Riccardo Di Liberto1, Daniele Andreuccetti2, Moreno Comelli2, Giancarlo Burriesci3.
Abstract
This work is the result of a campaign of measures of exposure levels to magnetic field gradients (GMF) generated by magnetic resonance imaging (MRI) tomographs, to which both healthcare staff and any persons accompanying patients who remain inside the magnet room are exposed while performing a diagnostic Investigation. The study was conducted on three MRI tomographs with a static magnetic induction field up to 1.5 T installed in two hospitals of Lombardy. The study aims to characterize electromagnetic emissions within the magnet room and the definition of a measurement method suitable for assessing the level of exposure of healthcare personnel and any persons accompanying patients. The measurements performed concerned the determination of the weighted peak index for magnetic induction, due to the diagnostic GMF, relating to the action levels for the workers and the reference levels for the general population, in force in the European Union. Thanks to the defined experimental setup, the use of two different measuring instruments, and the software resources of the WEBNIR platform, it was possible to identify, for both categories of exposed persons, the "clearance" space, i.e., the distance from the magnet of the tomograph that guarantees health protection concerning the exposure to GMF, according to the indications of the standards in force. The method used showed that the exposure levels to GMF are substantially safe for professionally exposed workers who do not carry specific risks. For workers particularly sensitive to the specific risk, as well as to individuals part of the population, it is however advisable to maintain a distance from the magnet of about one meter to prevent sensorial neuromuscular stimulation effects.Entities:
Keywords: MRI gradient magnetic field measurement; clearance zone; exposure levels; risk assessment
Year: 2021 PMID: 33801598 PMCID: PMC8037291 DOI: 10.3390/ijerph18073475
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
European Directive 2013/35/EU—Action Levels for exposure to magnetic fields from 1 Hz to 10 MHz.
| Frequency range | Magnetic Flux Density Low ALS (μT) (RMS) | Magnetic Flux Density High ALS (μT) (RMS) | Magnetic Flux Density ALS for Exposure of Limbs to a Localised Magnetic Field (μT) (RMS) |
|---|---|---|---|
| 1 ≤ | 2.0 × 105/ | 3.0 × 105/ | 9.0 × 105/ |
| 8 ≤ | 2.5 × 104/ | 3.0 × 105/ | 9.0 × 105/ |
| 25 ≤ | 1.0 × 103 | 3.0 × 105/ | 9.0 × 105/ |
| 300 Hz ≤ | 3.0 × 105/ | 3.0 × 105/ | 9.0 × 105/ |
| 3 kHz ≤ | 1.0 × 102 | 1.0 × 102 | 3.0 × 102 |
European Recommendation 1999/519/EC—Reference levels for electric, magnetic and electromagnetic fields (0 Hz to 300 GHz, unperturbed RMS values).
| Frequency Range | E-Field Strength (V/m) | H-Field Strength(A/m) | B-Field(μT) | Equivalent Plane WavePower Density Seq (W/m2) |
|---|---|---|---|---|
| 0–1 Hz | — | 3.2 × 104 | 4 × 104 | — |
| 1–8 Hz | 10,000 | 3.2 × 104/ | 4 × 104/ | — |
| 8–25 Hz | 10,000 | 4000/ | 5000/ | — |
| 0.025–0.8 kHz | 250/ | 4/ | 5/ | — |
| 0.8–3 kHz | 250/ | 5 | 6.25 | — |
| 3–150 kHz | 87 | 5 | 6.25 | — |
| 0.15–1 MHz | 87 | 0.73/ | 0.92/ | — |
| 1–10 MHz | 87/ | 0.73/ | 0.92/ | — |
| 10–400 MHz | 28 | 0.073 | 0.092 | 2 |
| 400–2000 MHz | 1.375 | 0.0037 | 0.0046 | |
| 2–300 GHz | 61 | 0.16 | 0.20 | 10 |
Figure 1Position of the measuring points: the “OP” and “REAR” points and the measuring points along: (a) Real setup; (b) Explanatory scheme.
Tomographs and GMF diagnostic sequences analysed; the “Code” column shows the reference with which the sequences are indicated in the result tables.
| Tomograph | Gradient Sequence | |
|---|---|---|
| Code | Description | |
| (A) Philips Panorama 1 T | A1 | SPIN-ECHO, Repetition Time 729 ms, Slice Gap 0.35 mm. |
| A2 | STIR, Repetition time 2500 ms, Inversion Time 120 ms, Echo time 70 ms. | |
| (B) Philips Ingenia 1.5 T | B1 | SPAIR Lumbar Spine 200 mm, Echo Time 80 ms, Repetition Time 4179 ms, Suppression of fat. |
| B2 | STIR Long TE Lumbar Spine (Sagittal-Lumbar), Echo Time 60 ms, Repetition Time 2500–4000 ms, Inversion Time 160 ms. | |
| (C) Siemens Magnetom Aera 1.5 T | C1 | T1 Coronal TIRM P2-320 Bilateral, Repetition Time 5110 ms, Echo time 52 ms, Slice Thickness 4.5 mm. |
| C2 | TURBO SPIN ECHO FatSat 256 bilateral, Repetition Time 481 ms, Echo Time 7.3 ms, Slice Thickness 4.5 mm. | |
Figure 2Instrumental chains for GFM measurement: (a) Narda ELT-400 with acquisition system; (b) Narda ELT-400 in a stand-alone version; (c) Narda-PMM EHP-50F.
Measurement results with the ELT-400 probe (Display values—Mode Exposure STD ICNIRP 1998 General Public—Range Low—Low Cut 30 Hz—Detector STND) and EHP-50F (Mode WP, magnetic field, full scale 100 µT).
| Sequence | Position | Probe ELT-400 | Probe EHP-50F | |||
|---|---|---|---|---|---|---|
| Distance (cm) | Height (cm) | RLs 1999/519/EC | RLs 1999/519/EC | Low ALs 2013/35/EU | High ALs 2013/35/EU | |
| A1 | 0 | 100 | 46.6% | |||
| A2 | 0 | 76.0% | 70.0% | 2.0% | 1.5% | |
| 50 | 7.0% | 9.0% | 0.3% | 0.3% | ||
| 100 | 2.4% | 2.4% | 0.1% | 0.1% | ||
| 150 | 1.7% | |||||
Measurement results with ELT-400 probe (postprocessing of data acquired from the analogue outputs—Mode Field Strength 320 μT—Range Low—Low Cut 30 Hz).
| Sequence | Distance (cm) | Height (cm) | RLs 1999/519/EC | Low ALs 2013/35/EU | High ALs 2013/35/EU |
|---|---|---|---|---|---|
| A2 | 0 | 100 | 69.5% | 3.36% | 3.56% |
Measurement results with the EHP-50F probe (WP mode, magnetic field, full scale of 100 µT).
| Sequence | Distance (cm) | Height (cm) | RLs 1999/519/EC | Low ALs 2013/35/EU | High ALs 2013/35/EU |
|---|---|---|---|---|---|
| B1 | 0 | 100 | 1389% | 45.0% | 45.6% |
| OP | 22.5% | 0.70% | 0.60% | ||
| B2 | 0 | 100 | 2356% | 76.0% | 72.5% |
| 40 | 401% | 13.0% | 12.0% | ||
| 100 | 83% | 2.50% | 2.10% | ||
| OP | 51% | 1.57% | 1.25% | ||
| 0 | 160 | 1332% | 40.1% | 40.8% | |
| 40 | 317% | 9.70% | 9.17% | ||
| 100 | 66% | 2.00% | 1.65% |
Measurements results with the ELT-400 probe (Postprocessing of the data acquired from the analogue outputs—Mode Field Strength 320 μT—Range Low—Low Cut 30 Hz).
| Sequence | Distance (cm) | Height (cm) | RLs 1999/519/EC | Low ALs 2013/35/EU | High ALs 2013/35/EU |
|---|---|---|---|---|---|
| B1 | 0 | 100 | 1423% | 52.4% | 53.2% |
Figure 3Interpolation of the B2 sequence at different distances and heights from the ground: (a) 100 cm from the ground (b) 160 cm from the ground.
Measurement results with the EHP-50F probe (WP mode, magnetic field, full scale 100 µT).
| Sequence | Distance (cm) | Height (cm) | RLs 1999/519/EC | Low ALs 2013/35/EU | High ALs 2013/35/EU |
|---|---|---|---|---|---|
| C1 | 0 | 100 | 2032% | 78.9% | 86.5% |
| C2 | 0 | 100 | 2642% | 88.0% | 92.0% |
| 50 | 338% | 11.4% | 10.7% | ||
| 100 | 100% | 3.36% | 2.82% | ||
| 150 | 44.4% | 1.51% | 1.21% | ||
| REAR | 916% | 31.0% | 29.5% |