| Literature DB >> 34069396 |
Zhanna Sagirova1, Natalia Kuznetsova2, Nana Gogiberidze1, Daria Gognieva2, Aleksandr Suvorov3, Petr Chomakhidze2, Stefano Omboni1,4, Hugo Saner1,5,6, Philippe Kopylov2.
Abstract
The availability of simple, accurate, and affordable cuffless blood pressure (BP) devices has the potential to greatly increase the compliance with measurement recommendations and the utilization of BP measurements for BP telemonitoring. The aim of this study is to evaluate the correlation between findings from routine BP measurements using a conventional sphygmomanometer with the results from a portable ECG monitor combined with photoplethysmography (PPG) for pulse wave registration in patients with arterial hypertension.Entities:
Keywords: CardioQVARK; PPG monitor; blood pressure; blood pressure measurement; photoplethysmography; portable ECG monitor; pulse wave analysis; smartphone; telemedicine
Mesh:
Year: 2021 PMID: 34069396 PMCID: PMC8158773 DOI: 10.3390/s21103525
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Characteristics of CardioQVARK device. On the left side, there is the electrode for I-lead ECG registration and on the right side the monitor for photoplethysmography PPG. The device is shown together with an example of the presentation of the recording.
Figure 2Recording of the ECG and the photoplethysmography image using the CardioQVARK® smartphone case. The index fingers are placed at the ECG electrode (left) and at the photoplethysmograph monitor (right).
Figure 3Pulse transit time and pulse wave acquired by photoplethysmography with points of interest. B1 = beginning of the pulse wave, B0 = point of maximum increase of the anterior leg, SEP = peak of the ejection pulse wave, DER3 = first positive peak of the 3th derivate, SEPMAX = first inflection of the pulse wave, SRP = peak of the reflected systolic wave, DP = peak of the diastolic wave, End = end of the pulse wave.
Figure 4Bland–Altman plot of systolic blood pressure values derived from cuff based versus cuffless measurements. The mean difference between the cuffless–cuff values for DBP was 0.35 ± 2.95 mm Hg, the 95% CI of mean difference was (0.093; 0.61), according to the Student’s t-test results, the mean was not equal to 0 (p = 0.008). Here the cuffless method also slightly underestimates DBP. The correlation between the 2 measurement techniques for DBP was r = 0.87 (Spearman’s correlation, p = 0.002) (Figure 5).
Figure 7Zonal error grid analysis.
Mean and median cuffless–cuff errors, mm Hg.
| Group and Subgroups According to Cuff Measurements | M ± SD | Me [25%; 75%] | Wilcoxon Test, |
|---|---|---|---|
| ++ | 0.499 | ||
| Whole group/DBP, mm Hg | −0.35 ± 2.96 | −0.24 [−1.55; 1.46] | 0.069 |
| <110/70 mm Hg/SBP, mm Hg | 2.66 ± 2.94 | 3.79 [2.82; 3.79] | 0.017 |
| <110/70 mm Hg/DBP, mm Hg | 2.75 ± 2.93 | 2.42 [0.86; 4.55] | 0.003 |
| 110/70–140/90 mm Hg/SBP, mm Hg | −0.17 ± 3.21 | 0.33 [−1.45; 1.84] | 0.159 |
| 110/70–140/90 mm Hg/DBP, mm Hg | −0.75 ± 2.45 | −0.36 [−1.82; 0.83] | <0.001 |
| >140/90 mm Hg/SBP, mm Hg | 0.77 ± 4.49 | 1.11 [−1.61; 4.29] | 0.140 |
| >140/90 mm Hg/DBP, mm Hg | −1.23 ± 3.18 | −2.49 [−2.66; 0.70] | 0.007 |
Reference blood pressure values in subgroups were determined based on blood pressure measurements using a cuff-based method. SBP = systolic blood pressure, DBP = diastolic blood pressure.