A novel quality assurance procedure for trajectory log validation using phantom-less real-time latency corrected EPID images.

Seng Boh Lim1, Benjamin J Zwan2,3, Danny Lee2,4, Peter B Greer2,5, Dale Michael Lovelock1.   

Abstract

The use of trajectory log files for routine patient quality assurance is gaining acceptance. Such use requires the validation of the trajectory log itself. However, the accurate localization of a multileaf collimator (MLC) leaf while it is in motion remains a challenging task. We propose an efficient phantom-less technique using the EPID to verify the dynamic MLC positions with high accuracy. Measurements were made on four Varian TrueBeams equipped with M120 MLCs. Two machines were equipped with the S1000 EPID; two were equipped with the S1200 EPID. All EPIDs were geometrically corrected prior to measurements. Dosimetry mode EPID measurements were captured by a frame grabber card directly linked to the linac. All leaf position measurements were corrected both temporally and geometrically. The readout latency of each panel, as a function of pixel row, was determined using a 40 × 1.0 cm2 sliding window (SW) field moving at 2.5 cm/s orthogonal to the row readout direction. The latency of each panel type was determined by averaging the results of two panels of the same type. Geometric correction was achieved by computing leaf positions with respect to the projected isocenter position as a function of gantry angle. This was determined by averaging the central axis position of fields at two collimator positions of 90° and 270°. The radiological to physical leaf end position was determined by comparison of the measured gap with that determined using a feeler gauge. The radiological to physical leaf position difference was found to be 0.1 mm. With geometric and latency correction, the proposed method was found to be improve the ability to detect dynamic MLC positions from 1.0 to 0.2 mm for all leaves. Latency and panel residual geometric error correction improve EPID-based MLC position measurement. These improvements provide for the first time a trajectory log QA procedure.
© 2021 The Authors. Journal of Applied Clinical Medical Physics is published by Wiley Periodicals LLC. on behalf of The American Association of Physicists in Medicine.

Entities:  

Keywords:  EPID; log file analysis; patient-specific QA; quality assurance

Mesh:

Year:  2021        PMID: 33634952      PMCID: PMC7984475          DOI: 10.1002/acm2.13202

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  14 in total

1.  Automating linear accelerator quality assurance.

Authors:  Tobias Eckhause; Hania Al-Hallaq; Timothy Ritter; John DeMarco; Karl Farrey; Todd Pawlicki; Gwe-Ya Kim; Richard Popple; Vijeshwar Sharma; Mario Perez; SungYong Park; Jeremy T Booth; Ryan Thorwarth; Jean M Moran
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  Commissioning and quality assurance for VMAT delivery systems: An efficient time-resolved system using real-time EPID imaging.

Authors:  Benjamin J Zwan; Michael P Barnes; Jonathan Hindmarsh; Seng B Lim; Dale M Lovelock; Todsaporn Fuangrod; Daryl J O'Connor; Paul J Keall; Peter B Greer
Journal:  Med Phys       Date:  2017-07-10       Impact factor: 4.071

3.  Measured vs simulated portal images for low MU fields on three accelerator types: possible consequences for 2D portal dosimetry.

Authors:  Mark Podesta; Sebastiaan M J J G Nijsten; Julia Snaith; Marc Orlandini; Tim Lustberg; Davy Emans; Trent Aland; Frank Verhaegen
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

4.  Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries.

Authors:  A Agnew; C E Agnew; M W D Grattan; A R Hounsell; C K McGarry
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

5.  A clinically observed discrepancy between image-based and log-based MLC positions.

Authors:  Brian Neal; Mahmoud Ahmed; Kunal Kathuria; Tyler Watkins; Krishni Wijesooriya; Jeffrey Siebers
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

6.  Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines.

Authors:  Weiliang Du; Song Gao; Wei Jiang; Rajat J Kudchadker
Journal:  J Appl Clin Med Phys       Date:  2018-06-29       Impact factor: 2.102

7.  Evaluation of IsoCal geometric calibration system for Varian linacs equipped with on-board imager and electronic portal imaging device imaging systems.

Authors:  Song Gao; Weiliang Du; Peter Balter; Peter Munro; Andrew Jeung
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

8.  Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT.

Authors:  Christina E Agnew; Denise M Irvine; Conor K McGarry
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

9.  An EPID-based system for gantry-resolved MLC quality assurance for VMAT.

Authors:  Benjamin J Zwan; Michael P Barnes; Todsaporn Fuangord; Cameron J Stanton; Daryl J O'Connor; Paul J Keall; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation.

Authors:  Hideaki Hirashima; Yuki Miyabe; Mitsuhiro Nakamura; Nobutaka Mukumoto; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2018-04-06       Impact factor: 2.102

View more
  1 in total

1.  Insensitivity of machine log files to MLC leaf backlash and effect of MLC backlash on clinical dynamic MLC motion: An experimental investigation.

Authors:  Michael Barnes; Dennis Pomare; Marcus Doebrich; Therese S Standen; Joshua Wolf; Peter Greer; John Simpson
Journal:  J Appl Clin Med Phys       Date:  2022-06-09       Impact factor: 2.243

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.