| Literature DB >> 33608759 |
Alice Finotello1, Giovanni Spinella2, Giulia Notini3, Domenico Palombo3, Giovanni Pratesi3, Simone Mambrini3, Ferdinando Auricchio4, Michele Conti4, Bianca Pane3.
Abstract
PURPOSE: To evaluate bridging stent geometry in patients who underwent branched endovascular aortic repair (B-EVAR) and to correlate the outcomes with intrinsic bridging stent characteristics aiming to identify the stent(s) that guarantees the best performance.Entities:
Keywords: B-EVAR; Balloon-expanding stent-graft; Bridging stent; Geometric analysis; Self-expanding stent-graft; Thoraco-abdominal aortic aneurysm
Mesh:
Year: 2021 PMID: 33608759 PMCID: PMC8060236 DOI: 10.1007/s00270-021-02773-w
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Fig. 1Schematic representation of computed parameters on CTA. A Pre-operative take-off angle (TOA); B post-operative TOA; C post-operative stent subdivision into zone 1 (light blue), zone 2 (green), and zone 3 (yellow) to compute centerline lengths and sections shape index
Patients’ demographics
| Patients | 11 |
| Sex; male/female | 10/1 |
| Age; years | 77 |
| Dislipidemia; n | 3 |
| Hypertension; n | 9 |
| Coronary artery disease; n | 6 |
| Diabetes mellitus; n | 1 |
| COPD; n | 6 |
| Renal Disease; n | 3 |
| Prior aortic repair; n | 5 |
| Custom-made; n (bridging stents adopted) | 3 (2 Fluency, 2 Covera, 6 VBX) |
| Off-the-shelf; n (bridging stents adopted) | 8 (8 Fluency, 11 Covera, 10 VBX) |
| Vessels; total n | 38 |
| CTR; n (bridging stents adopted) | 11 (3 Fluency, 2 Covera, 6 VBX) |
| SMA; n (bridging stents adopted) | 11 (3 Fluency, 4 Covera, 4 VBX) |
| LRA; n ((bridging stents adopted) | 8 (2 Fluency, 3 Covera, 3 VBX) |
| RRA; n (bridging stents adopted) | 8 (2 Fluency, 3 Covera, 3 VBX) |
| Fluency; n (vessels) | 10 (3 CTR, 3 SMA, 2 LRA, 2 RRA) |
| Covera; n (vessels) | 12 (2 CTR, 4 SMA, 3 LRA, 3 RRA) |
| VBX; n (vessels) | 16 (6 CTR, 4 SMA, 3 LRA, 3 RRA) |
CTR celiac trunk; SMA superior mesenteric artery; LRA left renal artery; RRA right renal artery
Pre-operative geometrical evaluations
| Total | CTR | SMA | LRA | RRA | ||
|---|---|---|---|---|---|---|
| Target vessel diameter, mm | 7.0 ± 1.1 | 8.1 ± 0.6 | 7.8 ± 1.1 | 6.1 ± 1.1 | 5.9 ± 1.1 | = 0.04* |
| Nominal stent diameter, mm | 7.6 ± 1.3 | 8.5 ± 0.9 | 8.7 ± 0.7 | 6.4 ± 0.7 | 6.4 ± 0.7 | = 0.05* |
| Pre-operative TOA, ° | 109.86 ± 28.65 | 117.5 ± 28.4 | 115.8 ± 28 | 104.5 ± 28.9 | 97.3 ± 29.6 | = 0.4 |
| Aneurysm diameter (TOA level), mm | 42.56 ± 11.88 | 50.9 ± 10.2 | 45.8 ± 13.9 | 34.7 ± 5.9 | 35.5 ± 6.4 | = 0.003*** |
| Lumen diameter (TOA level), mm | 32.17 ± 6.27 | 36.4 ± 6.1 | 34.3 ± 6.9 | 27.3 ± 2.5 | 28.5 ± 2.3 | = 0.004*** |
TOA take-off angle. Statistical significance: “*” for 0.05; “**” for 0.005, and “***” for < 0.005
Post-operative geometrical evaluations
| Total | Fluency | Covera | VBX | CTR | SMA | LRA | RRA | |||
|---|---|---|---|---|---|---|---|---|---|---|
| TOA, ° | 150.27 ± 21 | 156.75 ± 13.74 | 158.2 ± 11.7 | 140.6 ± 26.2 | 154.1 ± 15.8 | 156.5 ± 21.6 | 148.4 ± 22.5 | 140.0 ± 23.4 | ||
| Zone 1 —length, mm | 18.11 ± 0.93 | 17.71 ± 0.63 | 18.19 ± 0.92 | 18.37 ± 1.02 | 18.5 ± 1.3 | 18.2 ± 0.9 | 17.7 ± 0.7 | 18.0 ± 0.04 | ||
| Zone 1 —SI, - | 0.89 ± 0.04 | 0.88 ± 0.03 | 0.87 ± 0.04 | 0.91 ± 0.03 | 0.89 ± 0.04 | 0.89 ± 0.06 | 0.87 ± 0.06 | 0.88 ± 0.04 | ||
| Zone 2 –length, mm | 30.97 ± 10.0 | 27.7 ± 8.6 | 32.33 ± 12.4 | 32 ± 9.0 | 32.8 ± 11.7 | 31.2 ± 7.3 | 28.6 ± 6.2 | 30.5 ± 14.2 | ||
| Zone 2 —SI, - | 0.88 ± 0.06 | 0.81 ± 0.06 | 0.89 ± 0.03 | 0.92 ± 0.02 | 0.88 ± 0.07 | 0.88 ± 0.06 | 0.88 ± 0.05 | 0.88 ± 0.05 | ||
| Zone 3 —length, mm | 21.84 ± 10.27 | 23.5 ± 7.26 | 22.58 ± 11.4 | 19.25 ± 11.3 | 21.1 ± 8.8 | 19.2 ± 8.0 | 22.0 ± 11.9 | 24.4 ± 13.3 | ||
| Zone 3 —SI, - | 0.88 ± 0.06 | 0.81 ± 0.05 | 0.88 ± 0.05 | 0.93 ± 0.02 | 0.88 ± 0.06 | 0.89 ± 0.06 | 0.87 ± 0.06 | 0.89 ± 0.03 | ||
| Nominal stent length, mm | 70.26 ± 11.3 | 66.67 ± 10.0 | 72.73 ± 13.5 | 70.64 ± 10.5 | 68.7 ± 11.0 | 69.6 ± 10.3 | 69.6 ± 10.6 | 74.1 ± 12.2 | ||
| Stent total length, mm | 65.61 ± 11.3 | 63.00 ± 6.9 | 71.00 ± 12.6 | 63.08 ± 11.8 | 63.7 ± 11.8 | 66.1 ± 8.8 | 65.6 ± 11.2 | 68.8 ± 14.3 | ||
| Stent shortening, % | − 6.58 ± 6.42 | − 4.86 ± 4.63 | − 2.23 ± 2.18 | − 11.12 ± 5.65 | − 7.4 ± 5.8 | − 4.7 ± 4.3 | − 5.9 ± 5.3 | − 6.8 ± 6.6 |
TOA take-off angle; SI shape index. Statistical significance: “*” for 0.05; “**” for 0.005, and “***” for < 0.005
Fig. 2Bar plot showing differences in shape index between stent types in zones 1, 2, and 3 with significance of the different entries (“*” for p < 0.05). Error bars denote standard deviation
Fig. 3Shape index was computed for each section along the centerline path. Comparative outcomes concerning the three stent types (Fluency, Covera, and VBX) are reported. Color scale ranges from blue (minimum value; SI = 0.5) to red (maximum value; SI = 1). As highlighted, for the three comparative cases under consideration, better results are obtained for VBX stent