Wouter B Vehmeijer1, Kelly Jonkman2, Sveinn H Hardarson3, Leon Aarts2, Einar Stefansson3,4, Albert Dahan2, Nicoline E Schalij-Delfos1.
Abstract
PURPOSE: Retinal oximetry measures oxygen saturation in retinal vessels. With the introduction of a mobile handheld prototype oximeter, this technique will become available for a broader patient population including bedridden patients and newborn babies. The objective is to determine the sensitivity of this handheld oximeter in room air and during isocapnic hyperoxia. A comparison is made between the handheld oximeter and the Oxymap T1.Entities:
Keywords: fundus imaging; handheld oximeter; hyperoxia; retinal oximetry
Mesh:
Substances:
Year: 2021 PMID: 33650317 PMCID: PMC9543559 DOI: 10.1111/aos.14817
Source DB: PubMed Journal: Acta Ophthalmol ISSN: 1755-375X Impact factor: 3.988
Fig. 1Retinal oximetry in normoxia and hyperoxia with both oximeters. (A) Mobile Handheld Oximeter (Corimap Camera) in its stationary mode. Retinal oximetry images with the Handheld Oximeter (A = normoxia, C = hyperoxia). (a) The current ‘gold’ standard Oxymap T1. Retinal oximetry with the Oxymap T1 (b = normoxia and c = hyperoxia).
Physiological variables during normoxia, hyperoxia and recovery for both oxygenation sessions
| Session I | Session II | |||||
|---|---|---|---|---|---|---|
| Baseline | Hyperoxia | Recovery | Baseline | Hyperoxia | Recovery | |
| Physiological variables | ||||||
| Ventilation (L/min) | 9.4 ± 1.4 | 15.6 ± 3.0 | 13.5 ± 4.1 | 10.0 ± 1.9 | 17.4 ± 3.8 | 13.8 ± 3.5 |
| Tidal volume (ml) normoxia | 723.6 ± 170 | 915 ± 168 | 800 ± 249 | 671 ± 107 | 935 ± 171 | 756 ± 307 |
| Inspired carbon dioxide (PiCO2) | 0.17 ± 0.2 | 2.05 ± 0.7 | 1.77 ± 5.9 | 0.14 ± 0.2 | 2.4 ± 0.8 | 2.9 ± 6.7 |
| End‐tidal carbon dioxide (PetCO2) | 4.8 ± 1.5 | 5.3 ± 0.5 | 6.0 ± 5.4 | 4.9 ± 0.4 | 5.07 ± 0.9 | 6.2 ± 5.1 |
| Inspired oxygen (PiO2) | 20.0 ± 4.5 | 87.27 ± 2.5 | 23.0 ± 1.5 | 21.0 ± 1.1 | 81.7 ± 17.5 | 26.2 ± 5.6 |
| Expired oxygen (PeO2) | 15.3 ± 0.8 | 82.5 ± 2.3 | 20.7 ± 1.9 | 15.5 ± 0.9 | 77.3 ± 16.8 | 23.6 ± 5.7 |
| Heart rate (bpm) | 62 ± 12 | 62 ± 9 | 70 ± 18 | 57 ± 41 | 61 ± 11 | 71 ± 14 |
| Oxygen saturation (%) | 98.5 ± 1.2 | 99.8 ± 0.4 | 98.9 ± 1.9 | 99.0 ± 1.4 | 99.5 ± 0.6 | 99.2 ± 1.2 |
Retinal saturation and vessel diameter in normoxia, hyperoxia and after recovery
| Normoxia | Hyperoxia | Recovery | ||||
|---|---|---|---|---|---|---|
| Corimap camera | Oxymap T1 | Corimap camera | Oxymap T1 | Corimap camera | Oxymap T1 | |
| Oxygen saturation (Mean ± SD) | ||||||
| Arterioles | 91.3% ± 3.9% | 94.0% ± 2.8% | 94.6% ± 4.9% | 95.4% ± 3.2% | 89.2% ± 5.2% | 93.2% ± 2.6% |
| Venules | 56.3% ± 9.8% | 58.9% ± 5.3% | 82.2% ± 7.4% | 84.3% ± 4.0% | 54.9% ± 12.8% | 61.8% ± 5.5% |
| AV difference | 35.0% ± 7.3% | 35.0% ± 5.6% | 12.3% ± 4.2% | 11.2% ± 3.4% | 34.3% ± 9.3% | 31.4% ± 5.1% |
| Diameter (Pixels, Mean ± SD) | ||||||
| Arterioles | 14.4 ± 1.3 | 13.08 ± 1.18 | 13.5 ± 1.6 | 12.08 ± 1.51 | 13.6 ± 1.1 | 12.60 ± 1.20 |
| Venules | 18.4 ± 1.4 | 15.83 ± 1.22 | 16.7 ± 2.0 | 14.92 ± 1.34 | 17.3 ± 1.4 | 15.36 ± 0.99 |