| Literature DB >> 33844164 |
Simone F Kleiss1, Kirsten F Ma2, Mostafa El Moumni3, Richte C L Schuurmann2, Clark J Zeebregts2, Marieke Haalboom4, Reinoud P H Bokkers5, Jean-Paul P M de Vries2.
Abstract
PURPOSE: Hyperspectral imaging (HSI) is a noninvasive spectroscopy technique for determining superficial tissue oxygenation. The HyperView™ system is a hand-held camera that enables perfusion image acquisition. The evaluation of superficial tissue oxygenation is warranted in the evaluation of patients with peripheral arterial disease. The aim was to determine the reliability of repeated HSI measurements.Entities:
Keywords: Hyperspectral imaging; Lower extremity; Reproducibility of Results; Tissue oxygenation; Tissue perfusion; Transcutaneous oxygen pressure
Mesh:
Year: 2021 PMID: 33844164 PMCID: PMC9162963 DOI: 10.1007/s10877-021-00698-w
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 1.977
Fig. 1A flowchart of study procedures demonstrating the measurements used for the test–retest reliability and intra- and inter-observer reliability, respectively. During the measurement session, two consecutive images at each location were taken according to the standardized measurement protocol. The two consecutive images from one measurement session were used to determine test–retest reliability. The first image from measurement session 1 and 2 were used to determine intra-observer reliability. The first image from measurement session 1and 3 were used to determine inter-observer reliability. The same measurement sessions were used to determine the intra- and inter-observer reliability of TcPo2 measurements. * 2 images per measurement location; plantar side right foot, right calf, plantar side left foot, left calf. ** TcPo2 of the left and right calf
Hyperspectral imaging values, TcPo2 and local skin temperature of 50 healthy volunteers at different locations of the lower leg
| Right foot | Left foot | Right vs. Left foot ( | Right calf | Left calf | Right vs. Left calf ( | |
|---|---|---|---|---|---|---|
| OxyHb (a.u.) | 85.0 ± 25.9* | 78.2 ± 23.7* | .003 | 38.9 ± 10.5* | 42.9 ± 12.1* | .001 |
| DeoxyHb (a.u.) | 64.9 ± 14.3* | 71.3 ± 15.4* | <.001 | 42.2 ± 11.6* | 40.7 ± 11.2* | .205 |
| Oxygen saturation (%) | 55.8 ± 10.9* | 51.6 ± 10.8 | <.001 | 48.2 ± 10.7* | 51.2 ± 10.4 | .008 |
| TcP | … | … | 66.7 ± 13.5 | 61.1 ± 11.3 | <.001 | |
| Skin temperature (°C) | 27.6 ± 2.7* | 27.7 ± 2.8* | .879 | 32.7 ± 0.8* | 32.4 ± 0.9* | <.001 |
HSI values are the mean from the first image from one measurement session of 50 participants. Values are reported as mean ± standard deviation. a.u. arbitrary units, TcPo transcutaneous oxygen pressure. *Statistically significant difference between foot and calf within the ipsilateral leg P < .001
Short term test–retest reliability and measurement error of two consecutive hyperspectral images taken at one measurement session
| Location | OxyHb (a.u.) | DeoxyHb (a.u.) | Oxygen saturation (%) | |
|---|---|---|---|---|
| Right foot | SEMagree | 10.1 | 4.9 | 4.3 |
| LoA | (−28.3 to 28.4) | (− 4.0 to 13.7) | (−12.2 to 12.0) | |
| SDC | 28.1 | 13.7 | 11.6 | |
| ICCagree | 0.85 | 0.89 | 0.84 | |
| 95% CI | (0.75–0.91) | (0.82–0.94) | (0.73–0.91) | |
| Right calf | SEMagree | 3.5 | 3.7 | 3.8 |
| LoA | (−9.8 to 10.0) | (−11.5 to 8.4) | (− 9.5 to 11.3) | |
| SDC | 9.8 | 10.3 | 10.5 | |
| ICCagree | 0.89 | 0.90 | 0.88 | |
| 95% CI | (0.81–0.94) | (0.83–0.94) | (0.80–0.93) | |
| Left foot | SEMagree | 9.6 | 6.0 | 5.2 |
| LoA | (−28.0 to 25.2) | (−13.9 to 18.4) | (−15.6 to 12.7) | |
| SDC | 26.5 | 16.6 | 14.3 | |
| ICCagree | 0.82 | 0.84 | 0.75 | |
| 95% CI | (0.71–0.89) | (0.74–0.91) | (0.61–0.85) | |
| Left calf | SEMagree | 5.3 | 4.3 | 5.3 |
| LoA | (−13.4 to 15.7) | (−13.0 to 10.6) | (− 12.8 to 16.0) | |
| SDC | 14.6 | 11.9 | 14.6 | |
| ICCagree | 0.82 | 0.84 | 0.72 | |
| 95% CI | (0.70–0.89) | (0.73–0.91) | (0.56–0.83) |
Values are determined with the two subsequent images from one measurement session for 50 participants. SEM standard error of measurement for absolute agreement, LoA limits of agreement, ICC Intraclass correlation for absolute agreement, 95% CI 95% Confidence intervals for the ICC, SDC smallest detectable change
Intra-observer reliability and measurement error for HSI at two different measurement sessions performed by one observer together with TcPo2 measurements from these sessions
| Location | OxyHb (a.u.) | DeoxyHb (a.u.) | Oxygen saturation (%) | TcP | |
|---|---|---|---|---|---|
| Right foot | SEMcons | 21.2 | 8.7 | 8.6 | |
| LoA | (−59.5 to 57.9) | (−29.3 to 19.4) | (−22.1 to 25.5) | ||
| SDC | 58.7 | 24.3 | 23.8 | ||
| ICCcons | 0.24 | 0.71 | 0.33 | ||
| 95% CI | (−0.04–0.49) | (0.54–0.82) | (0.06–0.56) | ||
| Right calf | SEMcons | 6.1 | 6.9 | 6.9 | 9.1 |
| LoA | (−19.0 to 14.9) | (−20.3 to 18.1) | (−19.7 to 18.8) | (−25.0 to 25.7) | |
| SDC | 16.9 | 19.2 | 19.2 | 25.1 | |
| ICCcons | 0.62 | 0.59 | 0.57 | 0.54 | |
| 95% CI | (0.42–0.77) | (0.38–0.75) | (0.34–0.73) | (0.49–0.59) | |
| Left foot | SEMcons | 17.3 | 10.0 | 7.8 | |
| LoA | (−50.3 to 45.7) | (−33.6 to 21.8) | (−20.0 to 23.1) | ||
| SDC | 48.0 | 27.7 | 21.6 | ||
| ICCcons | 0.41 | 0.59 | 0.37 | ||
| 95% CI | (0.16–0.62) | (0.37–0.74) | (0.11–0.59) | ||
| Left calf | SEMcons | 7.4 | 8.2 | 8.8 | 6.9 |
| LoA | (−19.2 to 21.6) | (−26.0 to 19.2) | (−21.7 to 27.2) | (− 19.2 to 19.8) | |
| SDC | 20.4 | 22.6 | 24.5 | 19.2 | |
| ICCcons | 0.46 | 0.57 | 0.44 | 0.56 | |
| 95% CI | (0.20–0.65) | (0.34–0.73) | (0.18–0.64) | (0.52–0.60) |
Values are determined with the first image from the first session with the first image from the second session for 50 participants. SEM standard error of measurement for consistency, LoA limits of agreement, ICC Intraclass correlation for consistency, 95% CI 95% Confidence intervals for the ICC, TcPo transcutaneous oxygen pressure, SDC smallest detectable change
Inter-observer reliability and measurement error for HSI at two different measurement sessions performed by two observers together with TcPo2 measurements from these sessions
| Location | OxyHb (a.u.) | DeoxyHb (a.u.) | Oxygen saturation (%) | TcP | |
|---|---|---|---|---|---|
| Right foot | SEMagree | 15.9 | 12.0 | 8.1 | |
| LoA | (−39.2 to 48.3) | (−37.1 to 16.7) | (−14.2 to 25.5) | ||
| SDC | 44.2 | 33.3 | 22.6 | ||
| ICCagree | 0.58 | 0.43 | 0.45 | ||
| 95% CI | (0.36–0.74) | (0.07–0.67) | (0.16–0.66) | ||
| Right calf | SEMagree | 8.0 | 9.0 | 9.1 | 9.7 |
| LoA | (−22.3 to 22.4) | (−26.4 to 23.8) | (–24.7 to 26.4) | (–24.6 to 29.1) | |
| SDC | 22.1 | 25.0 | 25.3 | 26.8 | |
| ICCagree | 0.33 | 0.32 | 0.26 | 0.52 | |
| 95% CI | (0.05–0.55) | (0.05–0.55) | (–0.30 to 0.50) | (0.47–0.58) | |
| Left foot | SEMagree | 18.0 | 11.1 | 8.3 | |
| LoA | (–46.8 to 53.4) | (–34.5 to 16.5) | (–17.1 to 26.2) | ||
| SDC | 50.0 | 30.8 | 23.2 | ||
| ICCagree | 0.30 | 0.48 | 0.30 | ||
| 95% CI | (0.03–0.53) | (0.13–0.70) | (0.05–0.53) | ||
| Left calf | SEMagree | 6.9 | 7.6 | 8.5 | 10.0 |
| LoA | (–15.8 to 21.4) | (–23.8 to 13.4) | (−16.6 to 26.7) | (−23.5 to 31.3) | |
| SDC | 19.2 | 21.1 | 23.6 | 27.6 | |
| ICCagree | 0.46 | 0.57 | 0.42 | 0.31 | |
| 95% CI | (0.22–0.65) | 0.28–0.75) | (0.15–0.62) | (0.21–0.45) |
Values are determined with the first image from measurement session #1 performed by observer #1 and the first image from measurement session #3 performed by observer #2 for 50 participants. SEM standard error of measurement for agreement, LoA limits of agreement, ICC Intra Class Correlation for absolute agreement, 95% CI 95% Confidence intervals for the ICC, TcPo transcutaneous oxygen pressure, SDC smallest detectable change
Intra-observer reliability based on generalizability and decision studies of HSI measurements
| OxyHb | DeoxyHb | Oxygen saturation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Right foot | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.40* | 0.52 | 0.62 | 0.67 | 0.65* | 0.38* | 0.43 | 0.49 | 0.51 | |||
| 2 | 0.57 | 0.68 | 0.55 | 0.61 | 0.65 | 0.68 | ||||||
| 4 | 0.73 | 0.71 | ||||||||||
| 6 | ||||||||||||
| Left foot | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.40* | 0.49 | 0.58 | 0.63 | 0.60* | 0.72 | 0.35* | 0.43 | 0.51 | 0.55 | ||
| 2 | 0.57 | 0.66 | 0.74 | 0.52 | 0.60 | 0.67 | 0.71 | |||||
| 4 | 0.73 | 0.68 | ||||||||||
| 6 | ||||||||||||
| Right calf | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.37* | 0.54 | 0.70 | 0.65* | 0.65* | |||||||
| 2 | 0.46 | 0.63 | ||||||||||
| 4 | 0.52 | 0.68 | ||||||||||
| 6 | 0.55 | 0.71 | ||||||||||
| Left calf | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.55* | 0.71 | 0.69* | 0.57* | 0.71 | |||||||
| 2 | 0.71 | 0.73 | ||||||||||
| 4 | ||||||||||||
| 6 | ||||||||||||
HSI hyperspectral imaging, nm alternative number of measurements, no alternative number of observers. *generalizability coefficients determined with all measurements performed by the same observer. Coefficients in bold exceed 0.75, indicating good reliability. OxyHb Oxyhemoglobin, DeoxyHb deoxyhemoglobin
Inter-observer reliability based on generalizability and decision studies of HSI measurements
| OxyHb | DeoxyHb | Oxygen saturation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Right foot | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.42* | 0.59 | 0.74 | 0.53* | 0.69 | 0.36* | 0.52 | 0.69 | ||||
| 2 | 0.54 | 0.70 | 0.62 | 0.41 | 0.58 | 0.73 | ||||||
| 4 | 0.64 | 0.69 | 0.46 | 0.63 | ||||||||
| 6 | 0.70 | 0.72 | 0.48 | 0.65 | ||||||||
| Left foot | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.27* | 0.43 | 0.60 | 0.69 | 0.59* | 0.74 | 0.31* | 0.48 | 0.65 | |||
| 2 | 0.35 | 0.52 | 0.68 | 0.71 | 0.39 | 0.56 | 0.72 | |||||
| 4 | 0.42 | 0.59 | 0.74 | 0.46 | 0.63 | |||||||
| 6 | 0.45 | 0.62 | 0.50 | 0.66 | ||||||||
| Right calf | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.48* | 0.65 | 0.45* | 0.62 | 0.39* | 0.56 | 0.72 | |||||
| 2 | 0.65 | 0.55 | 0.71 | 0.48 | 0.65 | |||||||
| 4 | 0.61 | 0.54 | ||||||||||
| 6 | 0.64 | 0.56 | ||||||||||
| Left calf | ||||||||||||
| 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | 1 | 2 | 4 | 6 | |
| 1 | 0.55* | 0.71 | 0.59* | 0.74 | 0.48* | 0.65 | ||||||
| 2 | 0.71 | 0.70 | 0.61 | |||||||||
| 4 | 0.70 | |||||||||||
| 6 | 0.74 | |||||||||||
HSI hyperspectral imaging, nm alternative number of measurements, no alternative number of observers. *Generalizability coefficients determined with all measurements performed by different observers. Coefficients in bold exceed 0.75, indicating good reliability. OxyHb Oxyhemoglobin, DeoxyHb Deoxyhemoglobin