OBJECTIVES: Conventional probes (CPs) have been considered acceptable as diagnostic tools to measure probing pocket depth (PPD) and clinical attachment level (CAL) but are affected by multiple variables. Electronic probes (EPs) provide controlled force, digital readout and data storage in computers. The objectives were to compare the reproducibility in the measurement of PPD and CAL by CP and the newly introduced CEJ handpiece of EP and intra-examiner and inter-examiner errors done in two phases. METHODS AND MATERIAL: Selected 720 periodontal sites in 1st molar of 30 persons with chronic periodontitis ≤4 mm and >4 mm pockets were analysed by two trained investigators in two phases at 2 hours difference by CP and CEJ handpiece of EP. Standard deviation, mean difference, correlation coefficient, P value and student 't' test were done to analyse data.Entities:
Keywords: Clinical attachment level; conventional probe; electronic probe; pocket probing depth
Year: 2021 PMID: 34041063 PMCID: PMC8138386 DOI: 10.4103/jfmpc.jfmpc_1994_20
Source DB: PubMed Journal: J Family Med Prim Care ISSN: 2249-4863
Table showing the intra-examiner analysis
| Probe | Examiner | No. of sites | Phase 1 (mm±SD) | Phase 2 (mm±SD) | Mean difference 1st -2nd | Correlation coefficient | |||
|---|---|---|---|---|---|---|---|---|---|
| PPD ≤4 mm | |||||||||
| PPD | CP | Examiner 1 | 497 | 2.66±0.893 | 2.64±0.880 | 0.020±0.425 | 0.885 | 0.292 | 1.054 |
| Examiner 2 | 497 | 2.64±0.868 | 2.61±0.842 | 0.030±0.460 | 0.854 | 0.147 | 1.452 | ||
| EP | Examiner 1 | 497 | 2.66±0.891 | 2.65±0.882 | 0.012±0.291 | 0.946 | 0.355 | 0.926 | |
| Examiner 2 | 497 | 2.63±0.875 | 2.64±0.853 | 0.010±0.314 | 0.934 | 0.669 | 0.428 | ||
| CAL | CP | Examiner 1 | 497 | 2.64±1.290 | 2.62±1.280 | 0.022±0.404 | 0.951 | 0.222 | 1.223 |
| Examiner 2 | 497 | 2.63±1.270 | 2.60±1.240 | 0.024±0.507 | 0.920 | 0.288 | 1.061 | ||
| EP | Examiner 1 | 497 | 2.63±1.270 | 2.62±1.211 | 0.014±0.356 | 0.960 | 0.378 | 0.882 | |
| Examiner 2 | 497 | 2.62±1.24 | 2.60±1.180 | 0.024±0.470 | 0.926 | 0.253 | 1.145 | ||
| PPD >4 mm | |||||||||
| PPD | CP | Examiner 1 | 223 | 5.43±1.22 | 5.55±1.16 | 0.117±0.774 | 0.791 | 0.025 | 2.250 |
| Examiner 2 | 223 | 5.42±1.13 | 5.49±1.19 | 0.076±0.703 | 0.819 | 0.107 | 1.619 | ||
| EP | Examiner 1 | 223 | 5.35±1.04 | 5.34±1.061 | 0.013±0.383 | 0.933 | 0.603 | 0.521 | |
| Examiner 2 | 223 | 5.36±1.03 | 5.41±1.03 | 0.045±0.339 | 0.946 | 0.320 | 1.014 | ||
| CAL | CP | Examiner 1 | 223 | 4.52±1.77 | 4.48±1.69 | 0.049±0.705 | 0.918 | 0.297 | 1.044 |
| Examiner 2 | 223 | 4.53±1.78 | 4.49±1.71 | 0.040±0.639 | 0.934 | 0.347 | 0.943 | ||
| EP | Examiner 1 | 223 | 4.44±1.65 | 4.47±1.660 | 0.027±0.519 | 0.951 | 0.440 | 0.774 | |
| Examiner 2 | 223 | 4.45±1.64 | 4.48±1.64 | 0.031±0.488 | 0.956 | 0.337 | 0.961 |
PPD=probing pocket depth. CAL=clinical attachment level. CP=conventional probe. EP=electronic probe. SD=standard deviation. P=significant at≤0.05
Table showing the inter-examiner analysis
| Probe | Phases | No. of sites | Examiner 1 (mm±SD) | Examiner 2 (mm±SD) | Mean difference 1st-2nd | Correlation coefficient | |||
|---|---|---|---|---|---|---|---|---|---|
| PPD ≤4 mm | |||||||||
| PPD | CP | phase 1 | 497 | 2.66±0.893 | 2.64±0.868 | 0.018±0.056 | 0.919 | 0.746 | 0.324 |
| Phase 2 | 497 | 2.64±0.880 | 2.61±0.842 | 0.028±0.055 | 0.857 | 0.606 | 0.516 | ||
| EP | Phase 1 | 497 | 2.66±0.891 | 2.63±0.875 | 0.026±0.056 | 0.961 | 0.641 | 0.467 | |
| Phase 2 | 497 | 2.65±0.882 | 2.64±0.853 | 0.008±0.055 | 0.966 | 0.884 | 0.146 | ||
| CAL | CP | Phase 1 | 497 | 2.64±1.29 | 2.63±1.27 | 0.016±0.882 | 0.979 | 0.844 | 0.197 |
| Phase 2 | 497 | 2.62±1.28 | 2.60±1.24 | 0.018±0.080 | 0.970 | 0.821 | 0.226 | ||
| EP | Phase 1 | 497 | 2.63±1.27 | 2.62±1.24 | 0.010±0.080 | 0.983 | 0.900 | 0.126 | |
| Phase 2 | 497 | 2.62±1.21 | 2.60±1.18 | 0.020±0.076 | 0.965 | 0.791 | 0.265 | ||
| PPD >4 mm | |||||||||
| PPD | CP | phase 1 | 223 | 5.43±1.22 | 5.42±1.13 | 0.018±0.11 | 0.896 | 0.873 | 0.160 |
| Phase 2 | 223 | 5.55±1.16 | 5.49±1.19 | 0.058±0.111 | 0.863 | 0.601 | 0.523 | ||
| EP | Phase 1 | 223 | 5.35±1.04 | 5.36±1.03 | 0.009±0.098 | 0.937 | 0.927 | 0.091 | |
| Phase 2 | 223 | 5.34±1.04 | 5.41±1.03 | 0.067±0.099 | 0.930 | 0.498 | 0.679 | ||
| CAL | CP | Phase 1 | 223 | 4.52±1.77 | 4.53±1.78 | 0.009±0.169 | 0.990 | 0.958 | 0.053 |
| Phase 2 | 223 | 4.48±1.69 | 4.49±1.71 | 0.018±0.161 | 0.969 | 0.911 | 0.111 | ||
| EP | Phase 1 | 223 | 4.49±1.65 | 4.45±1.64 | 0.009±0.156 | 0.992 | 0.954 | 0.057 | |
| Phase 2 | 223 | 4.47±1.66 | 4.48±1.64 | 0.015±0.156 | 0.988 | 0.931 | 0.086 |
PPD=probing pocket depth. CAL=clinical attachment level. CP=conventional probe. EP=electronic probe. SD=standard deviation. P≤0.05
Table showing the inter-probe analysis within the examiner
| Examiners | Phases | No. of sites | CP (mm±SD) | EP (mm±SD) | Mean difference 1st-2nd | Standard error of mean | |||
|---|---|---|---|---|---|---|---|---|---|
| PPD ≤4 mm | |||||||||
| PPD | Examiner 1 | Phase 1 | 497 | 2.66±0.893 | 2.66±0.891 | 0.004±0.504 | 0.023 | 0.859 | 0.178 |
| Phase 2 | 497 | 2.64±0.880 | 2.65±0.882 | 0.004±0.412 | 0.018 | 0.828 | 0.218 | ||
| Examiner 2 | Phase 1 | 497 | 2.64±0.868 | 2.63±0.875 | 0.012±0.516 | 0.023 | 0.602 | 0.522 | |
| Phase 2 | 497 | 2.61±0.842 | 2.64±0.853 | 0.024±0.519 | 0.023 | 0.300 | 1.037 | ||
| CAL | Examiner 1 | Phase 1 | 497 | 2.64±1.29 | 2.63±1.270 | 0.012±0.500 | 0.022 | 0.591 | 0.538 |
| Phase 2 | 497 | 2.62±1.28 | 2.62±1.211 | 0.004±0.550 | 0.025 | 0.870 | 0.163 | ||
| Examiner 2 | phase 1 | 497 | 2.63±1.27 | 2.62±1.24 | 0.006±0.537 | 0.024 | 0.802 | 0.251 | |
| Phase 2 | 497 | 2.60±1.24 | 2.60±1.18 | 0.006±0.646 | 0.006 | 0.835 | 0.208 | ||
| PPD >4 mm | |||||||||
| PPD | Examiner 1 | Phase 1 | 223 | 5.43±1.22 | 5.35±1.04 | 0.081±0.796 | 0.053 | 0.131 | 1.515 |
| Phase 2 | 223 | 5.55±1.16 | 5.34±1.061 | 0.211±0.763 | 0.051 | 0.00 | 4.127 | ||
| Examiner 2 | Phase 1 | 223 | 5.42±1.13 | 5.36±1.03 | 0.054±0.702 | 0.047 | 0.253 | 1.145 | |
| Phase 2 | 223 | 5.49±1.19 | 5.41±1.03 | 0.085±0.842 | 0.056 | 0.132 | 1.511 | ||
| CAL | Examiner 1 | Phase 1 | 223 | 4.52±1.77 | 4.44±1.651 | 0.081±0.712 | 0.048 | 0.092 | 1.693 |
| Phase 2 | 223 | 4.48±1.69 | 4.47±1.660 | 0.004±0.780 | 0.052 | 0.932 | 0.086 | ||
| Examiner 2 | phase 1 | 223 | 4.53±1.78 | 4.45±1.64 | 0.081±0.706 | 0.047 | 0.089 | 1.708 | |
| Phase 2 | 223 | 4.49±1.71 | 4.48±1.64 | 0.009±0.741 | 0.05 | 0.857 | 0.181 |
PPD=Probing pocket depth. CAL=clinical attachment level. CP=conventional probe. EP=electronic probe. SD=standard deviation. P≤0.05