Immunochromatography and chemiluminescent enzyme immunoassay for COVID-19 diagnosis.

Toshiaki Ishii1, Masakazu Sasaki1, Kageto Yamada2, Daiki Kato3, Hiroyoshi Osuka3, Kotaro Aoki4, Toshisuke Morita5, Yoshikazu Ishii4, Kazuhiro Tateda4.   

Abstract

INTRODUCTION: The rapid and accurate detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is required to prevent the spread of COVID-19. This study evaluated the utility of two SARS-CoV-2 antigen detection methods.
METHODS: We evaluated two types of antigen detection methods using immunochromatography (Espline) and quantitative chemiluminescent enzyme immunoassay (Lumipulse). RT-PCR was performed as a standard procedure for COVID-19 diagnosis. Lumipulse and RT-PCR were performed for all 486 nasopharyngeal swabs and 136 saliva samples, and the Espline test was performed for 271 nasopharyngeal swabs and 93 saliva samples.
RESULTS: The sensitivity and specificity of the Espline test were 10/11 and 260/260 (100%), respectively for the nasopharyngeal swabs and 3/9 and 84/84 (100%), respectively for the saliva samples. High sensitivities for both saliva (8/9) and nasopharyngeal swabs (22/24) were observed in the Lumipulse test. The specificities of the Lumipulse test for nasopharyngeal swabs and saliva samples were 460/462 (99.6%) and 123/127 (96.9%), respectively.
CONCLUSION: The Espline test is not effective for saliva samples but is useful for simple and rapid COVID-19 tests using nasopharyngeal swabs because it does not require special devices. The Lumipulse test is a powerful high-throughput tool for COVID-19 diagnosis because it has high detection performance for nasopharyngeal swabs and saliva samples.
Copyright © 2021 Japanese Society of Chemotherapy and The Japanese Association for Infectious Diseases. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Antigen detection; Espline; Lumipulse; RT-PCR; SARS-CoV-2

Year:  2021        PMID: 33676843      PMCID: PMC7906513          DOI: 10.1016/j.jiac.2021.02.025

Source DB:  PubMed          Journal:  J Infect Chemother        ISSN: 1341-321X            Impact factor:   2.211


Coronavirus disease 2019 (COVID-19), which is caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is one of the most severe health concerns facing the world today [1]. A COVID-19 diagnosis is determined principally by detecting the viral genome RNA using a quantitative reverse transcription polymerase chain reaction (RT-PCR) in the patients’ nasopharynx and saliva [1]. Although RT-PCR has high sensitivity and specificity, it remains difficult to perform in all clinical laboratories because it requires special devices and reagents. While antibody detection is also a useful tool for investigating past SARS-CoV-2 infection, this method may not be sufficient for the diagnosis of COVID-19 [2]. Antigen detection methods based on immunochromatography and fluorescence immunochromatography have been developed for the diagnosis of COVID-19 [3]. However, the sensitivity of immunochromatography was low (ca. 30.2–75.5%) and a fluorescence analyzer was required for fluorescence immunochromatography [3]. More recently, a rapid antigen detection kit, Espline® SARS-CoV-2 (Espline; Fujirebio Inc., Tokyo, Japan), was developed for the detection of the viral nucleocapsid antigen based on immunochromatography [4]. Additionally, a quantitative antigen detection reagent with high sensitivity and specificity, Lumipulse® SARS-CoV-2 (Lumipulse; Fujirebio Inc.) based on a chemiluminescent enzyme immunoassay of nasopharyngeal swabs and saliva samples, was developed [5]. However, the information regarding these antigen detection tests for SARS-CoV-2 is limited, especially information with respect to the use of saliva samples. Therefore, this study evaluated the utility of two SARS-CoV-2 antigen detection tests, Espline and Lumipulse, using both nasopharyngeal swabs and saliva samples; these tests were compared to the RT-PCR. Overall, 486 nasopharyngeal swabs, including 24 RT-PCR-positive samples, and 136 saliva samples, including 9 RT-PCR-positive samples, were collected from 33 COVID-19 patients and 564 non-COVID-19 patients at Toho University Omori Medical Center between August and September 2020. All the samples were used unfrozen and fresh. The Lumipulse test and RT-PCR were performed on all the test samples, while the Espline test was performed on randomly selected 271 nasopharyngeal swabs and 93 saliva samples. Since RT-PCR and the Lumipulse test were given priority in this study, the Espline test could not be performed when there was a shortage of samples. Both the Espline and Lumipulse tests were performed according to the manufacturers’ recommendations. To compare the Lumipulse and Espline tests, although the saliva samples were not generally suited to immunochromatography, we tested both saliva and nasopharyngeal samples. The saliva samples were diluted 2-fold with a dedicated reagent. The same sample solution was used in both the Espline and Lumipulse tests. The sample solution was centrifuged at 12,000 rpm for 2 min. The centrifuged supernatant was dropped into the Espline test device, and the result was judged visually after 30 min. The Espline test was judged as positive when the positive line was observed with the naked eye. The Lumipulse test was used for quantitative antigen detection on the Lumipulse G1200 system (Fujirebio) for 35 min. The Lumipulse test measurement range was from 0.6 to 5000 pg/mL. A previous report indicated that the correlation between the virus load determined using RT-PCR and antigen amount determined using the Lumipulse test was maintained up to the antigen amount of 10,000 pg/mL [6]; thus, the antigen amount >5000 pg/mL was not diluted in this study. According to the manufacturer’s recommendations, the criteria for the Lumipulse test were as follows: positive (nasopharyngeal >10 pg/mL; saliva >4.0 pg/mL), indeterminate (nasopharyngeal, 1.0–10 pg/mL; saliva, 0.67–4.0 pg/mL), and negative (nasopharyngeal <1.0 pg/mL; saliva <0.67 pg/mL). The RT-PCR was performed as previously described using the TaqMan Fast Virus 1-step Master Mix (Thermo Fisher Scientific, Walthan, MA, USA) on the QuantStudio 5 Real-Time PCR System (Thermo Fisher Scientific) [7]. The viral RNA was extracted from the Lumipulse test suspension using the QIAamp Viral RNA Mini Kit (Qiagen, Hilden, Germany). A threshold cycle (Ct) value less than 35 was positive, and a Ct value of 35 or more was defined as below the limit of detection (LOD) [7]. Table 1 shows the results of each detection method. The Espline test for nasopharyngeal swabs was highly sensitive (10/11 were positive) and had high specificity (260/260, 100%), in addition to its high concordance with the RT-PCR results. In one false-negative result, the Ct value of the RT-PCR was 33.7 because of a low virus load. Conversely, the Espline test for saliva samples showed a significantly low sensitivity (3/9 were positive), suggesting that the Espline test was not suitable for saliva samples. According to previous reports on immunochromatography using nasopharyngeal swabs, the sensitivities were 30.2%–57.6% for the Respi-strip (CORIS BioConcept, Gembloux, Belgium) and 73.3%–75.5% for the Panbio COVID-19 Rapid Test Device (Abbot, Chicago, USA) [3,[8], [9], [10]]. These previous reports on the diagnosis of COVID-19 using immunochromatography showed lower sensitivities than of our study. The high sensitivity and specificity of the Espline test may be related to the use of centrifugation to process the sample. However, the main purpose of centrifugation is to remove impurities and increase specificity, which may not significantly affect sensitivity. In addition, previous reports have indicated that high sensitivities were obtained during the first week or within nine days of symptom onset and with high viral loads [4,8]. Therefore, the sensitivity of immunochromatography might depend on the timing of sampling. Although we studied only small numbers of positive samples, the Espline test may be useful for the diagnosis of COVID-19 using nasopharyngeal swabs. In addition, since the Espline test does not need special devices, it has a shorter turnaround time and is easy to judge.
Table 1

Results of each antigen detection methods compared with the RT-PCR: a. clinical evaluation of the Espline test; b. clinical evaluation of the Lumipulse test.

(a) Espline test
Sample typeRT-PCR
Predictive valuesSample typeRT-PCR
Predictive values
Nasopharyngeal swabsPositiveNegativeSalivaPositiveNegative
Espline
Positive100100%Positive30100%
Negative126099.6%Negative68493.3%
Total11260Total984

Sensitivities/specificities90.9%100%33.3%100%




(b) Lumipulse test

Sample typeRT-PCR
Predictive values
Sample typeRT-PCR
Predictive values
Nasopharyngeal swabs
Positive
Negative
Saliva
Positive
Negative
Lumipulse
Positive221a95.7%Positive8188.9%
Indeterminate01Indeterminate13
Negative246099.6%Negative0123100%
Total24462Total9127

Sensitivities/specificities91.7%99.6%88.9%96.9%

Previous RT-PCR in another sample for the same patient was positive.

Results of each antigen detection methods compared with the RT-PCR: a. clinical evaluation of the Espline test; b. clinical evaluation of the Lumipulse test. Previous RT-PCR in another sample for the same patient was positive. The sensitivity and specificity of the Lumipulse test for nasopharyngeal swabs were 22/24 (2 negative) and 460/462 (99.6%, 1 positive and 1 indeterminate), respectively. Although a sample that gave the false-positive result with the Lumipulse test had a 27.3 pg/mL antigen value and was below the LOD for the RT-PCR, SARS-CoV-2 RNA was detected using a RT-PCR when the same patient was tested again four days earlier. This patient was diagnosed with COVID-19 four days earlier with an antigen level of 20.9 and a Ct value of 33.7. Thus, this result was considered a false-negative for the RT-PCR, which may have been due to extraction of the RNA from the antigen test suspension. One indeterminate result showed a very low antigen level (1.0 pg/mL). Two false-negative results revealed high Ct values in RT-PCR (33.4 and 34.1); thus, it was considered that these samples contained only a low virus load. By contrast, the Lumipulse test for saliva samples showed a higher sensitivity than the Espline test (8/9, 1 indeterminate). The specificity of the Lumipulse test for saliva samples was 123/127 (96.9%, 1 positive and 3 indeterminate). One false-positive result had a 4.54 pg/mL antigen value, and the RT-PCR showed a result below the LOD. Among the four indeterminate results, one result showed that the Lumipulse test and Ct value of RT-PCR were 1.04 pg/mL and 32.6, respectively. In three of the four indeterminate results, all three RT-PCR results were negative, and the antigen quantification values were as follows: 0.71, 0.86, and 3.39 pg/mL. Our results indicated that the Lumipulse test had high sensitivity and specificity for both nasopharyngeal swabs and saliva samples. Particularly, one nasopharyngeal swab from a SARS-CoV-2 patient was negative in the RT-PCR and positive in the Lumipulse test. Thus, our results suggested that the Lumipulse test might have a high detection performance similar to the RT-PCR. This is the first report to evaluate the usefulness of the Lumipulse test in diagnosing COVID-19 using saliva. Previous studies of the Lumipulse test using nasopharyngeal swabs indicated that the sensitivities and specificities were 55.2%–91.7% and 97.3%–99.6%, respectively [5,11], and that these differences might be dependent on the timing of the sampling mentioned above. The detailed results of the COVID-19 positive patients are shown in Table 2 . Almost all the samples used in this study were obtained under 7 days after onset (Table 2). No association was found between the presence or absence of underlying diseases and testing results. The distribution of the SARS-CoV-2 antigen value in COVID-19 patients were 4.5 to >5000, and the geometric mean was 808.9 pg/mL. If positive results with low antigen value or indeterminate results are observed, it may be necessary to perform an additional test with using RT-PCR and to assess the patients’ symptoms. There have been reports of a case of confusion to a false-positive result from the Lumipulse test [11].
Table 2

Detailed testing results of the coronavirus disease 2019 patients. a. nasopharyngeal swab samples; b. saliva samples. Symptomatic means dysosmia, dysgeusia, fever, and pneumonia. Asymptomatic patients were close contact to coronavirus disease 2019 patients.

(a)
Sample type, nasopharyngeal swab
Elapsed daysaRT-PCR
Lumipulse
EsplineSymptomatic or asymptomatic
Serial no.AgeGenderUnderlying diseaseCt valueJudgementAntigen value (pg/mL)Judgement
150sMaleDiabetes017.2+>5000++Symptomatic
240sMaleUnknown422.8+>5000+N/ASymptomatic
320sFemaleTonsillitis224.3+>5000++Symptomatic
450sMaleAsthma, hypertension114.6+>5000+N/ASymptomatic
520sMaleUnknown019.1+>5000+N/ASymptomatic
620sFemaleUnknown018.3+>5000++Symptomatic
750sMaleUpper extremity deep vein thrombosis020.3+>5000++Symptomatic
840sMaleHypertension, hyperuricemia, hypercholesterolemia215.6+>5000++Symptomatic
990sMaleAngina, hypertension, aortic dissection021.0+>5000++Symptomatic
1050sMaleDiverticulitis014.5+>5000+N/ASymptomatic
1120sMaleUnknownN/Ab21.9+>5000+N/AAsymptomatic
1230sMaleUnknownN/A29.6+>5000+N/AAsymptomatic
1320sFemaleUnknownN/A23.8+>5000+N/AAsymptomatic
1430sMaleUnknown120.3+>5000++Symptomatic
1520sFemaleUnknown016.0+>5000+N/ASymptomatic
1630sMaleUnknown626.5+3238.70+N/ASymptomatic
1780sMaleAcute myocardial infarction127.8+1858.10++Symptomatic
1820sFemaleUnknownN/A29.4+1144.50+N/AAsymptomatic
1970sFemaleHypertension, old cerebral infarction034.7+526.60++Symptomatic
2060sFemaleAsthma431.1+362.40++Symptomatic
2120sFemaleAsthmaN/A33.0+109.70+N/AAsymptomatic
2210sMaleDiabetes033.7+20.90+Symptomatic
23<10MaleKawasaki Disease033.4+<0.60N/ASymptomatic
2430sMaleUnknownN/A34.1+<0.60N/AAsymptomatic
25
10s
Male
Diabetes
5
>35.0

27.30
+

Symptomatic



(b)

Sample type, saliva
Elapsed days
RT-PCR
Lumipulse
Espline
Symptomatic or asymptomatic

Serial no.
Age
Gender
Underlying disease
Ct value
Judgement
Antigen value (pg/mL)
Judgement
120sMaleUnknown118.3+>5000++Symptomatic
270sMalePancreatic cancer017.3+4791.02++Symptomatic
350sFemaleLichen planus918.6+2090.03++Symptomatic
420sMaleParoxysmal supraventricular tachycardia028.1+45.02+Symptomatic
520sMaleUnknown029.1+17.96+Symptomatic
620sFemaleUnknown631.4+17.34+Symptomatic
770sMaleHypertension, diabetes, chronic kidney disease, lung cancer1430.6+16.64+Symptomatic
830sFemaleUnknown028.8+8.48+Symptomatic
920sFemaleUnknown232.6+1.04IndeterminateSymptomatic

Onset-day was defined as day 0.

N/A, not available.

Detailed testing results of the coronavirus disease 2019 patients. a. nasopharyngeal swab samples; b. saliva samples. Symptomatic means dysosmia, dysgeusia, fever, and pneumonia. Asymptomatic patients were close contact to coronavirus disease 2019 patients. Onset-day was defined as day 0. N/A, not available. This study had a limitation. A limited number of positive SARS-CoV-2 samples (both nasopharyngeal swabs and saliva samples) were available for this study. Therefore, further evaluations of the Espline and Lumipulse tests might be needed. In conclusion, although the Espline test could not be used for saliva samples, it is useful for a simple point-of-care testing because it does not require a special device. In contrast, although the Lumipulse test requires a special device and reagents, it is useful in routine diagnosis because the Lumipulse test has a performance similar to the RT-PCR and has a simpler procedure and higher throughput than RT-PCR.

Authorship statement

All authors meet the ICMJE authorship criteria. All authors have made substantial contributions to the conception and design of the study, data acquisition, data analysis and interpretation, drafting or revision of the article for important intellectual content, and final approval of the version to be submitted.

Author contributions

TI helped in writing the original draft; he was also involved in the methodology and formal analysis of this paper. MS, KY, HO, DK, KA, TM, YI, and KT contributed to reviewing and editing and were also involved in the formal analysis and writing of the original draft.

Ethical approval

This study protocol was approved by the Ethics Committee of the Faculty of Medicine, Toho University (No. A20028_A20020_A20014_A19099).

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Declaration of competing interest

None.
  7 in total

1.  Development of Genetic Diagnostic Methods for Detection for Novel Coronavirus 2019(nCoV-2019) in Japan.

Authors:  Kazuya Shirato; Naganori Nao; Harutaka Katano; Ikuyo Takayama; Shinji Saito; Fumihiro Kato; Hiroshi Katoh; Masafumi Sakata; Yuichiro Nakatsu; Yoshio Mori; Tsutomu Kageyama; Shutoku Matsuyama; Makoto Takeda
Journal:  Jpn J Infect Dis       Date:  2020-02-18       Impact factor: 1.362

2.  SARS-CoV-2-IgG response is different in COVID-19 outpatients and asymptomatic contact persons.

Authors:  Nele Wellinghausen; Dietmar Plonné; Meike Voss; Ralitsa Ivanova; Reinhard Frodl; Susanne Deininger
Journal:  J Clin Virol       Date:  2020-07-06       Impact factor: 3.168

3.  Panbio antigen rapid test is reliable to diagnose SARS-CoV-2 infection in the first 7 days after the onset of symptoms.

Authors:  Manuel Linares; Ramón Pérez-Tanoira; Ana Carrero; Juan Romanyk; Felipe Pérez-García; Peña Gómez-Herruz; Teresa Arroyo; Juan Cuadros
Journal:  J Clin Virol       Date:  2020-10-16       Impact factor: 3.168

4.  Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis.

Authors:  Anaïs Scohy; Ahalieyah Anantharajah; Monique Bodéus; Benoît Kabamba-Mukadi; Alexia Verroken; Hector Rodriguez-Villalobos
Journal:  J Clin Virol       Date:  2020-05-21       Impact factor: 3.168

5.  Comparison of automated SARS-CoV-2 antigen test for COVID-19 infection with quantitative RT-PCR using 313 nasopharyngeal swabs, including from seven serially followed patients.

Authors:  Yosuke Hirotsu; Makoto Maejima; Masahiro Shibusawa; Yuki Nagakubo; Kazuhiro Hosaka; Kenji Amemiya; Hitomi Sueki; Miyoko Hayakawa; Hitoshi Mochizuki; Toshiharu Tsutsui; Yumiko Kakizaki; Yoshihiro Miyashita; Shintaro Yagi; Satoshi Kojima; Masao Omata
Journal:  Int J Infect Dis       Date:  2020-08-12       Impact factor: 3.623

6.  Another false-positive problem for a SARS-CoV-2 antigen test in Japan.

Authors:  Taku Ogawa; Tatsuya Fukumori; Yuji Nishihara; Takahiro Sekine; Nao Okuda; Tomoko Nishimura; Hiroyuki Fujikura; Nobuyasu Hirai; Natsuko Imakita; Kei Kasahara
Journal:  J Clin Virol       Date:  2020-08-25       Impact factor: 3.168

Review 7.  World Health Organization declares global emergency: A review of the 2019 novel coronavirus (COVID-19).

Authors:  Catrin Sohrabi; Zaid Alsafi; Niamh O'Neill; Mehdi Khan; Ahmed Kerwan; Ahmed Al-Jabir; Christos Iosifidis; Riaz Agha
Journal:  Int J Surg       Date:  2020-02-26       Impact factor: 6.071

  7 in total
  10 in total

Review 1.  Rapid, point-of-care antigen tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Pawana Sharma; Sarah Berhane; Susanna S van Wyk; Nicholas Nyaaba; Julie Domen; Melissa Taylor; Jane Cunningham; Clare Davenport; Sabine Dittrich; Devy Emperador; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Jan Y Verbakel; Yemisi Takwoingi; Sian Taylor-Phillips; Ann Van den Bruel; Jonathan J Deeks
Journal:  Cochrane Database Syst Rev       Date:  2022-07-22

Review 2.  Performance of Antigen Detection Tests for SARS-CoV-2: A Systematic Review and Meta-Analysis.

Authors:  Anastasia Tapari; Georgia G Braliou; Maria Papaefthimiou; Helen Mavriki; Panagiota I Kontou; Georgios K Nikolopoulos; Pantelis G Bagos
Journal:  Diagnostics (Basel)       Date:  2022-06-04

Review 3.  Novel impacts of saliva with regard to oral health.

Authors:  Hitoshi Uchida; Catherine E Ovitt
Journal:  J Prosthet Dent       Date:  2021-06-16       Impact factor: 3.426

Review 4.  Saliva as a diagnostic specimen for SARS-CoV-2 detection: A scoping review.

Authors:  Yifei Wang; Akshaya Upadhyay; Sangeeth Pillai; Parisa Khayambashi; Simon D Tran
Journal:  Oral Dis       Date:  2022-04-21       Impact factor: 4.068

5.  Effect of varying storage conditions on diagnostic test outcomes of SARS-CoV-2.

Authors:  Satoshi Oguri; Shinichi Fujisawa; Keisuke Kamada; Sho Nakakubo; Yu Yamashita; Junichi Nakamura; Hiroshi Horii; Kazuki Sato; Mutsumi Nishida; Takanori Teshima; Yoichi Ohiro; Ayato Takada; Satoshi Konno
Journal:  J Infect       Date:  2021-04-03       Impact factor: 6.072

6.  Comparison of a novel antigen detection test with reverse transcription polymerase chain reaction assay for laboratory diagnosis of SARS-CoV-2 infection.

Authors:  Osman Sezer Cirit; Esvet Mutlu; Banu Sancak; Tanil Kocagöz; Özge Can; Candan Çicek; Ayca Arzu Sayiner; Özgür Appak; Neval Yurttutan Uyar; Canan Külah; Aysegül Çopur Çiçek; Osman Birol Özgümüs; Yasemin Ay Altintop; Esma Saatçi; Tekin Karsligil; Yasemin Zer; Nevgün Sepin Özen; Yesim Çekin; Zeynep Ceren Karahan; Ebru Evren; Ayse Esra Karakoç; Sultan Gülbahçe Orhan; Derya Mutlu; Tugba Bozdemir; Yeliz Tanriverdi Çayci; Canberk Çinar; Meltem Tasbakan; Merve Mert; Ece Çinar; Oya Özlem Eren Kutsoylu; Sesin Kocagöz; Ayse Ertürk; Ilhami Çelik; Ayse Özlem Mete; Müge Günalp Eneyli; Irem Akdemir; Taliha Karakök; Dilara Inan; Aynur Atilla; Şevket Onur Taflan; Kağan Etka Yörük
Journal:  Infection       Date:  2022-05-05       Impact factor: 7.455

7.  Diagnostic Accuracy of Rapid Antigen Tests for COVID-19 Detection: A Systematic Review With Meta-analysis.

Authors:  Maniya Arshadi; Fatemeh Fardsanei; Behnaz Deihim; Zahra Farshadzadeh; Farhad Nikkhahi; Farima Khalili; Giovanni Sotgiu; Amir Hashem Shahidi Bonjar; Rosella Centis; Giovanni Battista Migliori; Mohammad Javad Nasiri; Mehdi Mirsaeidi
Journal:  Front Med (Lausanne)       Date:  2022-04-07

8.  The Detection of SARS-CoV2 Antigen in Wastewater Using an Automated Chemiluminescence Enzyme Immunoassay.

Authors:  Supranee Thongpradit; Somsak Prasongtanakij; Supanart Srisala; Suwannee Chanprasertyothin; Ekawat Pasomsub; Boonsong Ongphiphadhanakul
Journal:  Int J Environ Res Public Health       Date:  2022-06-24       Impact factor: 4.614

9.  Investigation of the diagnostic performance of the SARS-CoV-2 saliva antigen test: A meta-analysis.

Authors:  Cheng-Chieh Chen; Ke-Yu Hsiao; Chyi-Huey Bai; Yuan-Hung Wang
Journal:  J Microbiol Immunol Infect       Date:  2022-07-16       Impact factor: 10.273

10.  Accuracy of novel antigen rapid diagnostics for SARS-CoV-2: A living systematic review and meta-analysis.

Authors:  Lukas E Brümmer; Stephan Katzenschlager; Mary Gaeddert; Christian Erdmann; Stephani Schmitz; Marc Bota; Maurizio Grilli; Jan Larmann; Markus A Weigand; Nira R Pollock; Aurélien Macé; Sergio Carmona; Stefano Ongarello; Jilian A Sacks; Claudia M Denkinger
Journal:  PLoS Med       Date:  2021-08-12       Impact factor: 11.069

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.