| Literature DB >> 33768766 |
András Herein1,2, Gábor Stelczer1,2, Csilla Pesznyák1,2, Georgina Fröhlich1,3, Viktor Smanykó1, Norbert Mészáros1,4, Csaba Polgár1,4, Tibor Major1,4.
Abstract
BACKGROUND: The aim of the study was to dosimetrically compare multicatheter interstitial brachytherapy (MIBT) and stereotactic radiotherapy with CyberKnife (CK) for accelerated partial breast irradiation (APBI) especially concerning the dose of organs at risk (OAR-s). PATIENTS AND METHODS: Treatment plans of thirty-two MIBT and CK patients were compared. The OAR-s included ipsilateral non-target and contralateral breast, ipsilateral and contralateral lung, skin, ribs, and heart for left-sided cases. The fractionation was identical (4 x 6.25 Gy) in both treatment groups. The relative volumes (e.g. V100, V90) receiving a given relative dose (100%, 90%), and the relative doses (e.g. D0.1cm3, D1cm3) delivered to the most exposed small volumes (0.1 cm3, 1 cm3) were calculated from dose-volume histograms. All dose values were related to the prescribed dose (25 Gy).Entities:
Keywords: APBI; CyberKnife; dosimetric comparison; multicatheter interstitial brachytherapy
Year: 2021 PMID: 33768766 PMCID: PMC8042824 DOI: 10.2478/raon-2021-0016
Source DB: PubMed Journal: Radiol Oncol ISSN: 1318-2099 Impact factor: 2.991
Figure 1Representative DVH-s for multicatheter interstitial brachytherapy (MIBT) (solid line) and CyberKnife (CK) treatment (dashed line) for accelerated partial breast irradiation. The prescribed dose was 25 Gy. The target volume is indicated with red, the non-target breast with yellow, the ipsilateral lung with blue and the heart with orange. The solid red line clearly shows that with MIBT volumes irradiated by high doses (> 25 Gy) develop within the target volume.
Tumour localization according to breast quadrants in patients treated by multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK)
| Method of treatment | Laterality | Upper outer | Lower outer | Upper inner | Lower inner | Central |
|---|---|---|---|---|---|---|
| MIBT | right | 3 | 1 | 3 | 2 | 4 |
| left | 3 | 3 | 4 | 1 | 8 | |
| CK | right | 6 | 6 | 2 | 1 | 2 |
| left | 2 | 4 | 1 | 3 | 5 |
Absolute volumes of the body irradiated by 100%, 50% and 20% of the prescribed dose for multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK) treatments
| Body | MIBT | CK | p-value* |
|---|---|---|---|
| V100 (cm3) | 66.4 (28.9−193.8) | 77.1 (18.3−141.5) | 0.0356 |
| V50 (cm3) | 159.0 (73.4−444.2) | 189.3 (58.6−397.6) | 0.0184 |
| V20 (cm3) | 470.7 (253.4−1078.1) | 609.1 (326.7−1160.8) | 0.0013 |
*Mann-Whitney U test
Figure 2Representative dose distribution for multicatheter interstitial brachytherapy (MIBT) (A) and CyberKnife (CK) treatment (B). PTV: red, ipsilateral breast: yellow, contralateral breast: pink, ribs: green, heart: orange, ipsilateral lung: dark blue, contralateral lung: light blue.
Dosimetry of ipsilateral breast for multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK) treatments. The values are given in percentages
| MIBT | CK | p-value* | |
|---|---|---|---|
| V100 | 9.7 (2.2−29.7) | 9.6 (0.9−20.4) | 0.9381 |
| V50 | 23.2 (5.1−68.0) | 23.3 (2.7−47.1) | 0.9822 |
| V100 | 1.6 (0.4−4.2) | 0.7 (0.0−2.0) | < 0.0001 |
| V90 | 2.8 (0.6−8.1) | 2.3 (0.0−4.4) | 0.7019 |
| V50 | 12.9 (3.1−35.4) | 10.5 (1.7−18.9) | 0.5501 |
| V25 | 31.8 (9.4−68.4) | 29.9 (7.6−55.2) | 0.8772 |
*Mann-Whitney U test
Dosimetry of ipsilateral lung for multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK) treatments. The values are given in percentages
| Ipsilateral lung | MIBT | CK | p-value* |
|---|---|---|---|
| Mean lung dose | 4.9 (1.9−11.1) | 4.9 (1.6−11.8) | 0.9946 |
| D0.1cm3 | 41.0 (10.1−61.5) | 52.2 (15.6−90.5) | 0.0068 |
| D1cm3 | 36.1 (8.7−55.8) | 45.4 (14.4−84.5) | 0.0162 |
| D2cm3 | 33.6 (8−52.5) | 42.2 (13.7−80.5) | 0.0198 |
| D2 | 21.2 (7.4−32.9) | 24 (8.8−46.4) | 0.2454 |
| V5 | 30.6 (5.1−50.0) | 30.7 (5.8−63.2) | 0.9357 |
*Mann-Whitney U test
Dosimetry of heart for multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK) treatments. The values are given in percentages
| Heart | MIBT | CK | p-value* |
|---|---|---|---|
| Mean heart dose | 4.1 (1.0−7.7) | 3.3 (1.3−6.9) | 0.1706 |
| D0.1cm3 | 21.4 (4.0−47.2) | 18.9 (8.4−43.2) | 0.2822 |
| D1cm3 | 18.2 (3.2−42.3) | 16.6 (7.3−40.3) | 0.4878 |
| D2cm3 | 17 (3.0−40.0) | 15.6 (6.8−38.3) | 0.5324 |
| D2 | 13.3 (4.3−26.4) | 12.5 (5.2−31.5) | 0.5098 |
| V5 | 29.9 (0.0−62.5) | 21.2 (6.1−41.0) | 0.1653 |
*Mann-Whitney U test
Dosimetry of skin and ribs for multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK) treatments. The values are given in percentages except for V100 and V50
| Skin | MIBT | CK | p-value* |
|---|---|---|---|
| D0.1cm3 | 75.3 (18.0−164.2) | 88.8 (27.6−112.4) | 0.0080 |
| D0.2cm3 | 70.3 (17.4−140.7) | 86.8 (21.1−111.1) | 0.0053 |
| D1cm3 | 58.4 (15.0−97.1) | 79 (22.8−106.0) | 0.0007 |
| D0.1cm3 | 53.7 (8.1−92.7) | 78.9 (17.6−108.4) | 0.0003 |
| D1cm3 | 43.1 (5.3−76.4) | 69.1 (16.4−105) | 0.0004 |
| V100 (cm3) | 0 (0−0) | 0.2 (0.0−3.0) | 0.0101 |
| V50 (cm3) | 2 (0.0−16.4) | 4.2 (0.0−13.4) | 0.0105 |
*Mann-Whitney U test
Dosimetry of contralateral breast and lung for multicatheter interstitial brachytherapy (MIBT) and CyberKnife (CK) treatments. The values are given in percentages
| Contralateral breast | MIBT | CK | p-value* |
|---|---|---|---|
| D0.1cm3 | 4 (0.0−9.8) | 2.1 (0.3−11.2) | 0.0039 |
| D1cm3 | 2.6 (0.0−6.4) | 1.8 (0.3−9.9) | 0.0378 |
| D0.1cm3 | 5.5 (0.0−11.7) | 3 (1.0−5.6) | 0.0000 |
| D1cm3 | 3.6 (0.0−8.3) | 2.5 (0.6−4.8) | 0.0228 |
*Mann-Whitney U test
Treatment characteristics and mean dosimetric parameters in publications for accelerated partial breast irradiation (APBI) with CyberKnife
| Publication | Xu 201225 | Vermeulen 201451 | Goggin 201526 | Rault 201627 | Obayomi 201652 | Lozza 201853 | Bonfantini 201828 | Lee 201829 | Lee 202054 | Fröhlich 202036 | Current study |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Patient number | 14 | 21 | 9 | 10 | 10 | 20 | 10 | 10 | 103 | 25 | 32 |
| Applicator | Cones | - | Iris, MLC | Iris | - | Iris | Iris | Cones | - | MLC | MLC |
| Beam number | - | 151 | - , - | - | 155 | 180 | 122 | - | - | - | 40 (57)@ |
| Fractionation | 10 x 3.85 Gy | 10 x 3.4 Gy | 5 x 6 Gy | 10 x 4 Gy | 5 x 6 Gy | 6 5 Gy x | 5 x 6 Gy | 5 x 6 Gy | 5 x 6 Gy | 4 x 6.25 Gy | 4 x 6.25 Gy |
| Treatment Time (min) | - | 46 | 28, 14.7 | 40 | - | 60 | 60 | 60 | 33# | - | 33.3 |
| CTV-PTV margin (mm) | 2 | 2 | 2 | 3 | 5 | 5 | 5 | 2 | 0 | 2 | 2 |
| Mean PTV volume (cm3) | - | 114 | - | - | 70 | 88.1 | 121.6 | 37.7 | 73.6 | 71.6 | 71.7 |
| Ipsilateral breast | |||||||||||
| V100 (%) | 7# | 12 | 12.5, 12.2 | 14 | 28.7 | 10 | - | 9.6 | |||
| V50 (%) | 22.4# | 26 | 25.8, 24.2 | 31 | 75 | 24 | 35.5# | 23.3 | |||
| Ipsilateral non-target breast | |||||||||||
| V100 (%) | 9 | - | 1.7 | - | 0.7 | ||||||
| V50 (%) | 23.1 | 17.6 | 16.3 | - | 10.5 | 10.5 | |||||
| Ipsilateral lung | |||||||||||
| V30 (%) | 1.3 | 1.9, 1.6 | 2.3 | 2.1 | 2.2# | 1.0 | |||||
| V5 (%) | 31.4# | 39.4, 17.9 | 30.7 | ||||||||
| mean dose (%) | 4.9 | ||||||||||
| D0.1cm3 | 57.5 | 52.2 | |||||||||
| D1cm3 | 45 | 45.4 | |||||||||
| Heart | |||||||||||
| V5 (%) | 9.4 | 10 | 3.9 | 31 | 14 | 23.2 | 23 | - | 21.2 | ||
| mean dose (%) | 2.3# | 3.3 | |||||||||
| D0.1cm3 | 12.1 | 18.9 | |||||||||
| D1cm3 | 10.5 | 16.6 | |||||||||
| Skin | |||||||||||
| Dmax (%) | - | 108.8 | - | - | 106.6 | 98.2 | 94.7 | 103.3 | 88.7# | 99.6* | 88.8* |
| Contralateral breast | |||||||||||
| Dmax (%) | 1.6 | 3 | - | - | 10 | - | - | 10 | 2.7# | 3.8* | 2.1* |
* = D0.1cm3; # = median value; @ = segment number in parentheses; CTV = clinical target volume; MLC = multileaf collimator; PTV = planning target volume