| Literature DB >> 33816251 |
Michelle Lis1,2, Wayne Newhauser2,3, Marco Donetti4, Moritz Wolf1, Timo Steinsberger1,5, Athena Paz1, Marco Durante1,5, Christian Graeff1.
Abstract
BACKGROUND: Quality management and safety are integral to modern radiotherapy. New radiotherapy technologies require new consensus guidelines on quality and safety. Established analysis strategies, such as the failure modes and effects analysis (FMEA) and incident learning systems have been developed as tools to assess the safety of several types of radiation therapies. An extensive literature documents the widespread application of risk analysis methods to photon radiation therapy. Relatively little attention has been paid to performing risk analyses of nascent radiation therapy systems to treat moving tumors with scanned heavy ion beams. The purpose of this study was to apply a comprehensive safety analysis strategy to a motion-synchronized dose delivery system (M-DDS) for ion therapy.Entities:
Keywords: 4D therapy; carbon ion therapy; failure modes and effects analysis; motion-mitigation; motion-synchronized dose delivery; patient safety; quality assurance
Year: 2021 PMID: 33816251 PMCID: PMC8018284 DOI: 10.3389/fonc.2021.620388
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Numerical scale used to assign rank values to Severity Index, Occurrence Index, and Detectability Index for each failure.
| Rank value | Severity Index | Occurrence Index (mean time between failure) | Detectability Index |
|---|---|---|---|
| 1 | No effect on patient care | + 4 years | Impossible to miss |
| 2 | Inconvenience or delay in care | 2 years | “ |
| 3 | “ | 1 | Highly likely to notice |
| 4 | Minor dosimetric error | 6 months | Easy to detect |
| 5 | “ | 1 month | Fairly easy to detect |
| 6 | Limited toxicity or overdose | 2 weeks | Fairly difficult to detect |
| 7 | Potentially serious under- or overdose or toxicity | 1 week | “ |
| 8 | “ | 3 days | Nearly undetectable |
| 9 | Very serious under- or overdose or toxicity | 1 day | “ |
| 10 | Patient death | 1 hour | Undetectable |
These data have been adapted from the TG-100 report on failure modes and effects analysis (FMEA). For each case, a rank of 1 was considered harmless, and a rank of 10 was catastrophic. Chosen rank values were based on observed or estimated.
Description of the purpose and pass criteria for each quality assurance test.
| Test type | Quantity tested | Pass criteria |
|---|---|---|
|
| ||
| Field uniformity | Homogeneity index ( | ≥95% |
| Spot shape | FWHM in X and Y direction across scan field ( | Symmetrical |
| Spot position | Distance to agreement ( | < ± 1 mm |
| Motion-monitoring system functionality | Function test | Functioning |
|
| ||
| Beam monitor calibration reproducibility | Coefficient of variation ( | <1% |
|
| ||
| Dose distribution with homogeneous phantom | Percent error from expected dose ( | <5% |
| Dose distribution with heterogeneous phantom | Percent error from expected dose ( | <5% |
|
| ||
| Motion-monitoring system performance | Distance to agreement ( | <0.1 mm |
Figure 1Experimental setup for patient-specific quality assurance (PSQA) measurements with (A) a water tank and (B) a linear stage mounted on top. The linear stage is programmed to move a variety of attachments in periodic, respiratory-like motion patterns. Here, (C) a holder with (D) 12 small-sized ionization chambers (IC) inserted inside is attached. This IC holder aligns with the isocenter markings on the water tank phantom, which is filled with water.
Figure 2Schematic of the ionization chambers (IC) detector array setup. The IC array detector is placed inside of a 5 mm polymethylmethacrylate (PMMA) holder, mounted onto a linear stage. Water-equivalent plastic of thickness “d,” corresponding to the distal, middle or proximal depths of a target, are then placed in front of the detector.
Figure 3Schematic of the film stack. The film stack contains up to nine slabs of polymethylmethacrylate (PMMA) with precisely machined slots for holding radiochromic films (5” × 4”) as well as several slabs of additional PMMA to vary the delivery depth. The phantom measures 15 × 13 × 9 cm3 when nine slabs are in place. The slabs are aligned and held together by plastic screws at each corner of the phantom. The lateral cutouts (blue circle) are used for easy access of the irradiated films (yellow) without the need to disassemble the film stack phantom.
Figure 4Process map for motion-synchronized dose delivery as commonly found in ion therapy clinics. The treatment process is broken down into five categories: imaging, treatment planning, plan verification, initial treatment fraction, and subsequent treatments. Each step in these processes is described.
Summary of failure modes and effects analysis (FMEA) results for potential errors during patient therapy with motion-synchronized ion beams using a DDS with integrated motion compensation controls. Risk priority numbers (RPNs) of over 125 were considered high risk.
| Failure mode | Severity | Occurrence | Detectability | RPN |
|---|---|---|---|---|
| Patient movement | 7 | 6 | 7 | 294 |
| Absolute change in breathing position | 6 | 7 | 7 | 294 |
| Patient alignment | 7 | 5 | 6 | 210 |
| No gating during sporadic movements | 5 | 5 | 8 | 200 |
| Gating window too large | 6 | 8 | 4 | 192 |
| Beam sweeping distance dose | 3 | 8 | 9 | 192 |
| Sending incorrect motion phase info | 7 | 3 | 7 | 147 |
| Error creating of slice files | 7 | 3 | 7 | 147 |
| Failure to gate | 9 | 2 | 8 | 144 |
| Position calibration incorrect | 8 | 2 | 7 | 112 |
| Changes to respiration pattern | 7 | 8 | 2 | 112 |
| Setup of motion management device to wrong position | 9 | 2 | 6 | 108 |
| Patient not re-imaged after anatomy change | 7 | 3 | 5 | 105 |
| Position feedback missing/not working | 4 | 5 | 5 | 100 |
| Determined wrong number of motion phases | 6 | 2 | 5 | 90 |
| Incorrect motion direction | 9 | 2 | 5 | 90 |
| Loading wrong treatment plan | 4 | 2 | 10 | 80 |
| Motion data recording stops | 4 | 2 | 9 | 72 |
| Failure to progress to next slice | 6 | 2 | 5 | 60 |
| Plan incomplete | 6 | 2 | 5 | 60 |
| Motion signal loss | 9 | 2 | 3 | 54 |
Figure 5A fault tree analysis for the case of gating failures during motion-synchronized dose deliveries.
Summary of pass-fail tests and results.
| Error | Expected action | Result |
|---|---|---|
|
| ||
| Wrong number of beam spots in plan library | Setup error state | Passed |
| Missing motion information in plan header | “ | “ |
| Particle numbers below or above limitations | “ | “ |
| Plan library larger than size limitations | “ | “ |
|
| ||
| Motion signal lost | Beam aborted | “ |
| Motion trajectory deviating from expected trajectory | Temporary gate | “ |
| Scanning magnet failure | Interlock | “ |
| Gating magnet failure | “ | “ |
| Delivery of a beam spot skipped | Treatment is halted | “ |
| MMD file recording error | Treatment stops, errors message, and file dump | “ |
| Treatment stops prematurely | Delivery data recorded | “ |
| Motion calibration incomplete before delivery starts | Beam gate activated | “ |
|
| ||
| Wrong motion compensation strategy selected | Set automatically from plan | “ |
| Motion system not fully set up or not on | Delivery cannot start | “ |
Figure 6(A) Correlation of motion position to the measured signal of the motion-monitoring device (an optical distance laser sensor) for a step-wise motion pattern. The motion positions, in mm, (red) are uploaded onto the linear stage as a motion file and used to move the linear stage. The motion-monitoring device then records the relative displacement (blue) as an analog signal. The left and right portions of the motion signal show where the motion-monitoring device is out of signal range. (B) A plot of position accuracy between the optical distance laser signal (in arbitrary units) and the linear stage position (in mm).
Summary of patient specific quality assurance results for four measurements.
| QA test | Metric (pass criteria) | Static results | Motion mitigation results | |
|---|---|---|---|---|
| Pinpoints | Dose deviation (± 5%) | ± 2.4% | ± 8.9% | |
| Log files (planned to reconstructed) | Gamma index analysis (>90%) | 99.4% | 96.0% | |
| Film stacks | Gamma index analysis (>90%) | 92.4% | 90.4% | |
| IC detector to log file reconstructions | Gamma index analysis (>90%) |
| 84.1% (0.9883) | 91.6% (0.9901) |
|
| 100.0% (0.9947) | 98.6% (0.9954) | ||
|
| 99.4% (0.9983) | 90.9% (0.9971) | ||
These measurements included calculating 3D dose measurement agreement with 12 small-sized ionization chambers (IC), and calculating gamma index analysis pass rates for comparisons between log file reconstructions and treatment plans as well as IC array detector measurements. Acceptable criteria and analysis results are summarized.