Dosimetric characteristics of accelerated partial breast irradiation with CT image--based multicatheter interstitial brachytherapy: a single institution's experience.

Tibor Major1, Csaba Polgár, Katalin Lövey, Georgina Fröhlich.   

Abstract

PURPOSE: To present dosimetric assessment of CT-based high-dose-rate interstitial brachytherapy of breast cancer including dose-volume histogram analysis of dose to critical structures. METHODS AND MATERIALS: Treatment plans of 49 patients with early-stage breast cancer enrolled to the Groupe Européen de Curiethérapie-European Society for Therapeutic Radiology and Oncology trial and treated with interstitial partial breast irradiation were selected for the study. Treatment planning was performed with the use of pre- and postimplant CT imaging. Quantitative dosimetry on target volume including calculation of target volume coverage, dose homogeneity, and conformality indices was performed. Dose nonuniformity ratio was defined as ratio of volumes encompassed by 1.5 times the prescribed dose (PD) and the PD. Dose homogeneity index was defined as (V100-V150)/V100. Detailed dose-volume histogram analysis was carried out for the breast, lung and heart.
RESULTS: Three-, four-, five-, and six-plane implants were used in 17, 21, 10, and 1 patients, respectively. The median number of catheters was 15 (range, 6-28). The mean volume of the planning target volume (PTV) was 66.4 cm(3) (range, 15.5-176 cm(3)). On average, 92% (range, 90-96%) of the PTV received the PD. The mean values for dose nonuniformity ratio and dose homogeneity index were 0.32 (range, 0.25-0.41) and 0.65 (range, 0.50-0.76), respectively. The mean conformal index was 0.68 (range, 0.51-0.82) and breast tissue volume around PTV, which received at least the PD ranged between 5.3 and 51 cm(3). The mean maximum doses to skin, lung, and heart were 55%, 44%, and 21%, respectively. Volumes of the lung and heart receiving 5Gy were 50.6 and 21.0 cm(3), respectively.
CONCLUSIONS: Our implant technique based on two sets of CT scanning resulted in acceptable dose distributions with respect to target coverage, dose homogeneity, and conformality.
Copyright © 2011 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21353647     DOI: 10.1016/j.brachy.2010.12.006

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  9 in total

1.  Biological dose summation of external beam radiotherapy for the whole breast and image-guided high-dose-rate interstitial brachytherapy boost in early-stage breast cancer.

Authors:  Georgina Fröhlich; Norbert Mészáros; Viktor Smanykó; Csaba Polgár; Tibor Major
Journal:  J Contemp Brachytherapy       Date:  2020-10-30

2.  External beam boost versus interstitial high-dose-rate brachytherapy boost in the adjuvant radiotherapy following breast-conserving therapy in early-stage breast cancer: a dosimetric comparison.

Authors:  Martje Marie Terheyden; Corinna Melchert; György Kovács
Journal:  J Contemp Brachytherapy       Date:  2016-08-29

Review 3.  Treatment planning for multicatheter interstitial brachytherapy of breast cancer - from Paris system to anatomy-based inverse planning.

Authors:  Tibor Major; Csaba Polgár
Journal:  J Contemp Brachytherapy       Date:  2017-02-27

Review 4.  Dosimetric evaluation and systematic review of radiation therapy techniques for early stage node-negative breast cancer treatment.

Authors:  Tabitha Y Chan; Johann I Tang; Poh Wee Tan; Neill Roberts
Journal:  Cancer Manag Res       Date:  2018-10-23       Impact factor: 3.989

5.  Dosimetric comparison of inverse optimisation methods versus forward optimisation in HDR brachytherapy of breast, cervical and prostate cancer.

Authors:  Georgina Fröhlich; Gyula Geszti; Júlia Vízkeleti; Péter Ágoston; Csaba Polgár; Tibor Major
Journal:  Strahlenther Onkol       Date:  2019-09-03       Impact factor: 3.621

6.  Multicatheter interstitial brachytherapy versus stereotactic radiotherapy with CyberKnife for accelerated partial breast irradiation: a comparative treatment planning study with respect to dosimetry of organs at risk.

Authors:  András Herein; Gábor Stelczer; Csilla Pesznyák; Georgina Fröhlich; Viktor Smanykó; Norbert Mészáros; Csaba Polgár; Tibor Major
Journal:  Radiol Oncol       Date:  2021-03-25       Impact factor: 2.991

7.  The importance of the implant quality in APBI - Gliwice experience. Dosimetric evaluation.

Authors:  Agnieszka Cholewka; Marta Szlag; Brygida Białas; Sylwia Kellas-Ślęczka; Krzysztof Slosarek
Journal:  J Contemp Brachytherapy       Date:  2013-12-05

8.  Multi-catheter interstitial brachytherapy for partial breast irradiation: an audit of implant quality based on dosimetric evaluation comparing intra-operative versus post-operative placement.

Authors:  Lavanya Gurram; Tabassum Wadasadawala; Kishor Joshi; Reena Phurailatpam; Siji Paul; Rajiv Sarin
Journal:  J Contemp Brachytherapy       Date:  2016-04-14

9.  Does inverse planning improve plan quality in interstitial high-dose-rate breast brachytherapy?

Authors:  Tibor Major; Georgina Fröhlich; Norbert Mészáros; Viktor Smanykó; Csaba Polgár
Journal:  J Contemp Brachytherapy       Date:  2020-04-30
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.