A Gowland3,4, Gordon W Moran2,4, Caroline L Hoad6,7.
Abstract
OBJECTIVE: Quantitative Magnetic Resonance Imaging sequences have been investigated as objective imaging biomarkers of fibrosis and inflammation in Crohn's disease. AIM: To determine the repeatability and inter- and intra-observer agreement of these measures in the prepared small bowel wall.Entities:
Keywords: Multiparametric magnetic resonance imaging; Small intestine; Validation study
Mesh:
Year: 2021 PMID: 34089407 PMCID: PMC8578109 DOI: 10.1007/s10334-021-00931-2
Source DB: PubMed Journal: MAGMA ISSN: 0968-5243 Impact factor: 2.310
Fig. 1Protocol schematic for the repeatability study
Parameters of MRI scans
| MRI parameters | T2 scan | T1 scan | IVIM-DWI scan | MT scan |
|---|---|---|---|---|
| Sequence | Single coronal slice spin-echo prepared bTFE | Single slice T1 weighted inversion recovery spoiled TFE | Coronal single-shot spin-echo-EPI | Single coronal slice TFE |
| Slice thickness (mm) | 5 | 5 | 5 | 5 |
| TR/TE (ms/ms) | 3.4 / 1.68 | 10 / 2.3 | 1278 / 71 | 20 / 2.3 |
| Field of view (mm2) | 340 × 352 | 375 × 351 | 400 × 400 | 375 × 351 |
| Number of slices | 1 | 1 | 12 | 2 |
| Breath-holding (BH) | 1 BH per TE value with a 15 s gap between TE acquisitions | 1 BH per TI value with a 15 s gap between TI acquisitions | Respiratory triggered | 11 s BH per slice to cover both ‘on’ and ‘off’ resonance image acquisition |
| Parameter variation | TE-prep values: 20, 50, 80, 120, 180, 300 ms | TI values: 500, 650, 800, 950, 1000, 1150, 1300, 2500, 5000 ms | b-values: 0, 50, 100, 200, 300, 400, 600, 800, 1000 s/mm2 | Presaturation MT on /off pulse |
| Approximate scan time including waiting between BH (mins) | 2 | 3–4 | 4–6 | 1 |
Fig. 2An example image highlighting the bowel wall that is automatically selected using a combination of thresholding and edge detection to define the walls. (A) T2 weighted bTFE image (TE = 20 ms). (B) Mask of the small bowel wall (red) overlaid on the TE = 20 ms image
Fig. 3An exmaple of ROI definition on MT images. The ROI was drawn on the small bowel wall after motion correction of the images. a shows the raw MT-on image with cropped area highlighted which was the target for motion correction. b Cropped area after motion correction. c MTR image with the bowel wall ROI highlighted
Fig. 4Single slice coronal images of the abdomen showing the T1 recovery of the bowel wall with different TIs in ms
Fig. 5Single slice coronal image of the abdomen showing T2 decay of the bowel wall with different TEs in ms
Test–retest repeatability, mean coefficients of variation (CoVs), Bland–Altman limits of agreements (BA-LA) and Intra-class correlation coefficients (ICC) of T2, T1, MTR, ADC, D, PF and D* parameters. CoVs across all participants calculated from Observer 1 Visit 1 data is also presented to indicate the variations in the parameters across the different participants
| Test–retest (within-participant) | Across participants | ||||||
|---|---|---|---|---|---|---|---|
| Parameter | N | Visit 1 Median (IQR) | Visit 2 | CoV % | BA bias | ICC, p value | CoV % |
| T2 (s) | 8 | 0.067 (0.054–0.083) | 0.060 (0.049–0.080) | 21 | 0.006 [−0.046–0.057] | − 0.385, p = 0.818 (−0.978 – 0.445) | 21 |
| T1 (s) | 9 | 1.0 (0.90–1.0) | 0.90 (0.77–0.96) | 8 | 0.08 [−0.16–0.32] | 0.485, p = 0.05 (−0.095 – 0.846) | 13 |
| MTR | 10 | 0.29 (0.25–0.31) | 0.28 (0.25–0.29) | 7 | 0.01 [−0.05–0.07] | 0.337, p = 0.157 (−0.370 – 0.779) | 11 |
ADC (10–3 mm2/s) | 10 | 3.06 (2.44–3.32) | 3.06 (2.67–3.33) | 5 | -0.03 [−0.63–0.57] | 0.814, p = 0.002 (0.409 – 0.901) | 18 |
D (10–3 mm2/s) | 10 | 2.4 (2.02–2.72) | 2.5 (2.05–2.89) | 10 | 0.01 [−0.93–0.91] | 0.650, p = 0.020 (0.042 – 0.901) | 24 |
| PF | 10 | 0.37 (0.32–0.45) | 0.43 (0.30–0.50) | 20 | -0.03 [−0.28–0.20] | 0.273, p = 0.211 (−0.385 – 0.750) | 24 |
D* (10–3 mm2/s) | 10 | 20 (12–32) | 22 (19–30) | 31 | −2.5 [−27–22] | 0.178, p = 0.310 (-0.522 – 0.712) | 50 |
Fig. 6Bland–Altman plots showing test–retest repeatability (% Difference vs average) for all measured quantitative parameters calculated from the small bowel MRI scans. Dotted lines indicate the bias (central line) and Dashed lines the corresponding 95% limits of agreement
Inter- and Intra-observer agreement intra-class correlation coefficients (ICCs) and Bland–Altman (BA) Bias and limits of agreements (LA) of the measured parameters. The number of measurements for each set of data is given by N. The median (IQR) number of ROIs used for each measurement are also given
| Inter-observer Data: Observer 1 and Observer 2 | |||||
|---|---|---|---|---|---|
| Parameter | Median | Median | ICC | BA Bias | |
| T2 (s) | 16 | 0.065 (0.052–0.080) | 0.063 (0.052–0.070) | 0.89 [0.71–0.96] | 0.004 [−14–15] |
Number of ROIs for T2 | 16 | 86 (51–165) | 105 (68–177) | ||
| T1 (s) | 16 | 0.91 (0.84–1.01) | 0.88 (0.71–0.95) | 0.55 [0.13–0.81] | 0.06 [−0.20–0.33] |
Number of ROIs for T1 | 16 | 12 (10–12) | 10 (10–10) | ||
| MTR | 20 | 0.29 (0.25–0.31) | 0.32 (0.30–0.34) | 0.08 [−0.15–0.39] | −0.04 [-0.11–0.04] |
Number of ROIs for MT | 20 | 18 (13–26) | 29 (25–32) | ||
ADC (10–3 mm | 20 | 3.1 (2.52–3.30) | 3.35 (2.6–3.60) | 0.76 [0.49–0.9] | −0.13 [−0.84–0.57] |
D (10–3 mm2/s) | 20 | 2.45 (2.10–2.80) | 2.65 (2.0–2.95) | 0.86 [0.68–0.94] | −0.09 [− 0.62–0.43] |
| PF | 20 | 0.40 (0.32–0.48) | 0.35 (0.30–0.38) | 0.41 [0.01–0.71] | 0.05 [− 0.13–0.24] |
D* (10–3 mm | 20 | 22 (14–30) | 19 (14–26) | 0.14 [−0.33–0.54] | −0.6 [− 35–34] |
Number of ROIs for DWI | 20 | 5 (4–7) | 13 (11–15) | ||
| Intra-Observer Data: Observer 1 Only | |||||
| Parameter | Median (IQR) first measurement | Median (IQR) second measurement | ICC [95% CI] | Bias BA-LA | |
| T2 (s) | 8 | 0.067 (0.054–0.083) | 0.060 (0.051–0.084) | 0.91 [0.65–0.98] | 0.002 [− 0.010–0.015] |
| T1 (s) | 9 | 1.0 (0.90–1.0) | 0.95 (0.8–2-1) | 0.33 [− 0.40–0.80] | 0.033 [− 0.24–0.31] |
| MTR | 10 | 0.29 (0.25–0.32) | 0.28 (0.26–0.30) | 0.32 [− 0.36–0.77] | 0.01 [− 0.07–0.9] |
ADC (10–3 mm2/s) | 10 | 3.06 (2.44–3.32) | 3.10 (2.8–3.45) | 0.85 [0.18–0.96] | − 0.21 [− 0.62–0.20] |
D (10–3 mm2/s) | 10 | 2.4 (2.02–2.72) | 2.6 (2.07–2.82) | 0.83 [0.47–0.95] | − 0.08 [− 0.72–0.56] |
| PF | 10 | 0.37 (0.32–0.45) | 0.43 (0.40–0.56) | 0.05 [− 0.31–0.55] | − 0.09 [− 0.33–0.14] |
D* (10–3 mm | 10 | 20 (12–32) | 21 (12–32) | 0.22 [−0.42–0.72] | 3.8 [− 19–26] |
Fig. 7Correlation graphs for all the measured MRI parameters showing the Test–Retest, Inter- and intra-observer data