Literature DB >> 3386973

A comparison of two methods of measuring gamma camera uniformity.

D Pearson1, T Kapasi.   

Abstract

Two quantitative methods of measuring intrinsic uniformity of gamma cameras have been compared. One is a modified NEMA protocol, and the other a simpler method of data acquisition and processing that can be carried out in departments with restricted data-processing facilities. Both methods show the same trend in uniformity and lead to action being taken to rectify deterioration in performance at the same time. The use of cusum plots to monitor trends in performance and set action levels is discussed and compared with a simple plot of the data as a time series.

Mesh:

Year:  1988        PMID: 3386973

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  2 in total

1.  An evaluation of cusum analysis and control charts applied to quantitative gamma-camera uniformity parameters for automated quality control.

Authors:  A C Knight; E D Williams
Journal:  Eur J Nucl Med       Date:  1992

2.  Long-term quality control of DXA: a comparison of Shewhart rules and Cusum charts.

Authors:  D Pearson; S A Cawte
Journal:  Osteoporos Int       Date:  1997       Impact factor: 4.507

  2 in total