The role of trans-thoracic echocardiography in the assessment of aortic annular diameter.

Khaled I Naguib1, Mohamed A Attia1, Mohamed S Bashandy1, Muhammad Saad Reihan1, Tarek Ahmed Dabash1, Ahmed B Abd El-Salam2, Hosam H Helal1, Eshak I Bahbah3,4.   

Abstract

ABSTRACT: We aimed to compare two-dimension transthoracic echocardiogram (2D-TTE) and three-dimension transthoracic echocardiogram (3D-TTE) measurements of the aortic annular diameter using multi-detector CT (MDCT) as a gold standard.This prospective observational study included 50 consecutive patients who came to the cardiology department, Al-Azhar University Hospital, New Damietta, for MDCT coronary angiography. The study was carried out in the period from July 2016 until February 2017. All patients were subjected to informed consent, clinical history, physical examination, transthoracic echocardiography 2D and 3D, and MDCT.The aortic annular areas measured by MDCT and 3D-TTE were significantly larger than areas by 2D-TTE. A good correlation (r = 0.82) was observed between the areas obtained by 3D-TTE and MDCT; however, the correlation between the values by 2D-TTE and MDCT was rough (r = 0.30). Eccentricity Index (EI) values in 28% of the patients were greater than 0.1, that is, the aortic annulus was elliptical.Accuracy of aortic annular diameter measurement by 3D-TTE was superior to that by 2D-TTE. Three-D TTE and MDCT revealed that the shape of the aortic annulus was elliptical in 28% to 30% respectively of study subjects. There is a strong concordance between the minimum and the maximum diameter determine by 3D-TTE and MDCT.
Copyright © 2021 the Author(s). Published by Wolters Kluwer Health, Inc.

Entities:  

Mesh:

Year:  2021        PMID: 33655932      PMCID: PMC7939191          DOI: 10.1097/MD.0000000000024682

Source DB:  PubMed          Journal:  Medicine (Baltimore)        ISSN: 0025-7974            Impact factor:   1.817


Introduction

Two-dimension transthoracic echocardiogram (2D-TTE) is a widespread imaging technique used to screen an aortic annulus.[ Also, it is the most common non-invasive diagnostic modality, expert opinions are needed for data interpretation because of the intra-cardiac complex anatomy and its relationship with the surrounding tissues.[ However, this technique has some limitations as the variation of results and interpretations when used by different investigators based on their imaginary reconstructions of cardiac structures. Besides, it can be affected by the patient's position and condition and, Doppler angle errors.[ These limitations can be overcome using real-time three-dimension transthoracic echocardiogram (3D-TTE), which provides a detailed evaluation of cardiac physiology and anatomy in real-time.[ Three-D TTE is an emerging non-invasive technique based on real-time volumetric scanning and useful in evaluating cardiac chamber volumes and the left ventricular wall's motion.[ It does not need extensive reconstruction after dataset acquisition as with computed tomography (CT) and magnetic resonance imaging. There is an increase in the exact measurement of the aortic annular diameter when measured by 3D echocardiography in relation to 2D measures. Moreover, 3D echocardiography gives more detailed anatomy. Since 2002 and till now, further advances have been made in 3D computer software technology as a matrix-array system, leading to better screening and evaluation of valvular and Congenital Heart Disease (CHD) in pediatric and adult populations.[ Several studies have shown that the aortic annular diameter is underestimated with 2D echocardiography compared with the 3D when evaluated in patients with severe aortic stenosis.[ Multi-detector computed tomography (MDCT) is one of the most exciting technological revolutions in cardiac imaging, used as a non-invasive assessment of coronary artery diseases, with new clinical applications arising over the past years.[ Nevertheless, there is no evidence on the role of MDCT in valvular diseases.[ Therefore, echocardiography has retained its clinical practice status, despite its limitations in assessing valvular morphology. This study aimed to compare 2D-TTE and 3D-TTE measurements of the aortic annulus diameter using multi-detector CT as a gold standard.

Patients and methods

In this study, we followed the rules and guidelines of the Strengthening the Reporting of Observational Studies in Epidemiology statement of observational studies.

Study design

This prospective observational study included 50 consecutive patients who came to the cardiology department, Al-Azhar University Hospital, New Damietta, for MDCT coronary angiography, which was used to exclude the coronary artery diseases, coronary artery anomalies, fistulas, and aneurysms. The study was carried over a period from July 2016 until February 2017. Informed consent was obtained from all patients, and ethical approval was obtained from the council of Al-Azhar University Hospital.

Selection criteria

Inclusion criteria

We included all patients with normal cardiac anatomy and function assessed by physical examination and standard 2D echocardiography.

Exclusion criteria

Patients with structural heart disease diagnosed by 2D-TTE, which is defined as an abnormality of the heart that affects the blood flow. Many structural heart conditions are congenital, but these abnormalities can also form later in life after infection or result from other underlying conditions. These secondary causes, even in the mild form, can affect the actual measure of aortic annular diameter.

Techniques method

TTE

Patients underwent 2D-TTE and 3D-TTE at Echo-Lab in the cardiology department, Al-Azhar University Hospital, New Damietta, using a Philips IE-33 echocardiography machine. The probe used was the X5-1 phased array sector probe (frequency range 1.5–4.3 MHz). We used a hard disk for storage of imaging data before transformation into a personal computer for analysis. We recorded all imaging data of the long axis of a left ventricle (LV) and 3D zooming image of the aortic annulus in parasternal views in the full-volume for 4 consecutive beats.

Measurement of aortic annulus diameters by 2D and 3D echocardiography

The annulus diameters (r) were measured by 2D-TTE using the zoomed images at the insertion of the leaflets in the mid-systolic phase from the parasternal long-axis view. We used this equation [area = π (r/2)2] to calculate the annular area by presuming the annulus’ shape to be circular. We used the cutting image of 3D-TTE to evaluate the maximum and minimum diameters of the aortic annulus, where the shape of the aortic annulus was mostly circular, obtained by a piece of analysis software (Q-Lab 10). Besides, the aortic annulus was traced, and the area was measured in the same image (Fig. 1).
Figure 1

3D-TTE multi-planar reformations of the aortic root.

3D-TTE multi-planar reformations of the aortic root.

Multi-detector CT (MDCT)

Electrocardiogram (ECG)-gated MDCT images were acquired using a 160-slice Dual-Source CT scanner (Toshiba) machine. Contrast enhancement was achieved with ULTRAVIST (iopromide) injected into the antecubital vein (20–50 ml) followed by a 30-ml saline bolus. We obtained 3 cross-sectional views (axial view, sagittal view, and coronal view) after image reconstruction at the timing of 200 ms from the R wave of the ECG. Then, the maximum and minimum diameters and the aortic annular areas were measured using the cutting images involving the aortic annulus obtained along the short axis (Fig. 2).
Figure 2

(A) Analysis of the aortic annulus by MDCT sagittal view; (B) Analysis of the aortic annulus by MDCT coronal view; (C) Analysis of the aortic annulus by MDCT axial view. (D) 2-DTTE zoomed images at the insertion of the leaflets from the parasternal long-axis view.

(A) Analysis of the aortic annulus by MDCT sagittal view; (B) Analysis of the aortic annulus by MDCT coronal view; (C) Analysis of the aortic annulus by MDCT axial view. (D) 2-DTTE zoomed images at the insertion of the leaflets from the parasternal long-axis view.

Statistical analysis

We used the “Data Collection Form” to collect and tabulate the collecting data before its transformation into a computerized database. Continuous variables were presented as mean ± SD, while qualitative variables were expressed as numbers and percentages. Statistical analysis was performed using the Statistical Package for the Social Sciences (SPSS Software, version 22). In 50 randomly selected studies, aortic annulus diameters were remeasured by the same observers at 3 separate times to determine intraobserver concordance and by another 2 observers to determine interobserver agreement. Intra- and inter-observer concordance were evaluated for 2D and 3D echocardiography and CT. Correlations were done to test for linear relations between variables. Measurements determined by MDCT were compared with those obtained by 3D-TTE and 2D-TTE using Bland-Altman analysis. P value was considered significant if <.05.

Results

Demographic data

This study included 50 patients with a mean age of 37.5 ± 10.65 years. Out of them, 34 (68%) were males. Regarding the clinical presentation, 13 patients (26%) were hypertensive, and 10 (20%) were diabetics, systolic BP of the study population ranged between 90 and 160 mm Hg, their diastolic BP ranged between 60 and 100 mm Hg. Table 1 presents all demographic and clinical data for the enrolled population.
Table 1

Characteristics of the included patients in the study.

CharacteristicValue
Age, years37.50 ± 10.65
Gender
 Male34 (68%)
 Female16 (32%)
DM
 DM10 (20%)
 Not DM40 (80%)
HTN
 HTN13 (26%)
 Not HTN37 (74%)
SBP123.30 ± 15.99
DBP78.60 ± 9.95
BAS1.59 ± 0.22
Characteristics of the included patients in the study.

Comparison between MDCT and 2D findings

There were statistically significant differences between mean 2D systolic aortic annular diameter and mean minimum (P = .002), and maximum aortic annular diameter derived from MDCT (P < .001). Also, there was a statistically significant difference between the mean aortic annular area derived from 2D and MDCT (P < .001). Table 2 shows the finding of 2D-TTE.
Table 2

Measurements of aortic annular diameter using different echocardiographic modalities.

Aortic annulus at minimum diameterAortic annulus at maximum diameterAnnular areaEccentricity index
CT2.52 ± 0.30a2.71 ± 0.37a5.08 ± 0.44a0.13 ± 0.22a
2D-TTE2.36 ± 0.15b-4.32 ± 0.56b-
3D2.46 ± 0.26a2.64 ± 0.28a4.79 ± 0.47c0.07 ± 0.06a
Measurements of aortic annular diameter using different echocardiographic modalities. There was a weak positive correlation between 2D-TTE systolic aortic annular diameter and minimum (P = .002, r = 0.42) and maximum aortic annular diameter derived from MDCT (P = .006, r = 0.38). Also, the mean 2D-TTE aortic annular area presented a weak positive correlation with the mean aortic annular area derived from MDCT (P = .031, r = 0.31).

Comparison between MDCT and 3D findings

There was no statistically significant difference between mean 3D-TTE minimum aortic annular diameter and mean minimum aortic annular diameter derived from MDCT (P = .37). In terms of mean maximum aortic annular diameter, there was no statistically significant difference between mean 3D-TTE and MDCT (P = .31). Also, there was no statistically significant difference between the mean 3D-TTE eccentricity index (EI) and the mean EI derived from MDCT (P = .16). On the other hand, there was a statistically significant difference between the mean 3D-TTE derived aortic annular area and the mean aortic annular area derived from MDCT (P = .001). Table 2 shows the main finding of 3D-TTE and MDCT. There was an excellent positive correlation between 3D-TTE minimum aortic annular diameter and minimum aortic annular diameter derived from MDCT (P < .001, r = 0.81). Also, between 3D-TTE maximum aortic annular diameter and maximum aortic annular diameter derived from MDCT (P < 0.001, r = 0.8). Nevertheless, there was a weak positive correlation between 3D-TTE EI and EI derived from MDCT (P < .02, r = 0.43). Regarding the mean aortic annular area, there was an excellent positive correlation between 3D-TTE and MDCT (P < .001, r = 0.82). The concordance between minimum aortic annular diameter determined by 3D-TTE and MDCT (average bias 0.06 cm, 95% CI −0.32 to 0.44 cm) was high and showed a small difference (Fig. 3A). Also, there is a small difference between the maximum aortic annular diameter determined by 3D-TTE and MDCT (average bias 0.075 cm, 95% CI −0.36 to 0.516 cm) (Fig. 3B). The concordance between the annular area determined by 3D-TTE and MDCT (average bias 0.29 cm, 95% CI −0.24 to 0.83 cm) was better than the concordance between the annular area determined by 2D-TTE and MDCT (average bias 0.76 cm, 95% CI −0.39 to 1.92 cm) (Fig. 3 C and D).
Figure 3

(A) Bland-Altman plots demonstrating the concordance of aortic annulus diameters determined by 3D-TTE to minimal diameters and minimal diameter by MDCT; (B) Bland-Altman plots demonstrating the concordance of aortic annulus diameters determined by 3D-TTE to maximal diameters and maximal diameter by MDCT; (C) Bland-Altman plots demonstrating the concordance of the annular area determined by 3D-TTE to the aortic area by MDCT; (D) Bland-Altman plots demonstrating the concordance of the annular area determined by 2D-TTE to the aortic area by MDCT.

(A) Bland-Altman plots demonstrating the concordance of aortic annulus diameters determined by 3D-TTE to minimal diameters and minimal diameter by MDCT; (B) Bland-Altman plots demonstrating the concordance of aortic annulus diameters determined by 3D-TTE to maximal diameters and maximal diameter by MDCT; (C) Bland-Altman plots demonstrating the concordance of the annular area determined by 3D-TTE to the aortic area by MDCT; (D) Bland-Altman plots demonstrating the concordance of the annular area determined by 2D-TTE to the aortic area by MDCT.

Comparison between 2D-TTE and 3D-TTE findings

There was a statistically significant difference when compared 2D-TTE systolic aortic annular diameter against mean minimum (P = .017) and mean maximum aortic annular diameter derived from 3D-TTE (P < .001). Also, there was a statistically significant difference between the mean aortic annular area derived from 2D-TTE and 3D-TTE (P < .001). There was a weak positive correlation between 2D-TTE systolic aortic annular diameter and minimum (P = .003, r = 0.41) and maximum aortic annular diameter derived from 3D-TTE (P = .03, r = 0.36). On the other hand, there was no correlation between 2D-TTE aortic annular area and aortic annular area derived from 3D-TTE (P = .09, r = .24).

Correlation between demographic variables and aortic annular area

There are no correlations between [age, gender, diabetes mellitus (DM), or hypertension] and (annular area by MDCT, 3D-TTE, or 2D-TTE). Table 3 shows the relation between these variables.
Table 3

correlation between aortic annular diameters measured by different techniques and different parameters.

CT annular area3D annular area2D annular area
ParametersrP valuerP valuerP value
Age0.17.230.14.300.13.33
Sex0.03.820.15.270.08.57
DM0.04.770.07.600.14.33
HTN0.12.390.09.520.17.23
SBP0.13.340.19.180.09.53
DBP0.04.770.05.700.14.33
BSA0.11.440.03.790.29.04
correlation between aortic annular diameters measured by different techniques and different parameters.

Observer agreement

Intraobserver correlation of measurements of the aortic annulus diameter performed on 2D-systolic, 2D-annular, 3D-minimum vertical, 3D-maximum diameter, 3D-annular area, CT minimum diameter, and CT maximum diameter was high. In addition, the interobserver correlation of measurements performed the same techniques was high (Table 4).
Table 4

Intraobserver and interobserver correlation coefficient (r) and Confidence intervel of measurements of the aortic annulus diameter by different imaging modalities.

Interobserver correlationIntraobserver correlation
Correlation95% CICorrelation95% CI
2D systolicr = 0.710.58 to 0.80r = 0.920.87 to 0.95
2D annularr = 0.630.47 to 0.77r = 0.950.92 to 0.97
3D minimum verticalr = 0.920.88 to 0.95r = 0.970.95 to 0.98
3D maximum diameterr = 0.880.82 to 0.92r = 0.960.93 to 0.97
3D annular arear = 0.870.80 to 0.92r = 0.940.90 to 0.96
CT minimum diameterr = 0.900.85 to 0.94r = 0.960.95 to 0.98
CT maximum diameterr = 0.850.78 to 0.91r = 0.940.91 to 0.96
Intraobserver and interobserver correlation coefficient (r) and Confidence intervel of measurements of the aortic annulus diameter by different imaging modalities.

Discussion

Two-D TTE was widely used to screen patients and diagnose the aortic valve and other cardiac diseases.[ Due to the difference between the sizes of aortic valve 3 leaflets, the use of 2D-TTE imaging in assessing aortic annulus based on the distance between 2 valvular hinge points on a 2D-TTE image will give an inaccurate measurement.[ Therefore, many investigators tend to use 3D-TTE imaging techniques, which give more accurate measurements during the examination.[ Thus, the present study was designed to compare 2D-TTE and 3D-TTE measurements of the aortic annular diameter using MDCT as a gold standard. It included 50 patients with normal cardiac anatomy and function assessed by physical examination and standard 2D-TTE. Patients with structural heart disease with 2D-TTE were excluded. In the present study, the age of enrolled cases ranged between 18 years and 60 years, with a mean age of 37.5 ± 10.65. Out of the all patients, 34 (68%) were males, while 16 (32%) were females, 13 (26%) were hypertensive, 10 (20%) were diabetics, their body surface area ranged between 1.23 m2 and 2.1 m2 with (mean ± SD = 1.59 ± 0.22 m2). According to our analysis, there are no correlations between these variables and aortic annular diameter. In terms of systolic diameter in 2D-TTE, the mean diameter was 2.36 ± 0.15 cm. Those findings are somewhat similar to several recent studies such as Chikage et al where the mean diameter was 2.13 ± 0.23 cm.[ This discrepancy in results may be attributed to the previous study does not exclude patients with structural heart disease as 9% of their study population had valvular heart disease and 5% was diagnosed as cardiomyopathy. There was a statistically significant difference between 2D-TTE and MDCT (P = .002) and rough correlation (r = .422). In the present study, the mean annular area by 2D-TTE was 4.33 ± 0.55 cm2, with a significant difference compared with MDCT (P < .001). This finding was somewhat similar to Yoon et al,[ where the mean was 4.30 ± 0.76 cm2. The mean annular area by MDCT was 5.11 ± 0.46 cm2. This finding was somewhat similar to Gurvitch et al,[ where the mean was 5.16 ± 0.32 cm2. In contradiction with the results of Arnold et al, where the mean was 4.57 ± 0.74 cm2.[ This finding was somewhat similar to Chikage et al where they reported a significant difference between mean aortic annular area derived by 2D-TTE 3.60 ± 0.8 and MDCT 4.16 ± 0.7 cm2 with a significant difference (P > .001).[ In the term of 3D-TTE, the mean maximum diameter was 2.64 ± 0.28 cm, while in the study of Chikage et al, the mean was 2.35 ± 0.3 cm.[ There was no significant difference between the mean maximum diameter in 3D-TTE and MDCT (P = .31). These results are somewhat similar to many recent studies such as Schultz et al, where the mean was 2.69 ± 0.28 cm,[ while the results of the study of Khalique et al was 2.56 ± 0.26 cm.[ The aortic annular areas measured by MDCT and 3D-TTE were significantly larger than 2D-TTE (5.11 ± 0.46, 4.79 ± 0.47, and 4.33 ± 0.55 cm2, respectively). A good correlation (r = 0.82) was observed between the values obtained by 3D-TTE and MDCT; however, the correlation between the values by 2D-TTE and MDCT was weak (r = 0.30). Altiok et al reported a high correlation between 3D-TTE and MDCT concerning sagittal diameter r (r = 0.77) and coronal diameter (r = 0.88).[ In the present study, the EI by the 3D-TTE echo ranged between 0.01 and 0.21, with a mean of 0.07 ± 0.06 cm. Moreover, 14 of the 50 patients had EI ≥ 0.1, that is, the aortic annulus was elliptical in 28% of the patients. In contradiction with the results of Chikage et al, where mean EI was 0.11 ± 0.08 cm, and 70 of the 153 patients had EI≥ 0.1, that is, the aortic annulus was elliptical in 46% of the patients.[ This discrepancy in finding may be attributed to different inclusion criteria. Also, the EI by MDCT ranged between 0.01 and 0.93 with a mean of 0.11 ± 0.2 cm. Fifteen of the 50 patients had EI ≥ 0.1, that is, the aortic annulus was elliptical in 30% of the patients. In the comparison between 2D-TTE and 3D-TTE, our findings demonstrated a weak positive correlation between 2D-TTE systolic aortic annular diameter and minimum (P = .003, r = 0.41) and maximum aortic annular diameter derived from 3D-TTE (P = .03, r = .36). In contradiction with the results of Martin et al,[ where the correlation was good between the 2D-TTE and 3D-TTE minimum diameter (r = 0.81, P < .0001) and maximum diameter (r = 0.80, P < .0001). Besides, we reported a statistically significant difference when compared 2D-TTE systolic aortic annular diameter against mean minimum (P = .017) and mean maximum aortic annular diameter derived from 3D-TTE (P < .001). But Martin et al, reported insignificant difference (P = .44). This variation and difference may be due to the demographical variation they enrolled children with mean age 11 ± 3.6 years.[ Moreover, there was a statistically significant difference between the mean aortic annular area derived from 2D-TTE and 3D-TTE (P < .001). These findings are supported by Chikage et al, as they reported a significant difference in both techniques (P > .001). Our findings regarding the minimum diameter in MDCT (2.51 ± 0.13 cm) were somewhat similar to several recent studies such as Arnold et al, where the mean minimum diameter was 25.5 ± 2.1 cm.[ On the other hand, these findings were in contradiction with the findings of the study by Schultz et al, the mean minimum diameter was 2.17 ± 0.21 cm, and this discrepancy in findings may be attributed to the that their study mainly done on patients with severe aortic stenosis and 37% of them had impaired LV systolic function.[ Also, in contradiction with Khalique et al, the mean minimum diameter of their patients was 2.14 ± 0.28 cm due to the different inclusion criteria as their study included 100 patients with severe and symptomatic aortic stenosis who underwent trans-cutaneous aortic valve implantation (TAVI).[ This study highlighted the accuracy of measuring aortic annular diameter by 3D-TEE compared to the 2D-TTE and MDCT. This finding is very important, especially for cardiothoracic surgeons. The preoperative measurement of aortic annular size by 3D-TEE provides them accurate annular information and helps them make a detailed surgical plan before surgery, particularly predicting aortic annular enlargement during the operation. Besides, the accuracy of the 3D-TEE measurement we showed in this study indicates that 3D-TEE could be a feasible alternative to contrast CT for Transcatheter aortic valve implantation (TAVI) patients if contrast CT is not available or contraindicated for some reasons. We performed an image reconstruction by MDCT at the timing of 200 ms from the R wave of the ECG. On the other hand, we measured the aortic annular diameter by 3D-TTE in the mid-systolic phase when the aortic valve was open at maximum. Also, image qualities by 3D-TTE were inferior to those by a 3D-transesophageal echocardiogram (TEE). Another limitation is the small sample size of the included patients; however, this may be due to our restricted criteria.

Conclusion

We conclude that aortic annular areas measured by 3D-TTE had good concordance and correlation with MDCT. The accuracy of aortic annular diameter measurement by 3D-TTE was superior to that by 2D-TTE because the values by 2D-TTE were underestimated compared to those measured by 3D-TTE and MDCT. Three-D TTE and MDCT revealed that the shape of the aortic annulus was elliptical in 28% to 30% respectively of study subjects.

Author contributions

Conceptualization: Khaled I. Naguib, Mohamed A. Attia, Muhammad Saad Reihan, Tarek Ahmed Dabash. Data curation: Khaled I. Naguib, Ahmed B. Abd El-Salam, Hosam H. Helal. Formal analysis: Mohamed S. Bashandy, Ahmed B. Abd El-Salam, Eshak I Bahbah. Investigation: Mohamed A. Attia, Mohamed S. Bashandy, Muhammad Saad Reihan, Tarek Ahmed Dabash, Ahmed B. Abd El-Salam, Hosam H. Helal. Methodology: Mohamed A. Attia, Mohamed S. Bashandy, Muhammad Saad Reihan, Ahmed B. Abd El-Salam, Hosam H. Helal. Project administration: Muhammad Saad Reihan, Eshak I Bahbah. Software: Eshak I Bahbah. Supervision: Khaled I. Naguib, Muhammad Saad Reihan. Visualization: Ahmed B. Abd El-Salam, Eshak I Bahbah. Writing – original draft: Khaled I. Naguib, Mohamed A. Attia, Mohamed S. Bashandy, Muhammad Saad Reihan, Tarek Ahmed Dabash, Ahmed B. Abd El-Salam, Hosam H. Helal, Eshak I Bahbah. Writing – review & editing: Khaled I. Naguib, Mohamed A. Attia, Mohamed S. Bashandy, Muhammad Saad Reihan, Tarek Ahmed Dabash, Ahmed B. Abd El-Salam, Eshak I Bahbah.
  21 in total

1.  Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode.

Authors:  Hatem Alkadhi; Paul Stolzmann; Lotus Desbiolles; Stephan Baumueller; Robert Goetti; Andre Plass; Hans Scheffel; Gudrun Feuchtner; Volkmar Falk; Borut Marincek; Sebastian Leschka
Journal:  Heart       Date:  2010-06       Impact factor: 5.994

2.  Aortic annulus diameter determination by multidetector computed tomography: reproducibility, applicability, and implications for transcatheter aortic valve implantation.

Authors:  Ronen Gurvitch; John G Webb; Ren Yuan; Mark Johnson; Cameron Hague; Alexander B Willson; Stefan Toggweiler; David A Wood; Jian Ye; Robert Moss; Christopher R Thompson; Stephan Achenbach; James K Min; Troy M Labounty; Ricardo Cury; Jonathon Leipsic
Journal:  JACC Cardiovasc Interv       Date:  2011-11       Impact factor: 11.195

3.  EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography.

Authors:  Roberto M Lang; Luigi P Badano; Wendy Tsang; David H Adams; Eustachio Agricola; Thomas Buck; Francesco F Faletra; Andreas Franke; Judy Hung; Leopoldo Pérez de Isla; Otto Kamp; Jaroslaw D Kasprzak; Patrizio Lancellotti; Thomas H Marwick; Marti L McCulloch; Mark J Monaghan; Petros Nihoyannopoulos; Natesa G Pandian; Patricia A Pellikka; Mauro Pepi; David A Roberson; Stanton K Shernan; Girish S Shirali; Lissa Sugeng; Folkert J Ten Cate; Mani A Vannan; Jose Luis Zamorano; William A Zoghbi
Journal:  J Am Soc Echocardiogr       Date:  2012-01       Impact factor: 5.251

4.  Comparison of two- and three-dimensional transthoracic echocardiography for measurement of aortic annulus diameter in children.

Authors:  Romain Martin; Sébastien Hascoët; Yves Dulac; Marianne Peyre; Simon Mejean; Khaled Hadeed; Alexandre Cazavet; Bertrand Leobon; Philippe Acar
Journal:  Arch Cardiovasc Dis       Date:  2013-09-23       Impact factor: 2.340

5.  Aortic annular sizing using a novel 3-dimensional echocardiographic method: use and comparison with cardiac computed tomography.

Authors:  Omar K Khalique; Susheel K Kodali; Jean-Michel Paradis; Tamim M Nazif; Mathew R Williams; Andrew J Einstein; Gregory D Pearson; Kishore Harjai; Kendra Grubb; Isaac George; Martin B Leon; Rebecca T Hahn
Journal:  Circ Cardiovasc Imaging       Date:  2013-11-12       Impact factor: 7.792

6.  Three dimensional evaluation of the aortic annulus using multislice computer tomography: are manufacturer's guidelines for sizing for percutaneous aortic valve replacement helpful?

Authors:  Carl J Schultz; Adriaan Moelker; Nicolo Piazza; Apostolos Tzikas; Amber Otten; Rutger J Nuis; Lisan A Neefjes; Robert J van Geuns; Pim de Feyter; Gabriel Krestin; Patrick W Serruys; Peter P T de Jaegere
Journal:  Eur Heart J       Date:  2009-12-07       Impact factor: 29.983

7.  2017 ACC Expert Consensus Decision Pathway for Transcatheter Aortic Valve Replacement in the Management of Adults With Aortic Stenosis: A Report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents.

Authors:  Catherine M Otto; Dharam J Kumbhani; Karen P Alexander; John H Calhoon; Milind Y Desai; Sanjay Kaul; James C Lee; Carlos E Ruiz; Christina M Vassileva
Journal:  J Am Coll Cardiol       Date:  2017-01-04       Impact factor: 24.094

Review 8.  The role of noninvasive and invasive diagnostic imaging techniques for detection of extra-cranial venous system anomalies and developmental variants.

Authors:  Kresimir Dolic; Adnan H Siddiqui; Yuval Karmon; Karen Marr; Robert Zivadinov
Journal:  BMC Med       Date:  2013-06-27       Impact factor: 8.775

9.  Appreciating the strengths and weaknesses of transthoracic echocardiography in hemodynamic assessments.

Authors:  Stephen J Huang; Anthony S McLean
Journal:  Cardiol Res Pract       Date:  2012-02-20       Impact factor: 1.866

Review 10.  3D transoesophageal echocardiography in the TAVI sizing arena: should we do it and how do we do it?

Authors:  Caroline Bleakley; Mehdi Eskandari; Mark Monaghan
Journal:  Echo Res Pract       Date:  2017-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.