Literature DB >> 3372554

Socket design and cement pressurisation in the Charnley low-friction arthroplasty.

P Shelley1, B M Wroblewski.   

Abstract

An experimental evaluation was made of the efficacy of an unflanged cup, an ogee-flanged cup and the Exeter pressuriser in the pressurisation of acetabular cement. Only a very modest injection pressure could be generated through an unflanged socket and even this was readily lost as the socket "bottomed out". The ogee-flanged socket gave a consistently high injection pressure which could be maintained throughout the process of polymerisation. The Exeter pressuriser gave marginally better results. Of the two cements tested, Palacos with gentamicin showed better intrusion than CMW Type 1 cement, but it also allowed greater extrusion once pressure was released. The importance of maintaining a continuous pressure on the cement throughout polymerisation is emphasised.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3372554

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  7 in total

1.  Cement pressurisation in the acetabulum.

Authors:  F Wadia; M H A Malik; D Leonard; M L Porter
Journal:  Int Orthop       Date:  2006-04-26       Impact factor: 3.075

2.  Factors influencing survival of the socket after primary low-friction arthroplasty of the hip.

Authors:  S Kobayashi; K Terayama
Journal:  Arch Orthop Trauma Surg       Date:  1993       Impact factor: 3.067

3.  Basic science considerations in primary total hip replacement arthroplasty.

Authors:  Saqeb B Mirza; Douglas G Dunlop; Sukhmeet S Panesar; Syed G Naqvi; Shafat Gangoo; Saif Salih
Journal:  Open Orthop J       Date:  2010-05-11

4.  The use of a "rim cutter" device and a flanged cup for improving the mantle of the acetabular component of a cemented Exeter total hip arthroplasty.

Authors:  Jenaro Ángel Fernández-Valencia; Xavier Gallart; Guillem Bori; Juan Miguel Rodríguez-Roiz; Andreu Combalia
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-08-22

5.  Comparison of flanged and unflanged acetabular cup design. An experimental study using ceramic and cadaveric acetabuli.

Authors:  Mette Ørskov; Saba Abdulghani; Ian McCarthy; Kjeld Søballe; Gunnar Flivik
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

6.  Comparison of cement pressurisation in flanged and unflanged acetabular cups.

Authors:  Rajarshi Bhattacharya; Fahad G Attar; Sarah Green; Andrew Port
Journal:  J Orthop Surg Res       Date:  2012-02-03       Impact factor: 2.359

7.  The addition of a flange does not improve the pressure generated during cemented acetabular cup implantation.

Authors:  Alexander T Boote; David J Deehan; Kenneth S Rankin; David C Swailes; Philip J Hyde
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2022-06-03       Impact factor: 3.405

  7 in total