Laboratory assessment of three reflectance meters designed for self monitoring of blood glucose concentrations.

J P Begley1, A R Forrest.   

Abstract

Three monitors designed for self monitoring of blood glucose concentrations by diabetic patients were evaluated in a laboratory. All three assessments correlated positively with a laboratory reference method based on glucose oxidase. Coefficients of variation at all levels tested were less than 10% for each monitor. Stability of the colour development of each stick was assessed and the effects of changes in blood spot volume and incubation time were examined. Twelve ward nurses were each asked to measure two blood samples and the results obtained showed wide variability. Overall, our findings suggest that each of the three monitors tested is suitable for use in monitoring capillary blood glucose concentrations by those who are properly trained.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3366924      PMCID: PMC1141464          DOI: 10.1136/jcp.41.4.397

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  13 in total

1.  Editorial: "Control" and diabetes.

Authors:  G F Cahill; L D Etzwiler; N Freinkel
Journal:  N Engl J Med       Date:  1976-04-29       Impact factor: 91.245

2.  Incorrect least-squares regression coefficients in method-comparison analysis.

Authors:  P J Cornbleet; N Gochman
Journal:  Clin Chem       Date:  1979-03       Impact factor: 8.327

Review 3.  Patient, specimen and analysis as potential sources of error.

Authors:  C G Fraser; L R Watkinson
Journal:  Contemp Issues Clin Biochem       Date:  1985

Review 4.  Analytical goals for glucose analyses.

Authors:  C G Fraser
Journal:  Ann Clin Biochem       Date:  1986-07       Impact factor: 2.057

5.  Home blood glucose monitoring.

Authors:  R B Tattersall
Journal:  Diabetologia       Date:  1979-02       Impact factor: 10.122

Review 6.  Methods of self monitoring of diabetic control.

Authors:  I Peacock; R Tattersall
Journal:  Clin Endocrinol Metab       Date:  1982-07

7.  Quality assessment of blood glucose monitors in use outside the hospital laboratory.

Authors:  R F Drucker; D R Williams; C P Price
Journal:  J Clin Pathol       Date:  1983-08       Impact factor: 3.411

8.  Stabilization of blood glucose by cooling with ice: an effective procedure for preservation of samples from adults and newborns.

Authors:  Y L Lin; C H Smith; D N Dietzler
Journal:  Clin Chem       Date:  1976-12       Impact factor: 8.327

Review 9.  Home monitoring of diabetes mellitus--a quiet revolution.

Authors:  A L McCall; C J Mullin
Journal:  Clin Lab Med       Date:  1986-06       Impact factor: 1.935

10.  Correlation between plasma and urine glucose in diabetes.

Authors:  L R Morris; J A McGee; A E Kitabchi
Journal:  Ann Intern Med       Date:  1981-04       Impact factor: 25.391

View more
  2 in total

1.  Experience of a quality assessment scheme for non-laboratory glucose meters.

Authors:  T R Trinick; E Duly
Journal:  J Clin Pathol       Date:  1992-01       Impact factor: 3.411

2.  Assessment of the Glycotronic C blood glucose meter.

Authors:  A R Forrest; E J Pitt
Journal:  J Clin Pathol       Date:  1991-09       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.