Literature DB >> 33717758

Successful WATCHMAN Device Placement Despite Left Atrial Appendage Perforation in Two Cases.

Sangeeta Prabhakar Bhat1, Neil R Patel2, Nishant Sethi3.   

Abstract

The WATCHMAN (Boston Scientific, Marlborough, USA) is a device used to occlude the left atrial appendage (LAA) in patients of non-valvular atrial fibrillation (NVAF) with a high CHA2DS2-VASc score who are poor candidates for oral anticoagulation. LAA perforation is a well-known complication of the WATCHMAN device placement. Here we present two cases of NVAF where oral anticoagulation was not advisable due to recurrent bleeding episodes. They underwent the WATCHMAN procedure for stroke prevention. During the placement of the WATCHMAN device into the left atrial appendage (LAA) in both cases, pericardial staining was noted that worsened over the next few minutes. It was decided to deploy the 27 mm WATCHMAN device into the LAA. In one case, satisfactory hemostasis was achieved with the device deployment eliminating the need for cardiothoracic surgery. However, the second case led to pericardial tamponade and was managed by the placement of a pericardial window. To our knowledge, this is the first case series describing the use of a WATCHMAN device upon detection of LAA perforation.
Copyright © 2021, Prabhakar Bhat et al.

Entities:  

Keywords:  device deployment; hemostasis; left atrial appendage perforation; pericardial staining; pericardial tamponade; watchman

Year:  2021        PMID: 33717758      PMCID: PMC7952276          DOI: 10.7759/cureus.13251

Source DB:  PubMed          Journal:  Cureus        ISSN: 2168-8184


  9 in total

1.  Anatomy of the normal left atrial appendage: a quantitative study of age-related changes in 500 autopsy hearts: implications for echocardiographic examination.

Authors:  J P Veinot; P J Harrity; F Gentile; B K Khandheria; K R Bailey; J T Eickholt; J B Seward; A J Tajik; W D Edwards
Journal:  Circulation       Date:  1997-11-04       Impact factor: 29.690

2.  The PROTECT AF (WATCHMAN Left Atrial Appendage System for Embolic PROTECTion in Patients with Atrial Fibrillation) trial.

Authors:  Rebecca Brown Fountain; David R Holmes; Krishnaswamy Chandrasekaran; Douglas Packer; Samuel Asirvatham; Robert Van Tassel; Zoltan Turi
Journal:  Am Heart J       Date:  2006-05       Impact factor: 4.749

3.  Post-Approval U.S. Experience With Left Atrial Appendage Closure for Stroke Prevention in Atrial Fibrillation.

Authors:  Vivek Y Reddy; Douglas N Gibson; Saibal Kar; William O'Neill; Shephal K Doshi; Rodney P Horton; Maurice Buchbinder; Nicole T Gordon; David R Holmes
Journal:  J Am Coll Cardiol       Date:  2016-11-02       Impact factor: 24.094

Review 4.  Percutaneous left atrial appendage closure: Technical aspects and prevention of periprocedural complications with the watchman device.

Authors:  Sven Möbius-Winkler; Nicolas Majunke; Marcus Sandri; Norman Mangner; Axel Linke; Gregg W Stone; Ingo Dähnert; Gerhard Schuler; Peter B Sick
Journal:  World J Cardiol       Date:  2015-02-26

5.  Incidence of pericardial effusion after left atrial appendage closure: The impact of underlying heart rhythm-Data from the EWOLUTION study.

Authors:  Boris Schmidt; Timothy R Betts; Horst Sievert; Martin W Bergmann; Stephan Kische; Evgeny Pokushalov; Thomas Schmitz; Felix Meincke; Patrizio Mazzone; Kenneth M Stein; Hüseyin Ince; Lucas V A Boersma
Journal:  J Cardiovasc Electrophysiol       Date:  2018-05-21

Review 6.  The left atrial appendage: anatomy, function, and noninvasive evaluation.

Authors:  Roy Beigel; Nina C Wunderlich; Siew Yen Ho; Reza Arsanjani; Robert J Siegel
Journal:  JACC Cardiovasc Imaging       Date:  2014-12

Review 7.  Appendage obliteration to reduce stroke in cardiac surgical patients with atrial fibrillation.

Authors:  J L Blackshear; J A Odell
Journal:  Ann Thorac Surg       Date:  1996-02       Impact factor: 4.330

8.  Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial.

Authors:  David R Holmes; Vivek Y Reddy; Zoltan G Turi; Shephal K Doshi; Horst Sievert; Maurice Buchbinder; Christopher M Mullin; Peter Sick
Journal:  Lancet       Date:  2009-08-15       Impact factor: 79.321

9.  Implant success and safety of left atrial appendage closure with the WATCHMAN device: peri-procedural outcomes from the EWOLUTION registry.

Authors:  Lucas V A Boersma; Boris Schmidt; Timothy R Betts; Horst Sievert; Corrado Tamburino; Emmanuel Teiger; Evgeny Pokushalov; Stephan Kische; Thomas Schmitz; Kenneth M Stein; Martin W Bergmann
Journal:  Eur Heart J       Date:  2016-01-27       Impact factor: 29.983