Haedi E DeAngelis1, Anne M Grillet1, Martin B Nemer1, Maryla A Wasiolek1, Don J Hanson1, Michael A Omana1, Andres L Sanchez1, David W Vehar1, Paul M Thelen1.
Abstract
In response to personal protective equipment (PPE) shortages in the United States due to the Coronavirus Disease 2019, two models of N95 respirators were evaluated for reuse after gamma radiation sterilization. Gamma sterilization is attractive for PPE reuse because it can sterilize large quantities of material through hermetically sealed packaging, providing safety and logistic benefits. The Gamma Irradiation Facility at Sandia National Laboratories was used to irradiate N95 filtering facepiece respirators to a sterilization dose of 25 kGy(tissue). Aerosol particle filtration performance testing and electrostatic field measurements were used to determine the efficacy of the respirators after irradiation. Both respirator models exhibited statistically significant decreases in particle filtering efficiencies and electrostatic potential after irradiation. The largest decrease in capture efficiency was 40-50% and peaked near the 200 nm particle size. The key contribution of this effort is correlating the electrostatic potential change of individual filtration layer of the respirator with the decrease filtration efficiency after irradiation. This observation occurred in both variations of N95 respirator that we tested. Electrostatic potential measurement of the filtration layer is a key indicator for predicting filtration efficiency loss.Entities:
Mesh:
Year: 2021 PMID: 33831014 PMCID: PMC8031388 DOI: 10.1371/journal.pone.0248859
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Alanine dosimetry measurements.
| Pellet ID | Target Dose (kGy (CaF2:Mn)) | Spectral Titration | Measured Dose (kGy (Tissue)) | Estimated Uncertainty of Measured Dose |
|---|---|---|---|---|
| 1 | 25 | 281.46 | 28.5 | 4.2% |
| 2 | 25 | 279.41 | 28.2 | 4.2% |
| 3 | 25 | 278.33 | 28.1 | 4.2% |
| 4 | 25 | 274.27 | 27.6 | 4.2% |
| 5 | 25 | 271.99 | 27.3 | 4.2% |
| 6 | 25 | 270.26 | 27.0 | 4.2% |
Efficiency reduction of Model 1 respirator.
| 1 | -42 (5) |
| 2 | -45 (6) |
Model 1 respirator efficiency reduction measurements with particulate size 75nm for the FPT. Standard deviation in parentheses.
Trek 344 electrostatic potential measurement of various layers within respirators.
| 1 | 1 (Outer) | -2000 (0) | -2000 (0) | N/A |
| 2 | ||||
| 3 (Inner) | 0 (0) | 0 (0) | - | |
| 2 | 1 (Outer) | -1560 (590) | -630 (450) | 60% |
| 2 | ||||
| 3 | -1950 (100) | 1040 (760) | 153% | |
| 4 (Inner) | 0 (3) | -3 (2) | - |
Standard deviation given in parenthesis.
Trek 821HH electrostatic potential measurement of various layers within respirators.
| 1 | 1 (Outer) | -2040 (520) | -1374 (150) | 33% |
| 3 (Inner) | 1 (1) | 0 (1) | - | |
| 2 | 1 (Outer) | -1100 (160) | -320 (90) | 71% |
| 3 | -930 (80) | 820 (290) | 171% | |
| 4 (Inner) | 5 (2) | 7 (4) | - |
Standard deviation given in parenthesis.