| Literature DB >> 33751914 |
Frans Houwen1, Johannes Stemkens2, Peter J de Schipper3, Pim van der Wouw3, Martijn Heitink4, Herman van Langen5.
Abstract
Background: Measuring lymphedema with high accuracy is important for several reasons. The aim of this study was to assess the reliability and validity of three routinely used methods to estimate limb volumes. Methods andEntities:
Keywords: lymphedema; method; volume measurements
Mesh:
Substances:
Year: 2021 PMID: 33751914 PMCID: PMC8892971 DOI: 10.1089/lrb.2019.0082
Source DB: PubMed Journal: Lymphat Res Biol ISSN: 1539-6851 Impact factor: 2.589
Results of 30 Consecutive Measurements of the Upper Limb of a Volunteer Using Water Temperatures of 38°C and 16°C, Respectively
| Temperature | 38°C | 16°C |
| Mean volume (mL) | 2598 | 2591 |
| Standard deviation (mL) | 18.0 | 20.7 |
| Coefficient of variation % | 0.7 | 0.8 |
FIG. 1.(A) Average tailor tape measurements at different positions on the leg for two observers. (B) Standard deviations for the tailor tape measurement at different positions on the leg for two observers.
Calculated Volumes of a Leg Based on Tailor Tape Measurements at 17 Positions by Two Observers
| Observer 1 | Observer 2 | |
|---|---|---|
| Mean volume (mL) | 8296 | 8453 |
| Standard deviation (mL) | 45 | 44 |
| Coefficient of variation % | 0.5 | 0.5 |
Mean Volume, Standard Deviation, and Coefficient of Variation of 15 Regular and 15 Fast Measurements with the Perometer
| Regular | Fast | |
|---|---|---|
| Mean volume (mL) | 7868 | 7869 |
| Standard deviation (mL) | 31 | 30 |
| Coefficient of variation % | 0.4 | 0.4 |
Mean Volume, Standard Deviation of the Volume, and Coefficient of Variation of 50 Consecutive Measurements of a Mannequin Leg and of 100 Consecutive Lower Limb Measurements with the Perometer
| Leg of a mannequin | Leg of healthy volunteer | |
|---|---|---|
| Mean volume (mL) | 3847 | 7805 |
| Standard deviation (mL) | 14 | 49 |
| Coefficient of variation % | 0.4 | 0.6 |
Influence of Alignment of the Leg in the Perometer
| Proper alignment | Misalignment | |
|---|---|---|
| Mean volume (mL) | 8558 | 8783 |
| Standard deviation (mL) | 89 | 91 |
| Coefficient of variation % | 1 | 1 |
Mean volume, standard deviation, and coefficient of variation are presented for 50 consecutive measurements in a proper alignment and in misalignment.