Juha Koivisto1, Maureen van Eijnatten1, Timo Kiljunen2, Xie-Qi Shi3,4, Jan Wolff1.
Abstract
The objective of the present study was to assess and compare the effective doses in the wrist region resulting from conventional radiography device, multislice computed tomography (MSCT) device and two cone beam computed tomography (CBCT) devices using MOSFET dosemeters and a custom made anthropomorphic RANDO phantom according to the ICRP 103 recommendation. The effective dose for the conventional radiography was 1.0 μSv. The effective doses for the NewTom 5 G CBCT ranged between 0.7 μSv and 1.6 μSv, for the Planmed Verity CBCT 2.4 μSv and for the MSCT 8.6 μSv. When compared with the effective dose for AP- and LAT projections of a conventional radiographic device, this study showed an 8.6-fold effective dose for standard MSCT protocol and between 0.7 and 2.4-fold effective dose for standard CBCT protocols. When compared to the MSCT device, the CBCT devices offer a 3D view of the wrist at significantly lower effective doses.Entities: